SlideShare a Scribd company logo
Validation of Qualitative Lab
Test Methods
By: Dr Mostafa Mahmoud MD,
Consultant Microbiologist
Labs & Blood Banks Admin, Riyadh, MOH
Associate Professor of Medical Microbiology & Immunology
Faculty of Medicine – Ain Shams University
Procedures for validation
• Each method validation/ verification study is a collection of
experiments to assess performance and error in order to judge a
method’s suitability for use in the laboratory.
• Acceptability Criteria
• the laboratory must establish acceptance criteria.
• Parameters for accuracy, precision, sensitivity and specificity should
include a confidence level of at least 90%, or meet the claims of the
manufacturer.
Qualitative Methods:
• Includes semi quantitative testing that use cut offs such as hepatitis
testing and some molecular testing.
• No values/concentrations are included in the patient report.
• Test results are reported as positive/negative, normal/
borderline/abnormal, reactive/nonreactive, detected /not detected.
•1. FDA cleared or approved methods (CLIA Standards).
• Parameters to be validated:
1. Accuracy (correlation or comparison)
2. Precision (Replication)
3. Reportable range
4. Reference range (Normal values)
1- Accuracy (correlation or comparison)
• Demonstrates how close to the “true” value the new method can
achieve.
The materials (Samples) used:
1- Calibrators/controls,
2- Reference material,
3- Proficiency testing material with known values,
4- Samples tested by another lab using the same or similar method,
5- or by comparing results to an established comparative method.
Method & Number of samples:
• Most sources recommend comparing at least 40 patient specimens.
CLIA current guidance suggests a minimum of 20 samples.
• A method comparison must be used for 5 days and for another 5
days if failed.
• Document the results of the new method comparing the known
values from the reference sources, another certified laboratory’s
results or with results from the current method.
Calculations:
• Calculate the percent of positive, negative and total accuracy by
dividing observed results over known results multiplied by 100.
• Example: New method = 19 positives, 20 negatives. Current
method or reference material with known values = 20 positive, 20
negatives
• Percent positive accuracy 19/20 X 100 = 95%
• Percent negative accuracy 20/20 X 100 = 100%
• Total accuracy 39/40 X 100 = 98%
Validation of qualitative lab test  methods
2- Precision (Reproducibility ):
• Can the new method duplicate the same results?
•Materials used:
1- patient samples are the first choice followed by
2- control material and reference solutions.
Method & Number of samples:
- 2 negative samples and 2 positive samples run in triplicate for 5
days will provide data for within-run and between-run
components to estimate precision.
•
Calculation:
• Calculate the percent within-run (intra), between-run (inter) and
total precision by dividing observed results over known results
multiplied by 100.
• See next tables.
ID Day 1 Day 2 Day 3
Pos sample Pos Pos Pos Pos Pos Pos Pos Pos Pos
Pos sample Pos Pos Pos Neg Pos Pos Pos Pos Pos
Neg sample Neg Neg Neg Neg Neg Neg Neg Neg Neg
Neg sample Neg Neg Neg Neg Neg Neg Neg Neg Neg
Within run % 12/12/x 100 = 100% 11/12 x 100 = 92% 12/12/x 100 =
100%
ID Day 4 Day 5 Between run %
Pos sample Pos Pos Pos Pos Pos Pos 15/15x100 = 100%
Pos sample Pos Pos Pos Neg Pos Pos 14/15x100 =93%
Neg sample Neg Neg Neg Neg Neg Pos 14/15x100 =93%
Neg sample Neg Neg Neg Neg Neg Neg 15/15x100 = 100%
Within run % 12/12/x 100 = 100% 11/12 x 100 = 92%
Total Precision: 58/60 x 100 = 96.7%
3- Reportable Range:
• The highest and lowest test values that can be analyzed while
maintaining accuracy.
• For tests without high or low values, define method criteria for a
positive result.
Method & Number of samples:
• To verify reportable range, test at least 3-5 low and high positive
samples once.
• These samples can be combined with the accuracy/precision
experiments.
• Include both weak and strong positive samples.
• For methods depending on a cut-off value to determine positive
results, testing positive specimens near the cut-off can serve as
the cut-off validation.
•CAP Requirement:
• For qualitative tests that use a cut-off value to distinguish
positive from negative, the cut-off value is established initially,
and verified every 6 months thereafter.
• N.B. This requirement does not apply to FDA-cleared/approved in
vitro diagnostic assays that report the qualitative result based on a
predefined cut-off value.
• Applied to laboratory established threshold (cut-off value)
4- Reference Range (Normal Values):
• Provided by the manufacturer and verified by running known normal
patients.
• Manufacturer’s ranges or even published reference ranges in
textbooks are used if the patients are similar.
•Method & Number of samples:
• The Reference Range can be verified by testing 20 known normal
samples; if no more than 2 results fall outside the
manufacturer/published range then that reference range can be
considered to be verified.
• If the laboratory cannot reference the normal values, then the
reference range will need to be established. This involves a
selection of at least 120 reference samples for each group or
subgroup that needs to be characterized.
5- Sensitivity & Specificity:
• CLIA does not require that these parameters to be verified.
Validation Summary for FDA-approved qualitative tests:
• Clearly state the purpose of the verification, what
platform/method and the number of samples for each experiment.
Any discrepant results should be investigated and explained in the
Summary. Test results that show sample problems such as
contamination and degradation should not be used in the
assessment but still listed with an explanation.
• The Summary should also contain a Conclusion stating weather the
study met the acceptance criteria or not and its suitability for us in
the laboratory.
• Additional samples to be added if results are outside acceptance.
• If the results failed then test not to be implemented.
2. Non-FDA Cleared tests
• Qualitative methods developed in-House, non-FDA cleared
methods and FDA-cleared methods modified by the laboratory.
• - For Qualitative methods follow the instructions above for
Accuracy, Precision, Reportable Range and
Reference Range.
•Qualitative Sensitivity is not done by analytic sensitivity
but by diagnostic sensitivity.
• Diagnostic sensitivity: The percent of subjects (patients) with the
target condition whose test values are positive.
•Method &number of samples:
• dividing the number of true positives by the sum of the number of
true positives plus the number of false negatives and multiplying by
100.
Diagnostic Sensitivity: = [TP ÷(TP + FN)] x 100
- If a person has a disease, how often will the test be positive (true
positive rate).
- The higher the sensitivity the lower the False Negative and vice
versa.
•Qualitative Specificity
• Qualitative sensitivity not assessed by Analytical Specificity but
also by the Diagnostic Specificity.
• Diagnostic Specificity: the percent of subjects (Persons) without
the target condition whose test values are negative.
•Method & number of samples:
• Calculate by dividing the number of true negatives by the sum of
the number of true negatives plus the number of false positives
and multiplying by 100.
Diagnostic Specificity: [TN ÷(TN + FP)] x 100
-If a person does not have the disease, how often will the test be
negative ( true negative rate).
- The higher the specificity the lower the False Positive and vice
versa.
Interference:
• The laboratory must be aware of common interfering substances by
referencing studies performed elsewhere (manufacturer or literature)
or by performing studies.
• Common blood interferences are; hemolysis, bilirubin, lipemia,
preservatives and anticoagulants used in specimen collection.
CAP Accreditation requirements for validating
laboratory tests
FDA approved/cleared LDTs & modified FDA tests
Accuracy & Precision
(COM.40300, ph II)
Verify manufacturer’s results Establish (= validation)
Analytic sensitivity (LOD)
(COM.40400, ph II)
Verify: manufacturer or
literature documentation OK
Establish
Analytic specificity
(interferences)
(COM.40500, ph II)
Reference literature or
manufacturer documentation
Establish; studies by
manufacturer or in literature
OK
Reportable range (AMR)
(COM.40600, ph II)
Verify* Establish*
Reference range
(COM.50000, ph II)
Verify ** Establish**
Test method validation CBAHI standard
LB.10 The laboratory develops a process for test method validation.
LB.10.1 The laboratory implements policies and procedures on
test method validation including:
LB.10.1.1 Verification of accuracy/precision.
LB.10.1.2 Verification of sensitivity (lower detection limit).
LB.10.1.3 Verification of carryover acceptability.
LB.10.1.4 Verification of the Analytic Measurement (AMR)
LB.10.1.5 Approval of the method for clinical use (CRR??).
Test Method Validation
Evidence of compliance to (LB 10) standard:
1- Document Reviews:
- Policies, process and procedures on method validation.
2- Documented Evidences
- Records surveyor-selected method confirms compliance with
policies and procedures.
3- Staff Interview
- Senior personnel are knowledgeable about the concept of
method validation.
• LB.13 The laboratory has a system for instruments/methods
correlation.
LB.13.1 When the laboratory uses more than one method and/or
instruments to test for a given analyte, the laboratory develops
and implements policies and procedures on correlation to ensure
the following:
LB.13.1.1 The correlation studies are conducted every six months.
LB.13.1.2 There is clear description of the correlation study.
LB.13.1.3 There are clearly defined acceptance criteria.
LB.13.1.4 There is a process for review and approval of the
correlation results.
• Evidence of Compliance (LB.13):
• Document Review
- Policies, process and procedures on methods/instrument
correlation.
• Documented Evidences
- Records surveyor-selected methods/instruments confirms
compliance with policies and procedures.
• Staff Interview
- Senior personnel are knowledgeable about the purpose of
methods/instrument correlation.
• LB.14 The laboratory has a system for controlling the quality of test
methods.
LB.14.1 The laboratory implements policies and procedures on quality
control of test methods to satisfy the following:
LB.14.1.1 Assignment of performance and review responsibility (control
specimens are handled and tested in the same manner and by the same
laboratory personnel testing patient samples).
LB.14.1.2 Number and frequency of running controls.
LB.14.1.3 Tolerance limits of controls results.
LB.14.1.4 Corrective action to be taken in the event of unacceptable results.
LB.14.2 The laboratory quality control system conforms to the
manufacturer's instructions.
•Evidence of compliance to LB 14
• Document review:
- Policies, process and procedures on controlling the quality of test
methods.
- Policies and procedures on controlling the quality of test methods
confirm to the manufacturer instructions.
• Documented Evidences:
- Records surveyor-selected methods confirms compliance with
policies and procedures.
• Staff Interview:
- Laboratory personnel understanding of the purpose of controlling the
quality of test methods.
Validation of qualitative lab test  methods

More Related Content

PDF
Laboratory Method Verification, March 2017
PPTX
Validation of lab instruments and quantitative test methods
PDF
Method verification
PPTX
Medical records
PDF
Verifikasi dan Validasi Metode Pemeriksaan Laboratorium Klinik Bagian 1
PDF
كتيب الجودة في الفحوصات المخبريه
PPT
anaerobic infection ppt.
PPTX
Reticulocyte count
Laboratory Method Verification, March 2017
Validation of lab instruments and quantitative test methods
Method verification
Medical records
Verifikasi dan Validasi Metode Pemeriksaan Laboratorium Klinik Bagian 1
كتيب الجودة في الفحوصات المخبريه
anaerobic infection ppt.
Reticulocyte count

What's hot (20)

PPT
Clinical lab-method validation.ppt
PDF
Internal quality control in clinical laboratories hematology(2)
PPT
Quality control lecture CPath master 2014 Ain Shams
PDF
Troubleshooting Poor EQA/QC Performance in the Laboratory
PDF
Troubleshooting QC Problems: Your QC has failed, what do you do next?
PDF
How to Measure Uncertainty
PPTX
Quality Control In Clinical Laboratory
PPTX
Internal Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
PDF
Harmonization of Laboratory Indicators, 09 03-2017
PPT
PDF
Quality control in clinical laboratories
PPTX
Quality Control in a Medical Testing Laboratory
PPT
Quality control
PPTX
Total Quality Management in Medical Laboratories
PPTX
Quality Management System
PPTX
(IQC) - Internal Quality Control.pptx
PPTX
Controlling clinical laboratory errors
PDF
Laboratory audit ls
PPTX
Calibration and Quality controls of automated hematology analyzer
PPTX
Troubleshooting IQC / EQAS
Clinical lab-method validation.ppt
Internal quality control in clinical laboratories hematology(2)
Quality control lecture CPath master 2014 Ain Shams
Troubleshooting Poor EQA/QC Performance in the Laboratory
Troubleshooting QC Problems: Your QC has failed, what do you do next?
How to Measure Uncertainty
Quality Control In Clinical Laboratory
Internal Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
Harmonization of Laboratory Indicators, 09 03-2017
Quality control in clinical laboratories
Quality Control in a Medical Testing Laboratory
Quality control
Total Quality Management in Medical Laboratories
Quality Management System
(IQC) - Internal Quality Control.pptx
Controlling clinical laboratory errors
Laboratory audit ls
Calibration and Quality controls of automated hematology analyzer
Troubleshooting IQC / EQAS
Ad

Similar to Validation of qualitative lab test methods (20)

PPT
Method validation for calibration measure.ppt
PPTX
Analytical method validation
PDF
VERIFICATION AND VALIDATION TOOLKIT Determining Performance Characteristics o...
PDF
Analytical Method Validation basics by Dr. A. Amsavel
PPT
chem II 2019.ppt
PPTX
'Validation of system Protocol for Students'.pptx
PPTX
Accuracy and precision presentation
PPTX
Principles of Laboratory Medicine.pptx imp
PDF
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
PDF
How to Validate a New Test.pdf
PPTX
Analytical Method Validation
PPT
HPLC validation.ppt
PDF
method_validation_molinaro_laboratory.pdf
PPTX
Analytical Method Validation.pptx
PDF
Method Validation in Medical Laboratory..
PPTX
QUALITY CHECK IN A BIOCHEMISTRY LAB SYSTEM
PPTX
Control of analytical quality using stable control materials postgrad
PPTX
Validation of laboratory tests with important guidelines
PPTX
Chemistry QC, Calibration&Reportable Range REFINED.pptx
PPT
Validation of Analytical method.ppt
Method validation for calibration measure.ppt
Analytical method validation
VERIFICATION AND VALIDATION TOOLKIT Determining Performance Characteristics o...
Analytical Method Validation basics by Dr. A. Amsavel
chem II 2019.ppt
'Validation of system Protocol for Students'.pptx
Accuracy and precision presentation
Principles of Laboratory Medicine.pptx imp
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
How to Validate a New Test.pdf
Analytical Method Validation
HPLC validation.ppt
method_validation_molinaro_laboratory.pdf
Analytical Method Validation.pptx
Method Validation in Medical Laboratory..
QUALITY CHECK IN A BIOCHEMISTRY LAB SYSTEM
Control of analytical quality using stable control materials postgrad
Validation of laboratory tests with important guidelines
Chemistry QC, Calibration&Reportable Range REFINED.pptx
Validation of Analytical method.ppt
Ad

More from Mostafa Mahmoud (18)

PPTX
KPIs in Microbiology
PPTX
Serological Diagnosis of Infectious Diseases
PPTX
Lab diagnosis of Sexually transmitted Infections (STIs)
PPT
Conventional methods for bacterial identification
PPTX
Microbiology lab QC
PPT
Bacterial Culture media
PPT
Manual blood culture Techniques
PPTX
Antibiogram CLSI Recommendations
PPTX
Lab diagnosis of tb dr mostafa lecture
PPTX
Viral Hemorrhagic Fevers
PPTX
Multidrug Resistant Oraganisms (MDRO) infection control
PPTX
Hiv lab diagnosis
PPTX
Prevention and Control of AIDS for World AIDS day
PPTX
MERS CoV MOH Guidelines update 2015
PPTX
CSF Examination
PPTX
Laboratory detection of resistant bacteria
PPTX
Anaerobic Culture Methods
PPTX
KPIs in Microbiology
Serological Diagnosis of Infectious Diseases
Lab diagnosis of Sexually transmitted Infections (STIs)
Conventional methods for bacterial identification
Microbiology lab QC
Bacterial Culture media
Manual blood culture Techniques
Antibiogram CLSI Recommendations
Lab diagnosis of tb dr mostafa lecture
Viral Hemorrhagic Fevers
Multidrug Resistant Oraganisms (MDRO) infection control
Hiv lab diagnosis
Prevention and Control of AIDS for World AIDS day
MERS CoV MOH Guidelines update 2015
CSF Examination
Laboratory detection of resistant bacteria
Anaerobic Culture Methods

Recently uploaded (20)

PPT
Copy-Histopathology Practical by CMDA ESUTH CHAPTER(0) - Copy.ppt
PPT
Obstructive sleep apnea in orthodontics treatment
PPTX
Stimulation Protocols for IUI | Dr. Laxmi Shrikhande
PPTX
Acid Base Disorders educational power point.pptx
PPTX
DENTAL CARIES FOR DENTISTRY STUDENT.pptx
PPT
MENTAL HEALTH - NOTES.ppt for nursing students
PPTX
surgery guide for USMLE step 2-part 1.pptx
PPTX
Uterus anatomy embryology, and clinical aspects
PPTX
anaemia in PGJKKKKKKKKKKKKKKKKHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHH...
DOCX
RUHS II MBBS Microbiology Paper-II with Answer Key | 6th August 2025 (New Sch...
PDF
Oral Aspect of Metabolic Disease_20250717_192438_0000.pdf
PPTX
anal canal anatomy with illustrations...
PPTX
Important Obstetric Emergency that must be recognised
PPT
STD NOTES INTRODUCTION TO COMMUNITY HEALT STRATEGY.ppt
PPTX
ACID BASE management, base deficit correction
PPTX
Respiratory drugs, drugs acting on the respi system
PDF
شيت_عطا_0000000000000000000000000000.pdf
PDF
Intl J Gynecology Obste - 2021 - Melamed - FIGO International Federation o...
PPTX
LUNG ABSCESS - respiratory medicine - ppt
PDF
Handout_ NURS 220 Topic 10-Abnormal Pregnancy.pdf
Copy-Histopathology Practical by CMDA ESUTH CHAPTER(0) - Copy.ppt
Obstructive sleep apnea in orthodontics treatment
Stimulation Protocols for IUI | Dr. Laxmi Shrikhande
Acid Base Disorders educational power point.pptx
DENTAL CARIES FOR DENTISTRY STUDENT.pptx
MENTAL HEALTH - NOTES.ppt for nursing students
surgery guide for USMLE step 2-part 1.pptx
Uterus anatomy embryology, and clinical aspects
anaemia in PGJKKKKKKKKKKKKKKKKHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHHH...
RUHS II MBBS Microbiology Paper-II with Answer Key | 6th August 2025 (New Sch...
Oral Aspect of Metabolic Disease_20250717_192438_0000.pdf
anal canal anatomy with illustrations...
Important Obstetric Emergency that must be recognised
STD NOTES INTRODUCTION TO COMMUNITY HEALT STRATEGY.ppt
ACID BASE management, base deficit correction
Respiratory drugs, drugs acting on the respi system
شيت_عطا_0000000000000000000000000000.pdf
Intl J Gynecology Obste - 2021 - Melamed - FIGO International Federation o...
LUNG ABSCESS - respiratory medicine - ppt
Handout_ NURS 220 Topic 10-Abnormal Pregnancy.pdf

Validation of qualitative lab test methods

  • 1. Validation of Qualitative Lab Test Methods By: Dr Mostafa Mahmoud MD, Consultant Microbiologist Labs & Blood Banks Admin, Riyadh, MOH Associate Professor of Medical Microbiology & Immunology Faculty of Medicine – Ain Shams University
  • 2. Procedures for validation • Each method validation/ verification study is a collection of experiments to assess performance and error in order to judge a method’s suitability for use in the laboratory. • Acceptability Criteria • the laboratory must establish acceptance criteria. • Parameters for accuracy, precision, sensitivity and specificity should include a confidence level of at least 90%, or meet the claims of the manufacturer.
  • 3. Qualitative Methods: • Includes semi quantitative testing that use cut offs such as hepatitis testing and some molecular testing. • No values/concentrations are included in the patient report. • Test results are reported as positive/negative, normal/ borderline/abnormal, reactive/nonreactive, detected /not detected. •1. FDA cleared or approved methods (CLIA Standards). • Parameters to be validated: 1. Accuracy (correlation or comparison) 2. Precision (Replication) 3. Reportable range 4. Reference range (Normal values)
  • 4. 1- Accuracy (correlation or comparison) • Demonstrates how close to the “true” value the new method can achieve. The materials (Samples) used: 1- Calibrators/controls, 2- Reference material, 3- Proficiency testing material with known values, 4- Samples tested by another lab using the same or similar method, 5- or by comparing results to an established comparative method.
  • 5. Method & Number of samples: • Most sources recommend comparing at least 40 patient specimens. CLIA current guidance suggests a minimum of 20 samples. • A method comparison must be used for 5 days and for another 5 days if failed. • Document the results of the new method comparing the known values from the reference sources, another certified laboratory’s results or with results from the current method.
  • 6. Calculations: • Calculate the percent of positive, negative and total accuracy by dividing observed results over known results multiplied by 100. • Example: New method = 19 positives, 20 negatives. Current method or reference material with known values = 20 positive, 20 negatives • Percent positive accuracy 19/20 X 100 = 95% • Percent negative accuracy 20/20 X 100 = 100% • Total accuracy 39/40 X 100 = 98%
  • 8. 2- Precision (Reproducibility ): • Can the new method duplicate the same results? •Materials used: 1- patient samples are the first choice followed by 2- control material and reference solutions. Method & Number of samples: - 2 negative samples and 2 positive samples run in triplicate for 5 days will provide data for within-run and between-run components to estimate precision. •
  • 9. Calculation: • Calculate the percent within-run (intra), between-run (inter) and total precision by dividing observed results over known results multiplied by 100. • See next tables.
  • 10. ID Day 1 Day 2 Day 3 Pos sample Pos Pos Pos Pos Pos Pos Pos Pos Pos Pos sample Pos Pos Pos Neg Pos Pos Pos Pos Pos Neg sample Neg Neg Neg Neg Neg Neg Neg Neg Neg Neg sample Neg Neg Neg Neg Neg Neg Neg Neg Neg Within run % 12/12/x 100 = 100% 11/12 x 100 = 92% 12/12/x 100 = 100%
  • 11. ID Day 4 Day 5 Between run % Pos sample Pos Pos Pos Pos Pos Pos 15/15x100 = 100% Pos sample Pos Pos Pos Neg Pos Pos 14/15x100 =93% Neg sample Neg Neg Neg Neg Neg Pos 14/15x100 =93% Neg sample Neg Neg Neg Neg Neg Neg 15/15x100 = 100% Within run % 12/12/x 100 = 100% 11/12 x 100 = 92% Total Precision: 58/60 x 100 = 96.7%
  • 12. 3- Reportable Range: • The highest and lowest test values that can be analyzed while maintaining accuracy. • For tests without high or low values, define method criteria for a positive result. Method & Number of samples: • To verify reportable range, test at least 3-5 low and high positive samples once. • These samples can be combined with the accuracy/precision experiments. • Include both weak and strong positive samples.
  • 13. • For methods depending on a cut-off value to determine positive results, testing positive specimens near the cut-off can serve as the cut-off validation. •CAP Requirement: • For qualitative tests that use a cut-off value to distinguish positive from negative, the cut-off value is established initially, and verified every 6 months thereafter. • N.B. This requirement does not apply to FDA-cleared/approved in vitro diagnostic assays that report the qualitative result based on a predefined cut-off value. • Applied to laboratory established threshold (cut-off value)
  • 14. 4- Reference Range (Normal Values): • Provided by the manufacturer and verified by running known normal patients. • Manufacturer’s ranges or even published reference ranges in textbooks are used if the patients are similar. •Method & Number of samples: • The Reference Range can be verified by testing 20 known normal samples; if no more than 2 results fall outside the manufacturer/published range then that reference range can be considered to be verified.
  • 15. • If the laboratory cannot reference the normal values, then the reference range will need to be established. This involves a selection of at least 120 reference samples for each group or subgroup that needs to be characterized.
  • 16. 5- Sensitivity & Specificity: • CLIA does not require that these parameters to be verified.
  • 17. Validation Summary for FDA-approved qualitative tests: • Clearly state the purpose of the verification, what platform/method and the number of samples for each experiment. Any discrepant results should be investigated and explained in the Summary. Test results that show sample problems such as contamination and degradation should not be used in the assessment but still listed with an explanation. • The Summary should also contain a Conclusion stating weather the study met the acceptance criteria or not and its suitability for us in the laboratory. • Additional samples to be added if results are outside acceptance. • If the results failed then test not to be implemented.
  • 18. 2. Non-FDA Cleared tests • Qualitative methods developed in-House, non-FDA cleared methods and FDA-cleared methods modified by the laboratory. • - For Qualitative methods follow the instructions above for Accuracy, Precision, Reportable Range and Reference Range. •Qualitative Sensitivity is not done by analytic sensitivity but by diagnostic sensitivity. • Diagnostic sensitivity: The percent of subjects (patients) with the target condition whose test values are positive.
  • 19. •Method &number of samples: • dividing the number of true positives by the sum of the number of true positives plus the number of false negatives and multiplying by 100. Diagnostic Sensitivity: = [TP ÷(TP + FN)] x 100 - If a person has a disease, how often will the test be positive (true positive rate). - The higher the sensitivity the lower the False Negative and vice versa.
  • 20. •Qualitative Specificity • Qualitative sensitivity not assessed by Analytical Specificity but also by the Diagnostic Specificity. • Diagnostic Specificity: the percent of subjects (Persons) without the target condition whose test values are negative.
  • 21. •Method & number of samples: • Calculate by dividing the number of true negatives by the sum of the number of true negatives plus the number of false positives and multiplying by 100. Diagnostic Specificity: [TN ÷(TN + FP)] x 100 -If a person does not have the disease, how often will the test be negative ( true negative rate). - The higher the specificity the lower the False Positive and vice versa.
  • 22. Interference: • The laboratory must be aware of common interfering substances by referencing studies performed elsewhere (manufacturer or literature) or by performing studies. • Common blood interferences are; hemolysis, bilirubin, lipemia, preservatives and anticoagulants used in specimen collection.
  • 23. CAP Accreditation requirements for validating laboratory tests FDA approved/cleared LDTs & modified FDA tests Accuracy & Precision (COM.40300, ph II) Verify manufacturer’s results Establish (= validation) Analytic sensitivity (LOD) (COM.40400, ph II) Verify: manufacturer or literature documentation OK Establish Analytic specificity (interferences) (COM.40500, ph II) Reference literature or manufacturer documentation Establish; studies by manufacturer or in literature OK Reportable range (AMR) (COM.40600, ph II) Verify* Establish* Reference range (COM.50000, ph II) Verify ** Establish**
  • 24. Test method validation CBAHI standard LB.10 The laboratory develops a process for test method validation. LB.10.1 The laboratory implements policies and procedures on test method validation including: LB.10.1.1 Verification of accuracy/precision. LB.10.1.2 Verification of sensitivity (lower detection limit). LB.10.1.3 Verification of carryover acceptability. LB.10.1.4 Verification of the Analytic Measurement (AMR) LB.10.1.5 Approval of the method for clinical use (CRR??).
  • 25. Test Method Validation Evidence of compliance to (LB 10) standard: 1- Document Reviews: - Policies, process and procedures on method validation. 2- Documented Evidences - Records surveyor-selected method confirms compliance with policies and procedures. 3- Staff Interview - Senior personnel are knowledgeable about the concept of method validation.
  • 26. • LB.13 The laboratory has a system for instruments/methods correlation. LB.13.1 When the laboratory uses more than one method and/or instruments to test for a given analyte, the laboratory develops and implements policies and procedures on correlation to ensure the following: LB.13.1.1 The correlation studies are conducted every six months. LB.13.1.2 There is clear description of the correlation study. LB.13.1.3 There are clearly defined acceptance criteria. LB.13.1.4 There is a process for review and approval of the correlation results.
  • 27. • Evidence of Compliance (LB.13): • Document Review - Policies, process and procedures on methods/instrument correlation. • Documented Evidences - Records surveyor-selected methods/instruments confirms compliance with policies and procedures. • Staff Interview - Senior personnel are knowledgeable about the purpose of methods/instrument correlation.
  • 28. • LB.14 The laboratory has a system for controlling the quality of test methods. LB.14.1 The laboratory implements policies and procedures on quality control of test methods to satisfy the following: LB.14.1.1 Assignment of performance and review responsibility (control specimens are handled and tested in the same manner and by the same laboratory personnel testing patient samples). LB.14.1.2 Number and frequency of running controls. LB.14.1.3 Tolerance limits of controls results. LB.14.1.4 Corrective action to be taken in the event of unacceptable results. LB.14.2 The laboratory quality control system conforms to the manufacturer's instructions.
  • 29. •Evidence of compliance to LB 14 • Document review: - Policies, process and procedures on controlling the quality of test methods. - Policies and procedures on controlling the quality of test methods confirm to the manufacturer instructions. • Documented Evidences: - Records surveyor-selected methods confirms compliance with policies and procedures. • Staff Interview: - Laboratory personnel understanding of the purpose of controlling the quality of test methods.