DEVICES USED IN ICU &
MONITORING
1
Prof.Dr. RS Mehta, BPKIHS
INTRODUCTION
Intensive care unit (ICU) equipment
includes patient monitoring, respiratory
and cardiac support, pain management,
emergency resuscitation devices, and
other life support equipment .
2Prof.Dr. RS Mehta, BPKIHS
Contd…
They are designed to care for patients
who are
 seriously injured,
 have a critical or life-threatening illness,
or
 have undergone a major surgical
procedure thereby requiring 24-hour care
and monitoring.
3Prof.Dr. RS Mehta, BPKIHS
PURPOSE
 An ICU may be designed and equipped to
provide care to patients with a range of
conditions, or it may be designed and
equipped to provide specialized care to
patients with specific conditions.
4Prof.Dr. RS Mehta, BPKIHS
Contd…
 Neuromedical ICU cares for patients with
acute conditions involving the nervous
system or patients who have just had
neurosurgical procedures and require
equipment for monitoring and assessing
the brain and spinal cord.
5Prof.Dr. RS Mehta, BPKIHS
Contd…
 A neonatal ICU is designed and
equipped to care for infants who are ill,
born prematurely, or have a condition
requiring constant monitoring.
 A trauma/burn ICU provides specialized
injury and wound care for patients
involved in auto accidents and patients
who have gunshot injuries or burns.
6Prof.Dr. RS Mehta, BPKIHS
7Prof.Dr. RS Mehta, BPKIHS
TYPES OF DEVICES
Intensive care unit equipment includes
 Patient monitoring devices
 Life support and emergency resuscitation
devices, and
 Diagnostic devices.
8Prof.Dr. RS Mehta, BPKIHS
PATIENT MONITORING EQUIPMENT
 Arterial line
 Bed side monitor
 Blood pressure device (sphygmomanometer)
 Blood pressure monitor
 Electrocardiograph(ECG or EKG machine)
 Electroencephalograph(EEG machine)
 Intracranial pressure monitor
 Pulse Oximeter
 Glucometer
9Prof.Dr. RS Mehta, BPKIHS
LIFE SUPPORT AND EMERGENCY
RESUSCITATION DEVICES
 MechanicalVentilator
 Laryngoscope
 Airway
 Infusion pump
 Crash cart(Resuscitation cart)
 Intra aortic ballon pump
 Continuous positive air pressure
machine (CPAP)
 Defibrillator
10Prof.Dr. RS Mehta, BPKIHS
DIAGNOSTIC EQUIPMENT
 Mobile x-ray units
 portable clinical laboratory devices,
 Bronchoscope
 Colonoscope
 Endoscope
 Gastroscope
11Prof.Dr. RS Mehta, BPKIHS
OTHER ICU EQUIPMENT
Disposable ICU equipment includes
 Urinary catheter
 Urinary drainage collector
 Suction catheter
 Nasogastric (NG) tube
 Intravenous(IV) line or catheter
 Feeding tube
 Breathing tube( Endotracheal tube)
12Prof.Dr. RS Mehta, BPKIHS
Hemodynamic Monitoring
13Prof.Dr. RS Mehta, BPKIHS
Prof.Dr. RS Mehta, BPKIHS
Overview
 Blood pressure monitoring
◦ NIBP
◦ IBP
 Central venous pressure monitoring
 Pulmonary artery pressure
monitoring
 Mixed venous oxygen monitoring
 Cardiac output
15Prof.Dr. RS Mehta, BPKIHS
Why monitor BP?
◦ Alterations inherent
◦ Provides data for interpretation/therapeutic
decisions
◦ Important for determining organ perfusion
(MAP most important, except with the heart)
16Prof.Dr. RS Mehta, BPKIHS
Noninvasive Hemodynamic Monitoring
 Noninvasive BP
 Heart Rate, pulses
 Mental Status
 SkinTemperature
 Capillary Refill
 Urine Output
17Prof.Dr. RS Mehta, BPKIHS
Indications for
Arterial Blood Pressure
 Frequent titration of vasoactive drips
 Major surgery involving large fluid shifts
 CVP
 Aortic surgery
 Unstable blood pressures
 FrequentABGs or labs
 Unable to obtain Non-invasive BP
18Prof.Dr. RS Mehta, BPKIHS
Supplies to Gather
 Arterial Catheter
 PressureTubing
 Pressure Cable
 Sterile Gown
 SterileTowels
 Sterile Gloves
 Pressure Bag
 Flush – 500cc NS
 Suture (silk 2.0)
 Chlorhexidine Swabs
 Mask
19Prof.Dr. RS Mehta, BPKIHS
Potential Complications
Associated With Arterial Lines
 Hemorrhage
 Air Emboli
 Infection
 Altered Skin Integrity
 Impaired Circulation
20Prof.Dr. RS Mehta, BPKIHS
ARTERIAL LINE
DEFINITION:
It is the method of direct continuous
monitoring of systemic arterial pressure by
inserting a catheter into peripheral artery
either in arm or in leg. The catheter is
connected with a transducer with electrical
signals.
21Prof.Dr. RS Mehta, BPKIHS
PURPOSE:
The arterial line provides a way to
constantly measure a patient's blood
pressure and may be essential to the
stabilization of the patient.
Continuous measurement of arterial blood
pressure in case of open heart surgery.
22Prof.Dr. RS Mehta, BPKIHS
 Arterial lines may be useful in patients
with very high or low blood pressures.
The arterial line also provides access for
frequent blood sampling.
23Prof.Dr. RS Mehta, BPKIHS
COMPLICATIONS:
 The major complications associated with
the arterial line are bleeding, infection, and
rarely, a lack of blood flow to the tissue
supplied by the artery.
24Prof.Dr. RS Mehta, BPKIHS
NURSES ROLE / NURSING CARE
 Never give any medication through an arterial
line.
 Always check the pressure of the pressurized
bag and maintain a pressure of 300mm of hg.
 Cover the cannula cap with adhesive tape.
 Flush properly the arterial line every hour and
every time after a blood sample is drawn.
 Always compress the site after removal of
arterial line for 10 min.
25Prof.Dr. RS Mehta, BPKIHS
BEDSIDE MONITOR
 A bedside monitor is a
display of major body
functions on a device
that looks like a
television screen or
computer monitor.
26Prof.Dr. RS Mehta, BPKIHS
 It is a comprehensive patient monitoring
systems that can be configured to
continuously measure and display a
number of parameters via electrodes and
sensors that are connected to the patient.
27Prof.Dr. RS Mehta, BPKIHS
 These may include the electrical activity
of the heart via an EKG, respiration rate
(breathing), blood pressure, body
temperature, cardiac output, and amount
of oxygen and carbon dioxide in the
blood.
28Prof.Dr. RS Mehta, BPKIHS
 Each patient bed in an ICU has a
physiologic monitor that measure these
body activities. All monitors are
networked to a central nurses' station.
29Prof.Dr. RS Mehta, BPKIHS
PURPOSES:
 The monitor is typically used when the
doctor wants to measure functions like
the heart rate, respiratory rate, blood
pressure and temperature. In addition,
special functions such as capnography,
oximetry, electroencephalography and
pulmonary artery catheter readings are
also used in certain situations.
30Prof.Dr. RS Mehta, BPKIHS
 The bedside monitor has alarms that signal
the nurse if a body function needs
attention.
31Prof.Dr. RS Mehta, BPKIHS
NURSES ROLE:
 Check properly each connection so as to
get a desired reading.
 Any abnormality in a reading is signalled by
an alarm so inform doctor immediately.
32Prof.Dr. RS Mehta, BPKIHS
CentralVenous Line or Catheter
 A central venous catheter is a special IV
line that is inserted into a large vein in the
body. Several veins are used for central
venous catheters including those located
in the shoulder (subclavian vein), neck
(jugular vein), and groin (femoral vein)
33Prof.Dr. RS Mehta, BPKIHS
34Prof.Dr. RS Mehta, BPKIHS
Common sites for central venous
catheter insertion
1
Prof.Dr. RS Mehta, BPKIHS
PROCEDURE
 The most
common used
method is
seldinger
technique.
Prof.Dr. RS Mehta, BPKIHS
 In some patients, a central venous
catheter may be inserted into the
elbow vein (anticubital vein) and
advanced into the subclavian vein.
37Prof.Dr. RS Mehta, BPKIHS
38Prof.Dr. RS Mehta, BPKIHS
PURPOSE
 These special IVs are used when the
patient either does not have adequate
veins in the arms or needs special
medications and/or nutrition that cannot
be given through the smaller arm veins.
 Serve as a guide of fluid balance in
critically ill patients.
 Determine the function of the right side
of the heart
39Prof.Dr. RS Mehta, BPKIHS
complication
 Bleeding and infection are complications
associated with IV catheters. As previously
mentioned, collapse of a lung is a rare
complication of central venous catheters.
If this occurs, a chest tube (thoracostomy
tube) may be required to re-expand the
lung.
40Prof.Dr. RS Mehta, BPKIHS
 Arterial puncture, cardiac puncture
 Pneumothorax, Hemomothorax
 Air emboli,Thrombosis
 Cardiac temponade
 Cardiac arrhythmias
Carotid Artery Puncture
Perforation of SVC or R.Atrium/Ventricle
Pleural Effusion
41Prof.Dr. RS Mehta, BPKIHS
NURSES ROLE
 Monitor for the signs of complications.
 Assess for patency of the CVP line.
 Sterile dressing should be done to
prevent infection( CVP care per the
hospital protocol)
 The length of the indwelling catheter
should be recorded and regularly
monitored.
42Prof.Dr. RS Mehta, BPKIHS
ICP monitor
 ICU patients who have sustained head
trauma, brain hemorrhage, brain surgery,
or conditions in which the brain may
swell might require intracranial pressure
monitoring.
43Prof.Dr. RS Mehta, BPKIHS
PURPOSE
 The purpose of ICP monitoring is to
continuously measure the pressure
surrounding the brain. If the pressure
surrounding the brain gets too high, it can
cause decreased blood flow to the brain
and potentially lead to brain damage.
44Prof.Dr. RS Mehta, BPKIHS
 The ICP monitor is usually inserted by a
neurosurgeon while the patient is in the
ICU or operating room. After using
numbing medicine (local anesthetics), the
neurosurgeon makes a skin incision and
inserts the ICP monitor into the brain
through a very small hole created in the
skull.
45Prof.Dr. RS Mehta, BPKIHS
 The ICP monitor is usually inserted in the
left or right top-front part of the brain.
Some ICP monitors can drain spinal fluid if
necessary.
46Prof.Dr. RS Mehta, BPKIHS
complication
 Potential complications associated with
ICP monitoring include infection and brain
hemorrhage, which are very infrequent.
47Prof.Dr. RS Mehta, BPKIHS
Nurses role
 Optimizing cerebral tissue perfusion.
 Preventing infection.
 Maintaining patient airway.
 Maintaining negative fluid balance.
 Prevent infection( dressing)
48Prof.Dr. RS Mehta, BPKIHS
PULSE OXIMETER
A pulse oximeter is the device that
measures and displays the oxygen arterial
saturation. The study is called pulse
oxymetry.
The pulse oximeter is a small device that
has to be in contact with the skin to detect
the oxygen saturation.
49Prof.Dr. RS Mehta, BPKIHS
 The device is usually place on the
patient's finger, earlobe, toe or nose. The
pulse oximeter gives off light that
determines the oxygen saturation of the
blood.
50Prof.Dr. RS Mehta, BPKIHS
Breathing Machine (Mechanical
Ventilator
 A breathing machine
helps the patient
breathe. It is designed
to help patients who
cannot breathe
adequately on their
own. The breathing
machine does not fix
any problems of the
lungs.
51Prof.Dr. RS Mehta, BPKIHS
 It is a device that simply pushes air and
oxygen into the lungs and withdraws
carbon dioxide from the lungs. The lungs
must function in order for the breathing
machine to be effective.
52Prof.Dr. RS Mehta, BPKIHS
PURPOSE
 A breathing machine is
used whenever a
patient cannot breathe
without assistance.
Doctors, nurses and
respiratory therapists
all work to make sure
a breathing machine is
not used any longer
than necessary.
53Prof.Dr. RS Mehta, BPKIHS
 The goal when a breathing machine is
first used is to get the patient to be able
to breathe on their own, so that the
breathing machine can be removed.
54Prof.Dr. RS Mehta, BPKIHS
complications
 Patients who require breathing machine
support are at increased risk to develop
pneumonia. Occasionally, patients may
develop a collapsed lung. Both of these
complications require treatment
55Prof.Dr. RS Mehta, BPKIHS
NURSES ROLE
 Promoting effective airway clearance.
 preventing trauma and infection.
 Check
 Ventilator functioning properly
 Blockage of air passage
 Too much sputum, secretions
 When sedation drugs are used
 ABG, hypoxia
56Prof.Dr. RS Mehta, BPKIHS
b. Suction periodically as per need
c. Change the mode setup as adviced.
d. Give sedatives as adviced.
57Prof.Dr. RS Mehta, BPKIHS
INFUSION PUMP
 An intravenous (IV) infusion pump is a
machine that carefully controls the rate at
which IV fluids and/or IV medications are
given.
58Prof.Dr. RS Mehta, BPKIHS
PURPOSE
 Under some circumstances, the rate at
which IV fluids and/or IV medications are
given needs to be closely controlled.
59Prof.Dr. RS Mehta, BPKIHS
 These pumps are very reliable. Mechanical
problems are possible, but very rare. If the
IV infusion pump does not work correctly,
an alarm will sound.
60Prof.Dr. RS Mehta, BPKIHS
NURSES ROLE
 Using aseptic technique and universal
precautions, iv infusion should be set.
 Set the flow rate as prescribed calculating
the amount of fluid.
 Observe for the signs of infiltration or
other complications such as
thrombophlebitis. Fluid or electrolyte
overload and embolism before
administration.
61Prof.Dr. RS Mehta, BPKIHS
Resuscitation Cart (Crash Cart)
 The resuscitation cart
contains all of the
equipment and
medications needed
for advanced life
support and CPR
(cardiopulmonary
resuscitation).
62Prof.Dr. RS Mehta, BPKIHS
purpose
 This emergency equipment is used only if
the patient's heart or lungs stop working.
The cart is brought to the patient's
bedside when the patient's heart or lungs
are failing or have failed.
63Prof.Dr. RS Mehta, BPKIHS
NURSES ROLE
 Keep the resuscitation cart ready all the
time.
 Check the devices and ensure that the
devices are kept in charging.
 Check for the emergency (life saving)
medication for their expiry date.
64Prof.Dr. RS Mehta, BPKIHS
DEFIBRILLATOR
 A defibrillator is a device that is designed
to pass electrical current through a
patient’s heart. The passing of electrical
current through the heart is called
defibrillation. A defibrillation is done
through pads placed on the patient’s chest.
65Prof.Dr. RS Mehta, BPKIHS
purpose
 A defibrillation is used to restore a
patient’s heart rhythm to normal.
Abnormal heart rhythms may be treated
with medications while other rhythms
need to be treated with defibrillation.
66Prof.Dr. RS Mehta, BPKIHS
 Life threatening heart rhythms need
defibrillation immediately while other heart
rhythms may be defibrillated in a scheduled
fashion.
 Defibrillation may be done using the manual
defibrillator or the automatic external
defibrillator (AED).
67Prof.Dr. RS Mehta, BPKIHS
Complication
 The defibrillator pads may cause a skin
irritation and leave a temporary redden
area where they contacted the chest.
Unfortunately defibrillation does not
always return the patient’s heart rhythm
back to normal.
68Prof.Dr. RS Mehta, BPKIHS
NURSES ROLE
 Keep the patient in comfortable position
and obtain 12 lead ECG.
 Give the patient 100 % oxygen by
inhalation.
 Apply electrode paste on the DC paddle,
rub it and apply the paste at the patient’s
chest in the second intercostal space at the
right side of breast line and at the apex of
the heart.
69Prof.Dr. RS Mehta, BPKIHS
 TURN OFF the oxygen to the patient as a
spark from paddle could blow the oxygen
on the fire.
 Be sure to say “ ALL CLEAR”. No one
should touch the patient or the bed during
cardioversion.
 Check the rhythm on ECG monitor.
 Keep the patient in comfortable position
and give 100% oxygen by inhalation.
70Prof.Dr. RS Mehta, BPKIHS
 Report and record the procedure and
clean the paddle area with spirit swab.
 Keep the difibrilator on continue electrical
charging.
71Prof.Dr. RS Mehta, BPKIHS
MAINTENANCE OF ICU
EQUIPMENTS
 Since ICU equipment is used continuously
on critically ill patients, it is essential that
equipment be properly maintained,
particularly devices that are used for life
support and resuscitation.
72Prof.Dr. RS Mehta, BPKIHS
Contd…
 Staff in the ICU should perform daily
checks on equipment and inform
biomedical engineering staff when
equipment needs maintenance, repair, or
replacement.
 For mechanically complex devices, service
and preventive maintenance contracts are
available from the manufacturer or third-
party servicing companies, and should be
kept current at all times.
73Prof.Dr. RS Mehta, BPKIHS
Health care team roles
 Equipment in the ICU is used by a team
specialized in their use. The team usually
comprises a critical care attending
physician (also called an intensivist), critical
care nurses, an infectious disease team,
critical care respiratory therapists,
pharmacologists, physical therapists, and
dietitians.
74Prof.Dr. RS Mehta, BPKIHS
 Radiologic technologists perform mobile x
ray examinations (bedside radiography).
Either nurses or clinical laboratory
personnel perform point-of-care blood
analysis. Equipment in the ICU is
maintained and repaired by hospital
biomedical engineering staff and/or the
equipment manufacturer.
75Prof.Dr. RS Mehta, BPKIHS
Thank you
76Prof.Dr. RS Mehta, BPKIHS
3.CARDIAC PACING
 Cardiac Pacing is
the repetitive
delivery of very
low electrical
energies to the
heart to initiate
and maintain
cardiac rhythm.
77Prof.Dr. RS Mehta, BPKIHS
METHODS
 Percussive pacing
 Transcutaneous
 Epicardial
 Transvenous
 Permanent pacing
78Prof.Dr. RS Mehta, BPKIHS
Types of PA catheters
1. The thermo dilution catheter:
 is the one described above; using this
catheter, thermo dilution cardiac output &
other divided haemodynamic parameters
may be measured
2. Pacing:
 Some PAC’s have the capacity to provide
intra cardiac pacing
79Prof.Dr. RS Mehta, BPKIHS
3. Mixed venous oxygen saturation:
 Special fiber-optic PAC can be used to monitor
mixed venous oxygen saturation SVO2
continuously by the principle of absorption and
reflectance of light through blood
 The normal SVO2 is 75% and a 5– 10 %
increase or decrease is considered significant
 A significant decrease in SVO2 may be due to:
(a) a decrease in the cardiac output
(b) increase in metabolic rate
(c) decrease in arterial oxygen saturation.
80Prof.Dr. RS Mehta, BPKIHS
4. Ejection fraction catheter :
New-catheters with faster thermistor response times
can be used to determine the right ventricular ejection
fraction in addition to the cardiac output
5. Continuous cardiac output measurement
:
Continuous cardiac output measuring PACs contain
an integrated thermal filament at level of the RV
This filament is activated in a programmed sequence
to provide small amounts of heat, which is then
detected in the PA by a thermistor
The data by the device yields a rapidly updated, near
continuous value for cardiac output
81Prof.Dr. RS Mehta, BPKIHS
82Prof.Dr. RS Mehta, BPKIHS
PULMONARY ARTERY CATHETER
83Prof.Dr. RS Mehta, BPKIHS
Components of a Pulmonary Artery
Catheter
84Prof.Dr. RS Mehta, BPKIHS
85Prof.Dr. RS Mehta, BPKIHS
POSITION OF PATIENT
3-way tap
manometer
Fluid
Bag
Patient in supine position
Central
Venous
Access
86Prof.Dr. RS Mehta, BPKIHS
87Prof.Dr. RS Mehta, BPKIHS
DEFINITIONS
 Cardiac output = CO (mL/min): volume
of blood pumped by heart each minute
• Stroke volume = SV (mL/beat): volume of
blood pumped by heart with each
ventricular contraction
• Heart rate = HR (beats/min)
• Starlings law: the degree to which the
ventricular walls are stretched by returning
blood determines the stroke volume (more
in = more out)
88Prof.Dr. RS Mehta, BPKIHS
Regulation of Cardiodynamics
 Intrinsic: Within the heart (SV) – force of
contraction related to degree of stretch of
myocardium
 Lots of stretch = increased force production
 Extrinsic: outside the heart (NS: Autonomic or
Hormonal) – Heart rate influenced by both
sympathetic and parasympathetic (autonomic)
nervous system – Stroke volume influenced by
blood pressure
89Prof.Dr. RS Mehta, BPKIHS
90Prof.Dr. RS Mehta, BPKIHS
Definition
 Hemodynamic regulation is known as
optimization of heart rate, preload, afterlo
ad, and contractility.
 Heart rate, or heart pulse, is the speed
of the heartbeat measured by the number
of poundings of the heart per unit
of time — typically beats per
minute (bpm).
91Prof.Dr. RS Mehta, BPKIHS
Contd…
 Preload is the end diastolic pressure
that stretches the right or left ventricle of
the heart to its greatest geometric
dimensions under variable physiologic
demand
 Afterload is the tension or stress
developed in the wall of the left ventricle
during ejection.
92Prof.Dr. RS Mehta, BPKIHS
MONITORING
 Oxygenation
 Ventilation
 Circulation
 Temperature
 Monitoring patients on ventilator
 Monitoring patients with raised ICP
93Prof.Dr. RS Mehta, BPKIHS
MONITORING CIRCULATION
 Clinical parameters
◦ Pulse- Rate, Rhythm,
Volume, Character
◦ NIBP
◦ Pulse Oximeter-
Plethysomography
◦ Cyanosis
◦ Temperature
◦ Capillary Refill
◦ Urine Output
◦ Peripheral
Temperature
◦ JVP
◦ Pedal Edema
◦ Basal Rales
94Prof.Dr. RS Mehta, BPKIHS
Role of
Nurses in
ICU
95Prof.Dr. RS Mehta, BPKIHS
 ICU nurses play a vital role in the patient’s
care, including the following:
◦ Taking regular blood tests
◦ Changing the patient’s treatment in line with test
results
◦ Giving the patient the drugs and fluids that the
doctors have prescribed
◦ Recording a patient’s blood pressure, heart rate
and oxygen levels
◦ Clearing fluid and mucus from the patient’s chest
using a suction tube
◦ Turning the patient in his or her bed every few
hours to prevent sores on the skin
96Prof.Dr. RS Mehta, BPKIHS
Contd…
◦ Cleaning the patient’s teeth and moistening
the mouth with a wet sponge
◦ Washing the patient in bed
◦ Changing the sheets
◦ Changing a patient’s surgical stockings, which
help circulation when he or she is inactive
(lying still) for a long time
◦ Putting drops in the patient’s eyes to make it
easier to blink
97Prof.Dr. RS Mehta, BPKIHS
Nurses role to patient with CVP
 Position the patient in Semi Fowler
position.
 Removes clothing that could constrict the
neck or upper chest
 Provide adequate lightening to visualize
effectively the external jugular veins.
 Prevent the infection from the ports by
change dressing.
98Prof.Dr. RS Mehta, BPKIHS
Contd…
 Label the date of insertion and change.
 Observe for complication such as
pneumothorax, hemothorax, hematoma,
cardiac tamponade, air embolism and
colonization of micro-organism.
99Prof.Dr. RS Mehta, BPKIHS
THANK YOU!
100Prof.Dr. RS Mehta, BPKIHS

More Related Content

PPTX
Central venous pressure
PPT
ICU Care Bundles
PPTX
Arterial lines by Dr.Tinku Joseph
PPTX
Intubation ppt
PDF
Intra Aortic Balloon Pump by Rubina Shehzadi RN
PDF
Post anaesthesia discharge criteria and complications
PPTX
Basics of icu care
PPT
Transport of critically ill patient
Central venous pressure
ICU Care Bundles
Arterial lines by Dr.Tinku Joseph
Intubation ppt
Intra Aortic Balloon Pump by Rubina Shehzadi RN
Post anaesthesia discharge criteria and complications
Basics of icu care
Transport of critically ill patient

What's hot (20)

PPTX
Central venous pressure monitoring
PPTX
Defibrillator
PPTX
Icu psychosis
PPTX
The crash cart
PPTX
Settings Use and Maintenance of Mechanical Ventilator
PPTX
Advanced cardiac life support(acls)
PPTX
PATIENT POSITIONING AND ANAESTHESIA ppt
PPTX
Syringe Pump
PPTX
RESUSCIATION EQUIPMENTS IN INTENSISIVE CARE UNITS
PPT
Concept Of Critical Care
PPTX
Cardiac monitoring ppt
PPTX
initial assessment of critically ill patients(PRESENTED AT IMA HOUSE,CUTTACK)
PPTX
Defibrillation & Cardioversion
PPT
Care of ventilated patient
PPTX
Central venous catheterization
PPTX
MECHANICAL VENTILATOR
PPSX
PPTX
Dynamics of healing in icu
Central venous pressure monitoring
Defibrillator
Icu psychosis
The crash cart
Settings Use and Maintenance of Mechanical Ventilator
Advanced cardiac life support(acls)
PATIENT POSITIONING AND ANAESTHESIA ppt
Syringe Pump
RESUSCIATION EQUIPMENTS IN INTENSISIVE CARE UNITS
Concept Of Critical Care
Cardiac monitoring ppt
initial assessment of critically ill patients(PRESENTED AT IMA HOUSE,CUTTACK)
Defibrillation & Cardioversion
Care of ventilated patient
Central venous catheterization
MECHANICAL VENTILATOR
Dynamics of healing in icu
Ad

Similar to Monitoring & devices used in icu ccu (20)

PPSX
3. monitoring & devices used in icu ccu
PDF
1.6. critical care devices used in icu- ext
PDF
3. monitoring & devices used in icu ccu
PDF
Monitoring system in icu
PPTX
HEMODYNAMIC MONITORING for cardiovascular cases
PPTX
Hemodynamic monitoring ppt
PPTX
CRITICAL AND CARE NURSING hemodynamics.pptx
PPTX
Hemodynamic monitoring
PDF
NCM118_LECTURE_PRELIMMIDTERM_REVIEWER.pdf.pdf
PPTX
Power point presentation MUCLecture_2023_11253802.pptx
PPT
Details of the Help-Invasive-Hemodynamic.ppt
PDF
عناية مركزة م4 غير مترجم.pdf
PPTX
4. monitoring and interpreting medical investigations in icu
PPTX
NISHI%20-%20CARDIAC%20MONITORING(%20IBP,%20CVP,%20PA).pptx
PPTX
NISHI%20-%20CARDIAC%20MONITORING(%20IBP,%20CVP,%20PA).pptx
PPTX
#Monitoring equipment
PPTX
Patient monitoring system
PPTX
anesthesia ICU equipments use and maintenance
3. monitoring & devices used in icu ccu
1.6. critical care devices used in icu- ext
3. monitoring & devices used in icu ccu
Monitoring system in icu
HEMODYNAMIC MONITORING for cardiovascular cases
Hemodynamic monitoring ppt
CRITICAL AND CARE NURSING hemodynamics.pptx
Hemodynamic monitoring
NCM118_LECTURE_PRELIMMIDTERM_REVIEWER.pdf.pdf
Power point presentation MUCLecture_2023_11253802.pptx
Details of the Help-Invasive-Hemodynamic.ppt
عناية مركزة م4 غير مترجم.pdf
4. monitoring and interpreting medical investigations in icu
NISHI%20-%20CARDIAC%20MONITORING(%20IBP,%20CVP,%20PA).pptx
NISHI%20-%20CARDIAC%20MONITORING(%20IBP,%20CVP,%20PA).pptx
#Monitoring equipment
Patient monitoring system
anesthesia ICU equipments use and maintenance
Ad

More from BP KOIRALA INSTITUTE OF HELATH SCIENCS,, NEPAL (20)

PDF
3. Elderly helath care RS Mehta pdf.pdf
PDF
1. Retired and Elderly helath care RS Mehta pdf.pdf
PDF
2. Retired helath care RS Mehta pdf.pdf
PDF
Telephone Health Service to improve the quality of life
PDF
Hand Book of Oncology Nursing - Cancer Nursing Book
PDF
Text Book of Operation Theater Nursing OT Nursing
PDF
Text Book of Emergency Disaster and Trauma Nursing Book
PDF
Text Book of Critical Care Nursing ICU NURSING
PDF
Text book of Orthopedic Nursing by RS Mehta
PDF
Text book of comprehensive Medical Surgical Nursing
PPSX
M.Sc. Nursing Orientation Programme 2015.ppsx
PDF
Paradigm shift in nursing research by RS MEHTA
PDF
PDF
Ph.D. Thesis on HBC by RS Mehta.pdf
PDF
M. Sc. Nursing Thesis by RS Mehta.pdf
PDF
Ph.D. Thesis on HBC by RS Mehta.pdf
PDF
12-lead EKG Interpretation1.pdf
PDF
PDF
3. Legal Aspects in Nursing.pdf
3. Elderly helath care RS Mehta pdf.pdf
1. Retired and Elderly helath care RS Mehta pdf.pdf
2. Retired helath care RS Mehta pdf.pdf
Telephone Health Service to improve the quality of life
Hand Book of Oncology Nursing - Cancer Nursing Book
Text Book of Operation Theater Nursing OT Nursing
Text Book of Emergency Disaster and Trauma Nursing Book
Text Book of Critical Care Nursing ICU NURSING
Text book of Orthopedic Nursing by RS Mehta
Text book of comprehensive Medical Surgical Nursing
M.Sc. Nursing Orientation Programme 2015.ppsx
Paradigm shift in nursing research by RS MEHTA
Ph.D. Thesis on HBC by RS Mehta.pdf
M. Sc. Nursing Thesis by RS Mehta.pdf
Ph.D. Thesis on HBC by RS Mehta.pdf
12-lead EKG Interpretation1.pdf
3. Legal Aspects in Nursing.pdf

Recently uploaded (20)

PPTX
Fundamentals of human energy transfer .pptx
PPTX
History and examination of abdomen, & pelvis .pptx
PPT
CHAPTER FIVE. '' Association in epidemiological studies and potential errors
PPTX
Electromyography (EMG) in Physiotherapy: Principles, Procedure & Clinical App...
PPTX
LUNG ABSCESS - respiratory medicine - ppt
PPTX
CEREBROVASCULAR DISORDER.POWERPOINT PRESENTATIONx
PPT
Breast Cancer management for medicsl student.ppt
PPTX
CME 2 Acute Chest Pain preentation for education
PPTX
Slider: TOC sampling methods for cleaning validation
PPTX
Important Obstetric Emergency that must be recognised
PDF
Neuro ED Bet Sexologist in Patna Bihar India Dr. Sunil Dubey
PDF
Medical Evidence in the Criminal Justice Delivery System in.pdf
PPTX
JUVENILE NASOPHARYNGEAL ANGIOFIBROMA.pptx
PPTX
neonatal infection(7392992y282939y5.pptx
PPTX
NEET PG 2025: Memory-Based Recall Questions Compiled by Dr. Shivankan Kakkar, MD
PPTX
Patholysiology of MAFLD/MASLD and Role of GLP 1 agonist in obesity and cardio...
PPT
ASRH Presentation for students and teachers 2770633.ppt
PPTX
NEET PG 2025 Pharmacology Recall | Real Exam Questions from 3rd August with D...
PPT
Obstructive sleep apnea in orthodontics treatment
PPTX
Neuropathic pain.ppt treatment managment
Fundamentals of human energy transfer .pptx
History and examination of abdomen, & pelvis .pptx
CHAPTER FIVE. '' Association in epidemiological studies and potential errors
Electromyography (EMG) in Physiotherapy: Principles, Procedure & Clinical App...
LUNG ABSCESS - respiratory medicine - ppt
CEREBROVASCULAR DISORDER.POWERPOINT PRESENTATIONx
Breast Cancer management for medicsl student.ppt
CME 2 Acute Chest Pain preentation for education
Slider: TOC sampling methods for cleaning validation
Important Obstetric Emergency that must be recognised
Neuro ED Bet Sexologist in Patna Bihar India Dr. Sunil Dubey
Medical Evidence in the Criminal Justice Delivery System in.pdf
JUVENILE NASOPHARYNGEAL ANGIOFIBROMA.pptx
neonatal infection(7392992y282939y5.pptx
NEET PG 2025: Memory-Based Recall Questions Compiled by Dr. Shivankan Kakkar, MD
Patholysiology of MAFLD/MASLD and Role of GLP 1 agonist in obesity and cardio...
ASRH Presentation for students and teachers 2770633.ppt
NEET PG 2025 Pharmacology Recall | Real Exam Questions from 3rd August with D...
Obstructive sleep apnea in orthodontics treatment
Neuropathic pain.ppt treatment managment

Monitoring & devices used in icu ccu

  • 1. DEVICES USED IN ICU & MONITORING 1 Prof.Dr. RS Mehta, BPKIHS
  • 2. INTRODUCTION Intensive care unit (ICU) equipment includes patient monitoring, respiratory and cardiac support, pain management, emergency resuscitation devices, and other life support equipment . 2Prof.Dr. RS Mehta, BPKIHS
  • 3. Contd… They are designed to care for patients who are  seriously injured,  have a critical or life-threatening illness, or  have undergone a major surgical procedure thereby requiring 24-hour care and monitoring. 3Prof.Dr. RS Mehta, BPKIHS
  • 4. PURPOSE  An ICU may be designed and equipped to provide care to patients with a range of conditions, or it may be designed and equipped to provide specialized care to patients with specific conditions. 4Prof.Dr. RS Mehta, BPKIHS
  • 5. Contd…  Neuromedical ICU cares for patients with acute conditions involving the nervous system or patients who have just had neurosurgical procedures and require equipment for monitoring and assessing the brain and spinal cord. 5Prof.Dr. RS Mehta, BPKIHS
  • 6. Contd…  A neonatal ICU is designed and equipped to care for infants who are ill, born prematurely, or have a condition requiring constant monitoring.  A trauma/burn ICU provides specialized injury and wound care for patients involved in auto accidents and patients who have gunshot injuries or burns. 6Prof.Dr. RS Mehta, BPKIHS
  • 8. TYPES OF DEVICES Intensive care unit equipment includes  Patient monitoring devices  Life support and emergency resuscitation devices, and  Diagnostic devices. 8Prof.Dr. RS Mehta, BPKIHS
  • 9. PATIENT MONITORING EQUIPMENT  Arterial line  Bed side monitor  Blood pressure device (sphygmomanometer)  Blood pressure monitor  Electrocardiograph(ECG or EKG machine)  Electroencephalograph(EEG machine)  Intracranial pressure monitor  Pulse Oximeter  Glucometer 9Prof.Dr. RS Mehta, BPKIHS
  • 10. LIFE SUPPORT AND EMERGENCY RESUSCITATION DEVICES  MechanicalVentilator  Laryngoscope  Airway  Infusion pump  Crash cart(Resuscitation cart)  Intra aortic ballon pump  Continuous positive air pressure machine (CPAP)  Defibrillator 10Prof.Dr. RS Mehta, BPKIHS
  • 11. DIAGNOSTIC EQUIPMENT  Mobile x-ray units  portable clinical laboratory devices,  Bronchoscope  Colonoscope  Endoscope  Gastroscope 11Prof.Dr. RS Mehta, BPKIHS
  • 12. OTHER ICU EQUIPMENT Disposable ICU equipment includes  Urinary catheter  Urinary drainage collector  Suction catheter  Nasogastric (NG) tube  Intravenous(IV) line or catheter  Feeding tube  Breathing tube( Endotracheal tube) 12Prof.Dr. RS Mehta, BPKIHS
  • 15. Overview  Blood pressure monitoring ◦ NIBP ◦ IBP  Central venous pressure monitoring  Pulmonary artery pressure monitoring  Mixed venous oxygen monitoring  Cardiac output 15Prof.Dr. RS Mehta, BPKIHS
  • 16. Why monitor BP? ◦ Alterations inherent ◦ Provides data for interpretation/therapeutic decisions ◦ Important for determining organ perfusion (MAP most important, except with the heart) 16Prof.Dr. RS Mehta, BPKIHS
  • 17. Noninvasive Hemodynamic Monitoring  Noninvasive BP  Heart Rate, pulses  Mental Status  SkinTemperature  Capillary Refill  Urine Output 17Prof.Dr. RS Mehta, BPKIHS
  • 18. Indications for Arterial Blood Pressure  Frequent titration of vasoactive drips  Major surgery involving large fluid shifts  CVP  Aortic surgery  Unstable blood pressures  FrequentABGs or labs  Unable to obtain Non-invasive BP 18Prof.Dr. RS Mehta, BPKIHS
  • 19. Supplies to Gather  Arterial Catheter  PressureTubing  Pressure Cable  Sterile Gown  SterileTowels  Sterile Gloves  Pressure Bag  Flush – 500cc NS  Suture (silk 2.0)  Chlorhexidine Swabs  Mask 19Prof.Dr. RS Mehta, BPKIHS
  • 20. Potential Complications Associated With Arterial Lines  Hemorrhage  Air Emboli  Infection  Altered Skin Integrity  Impaired Circulation 20Prof.Dr. RS Mehta, BPKIHS
  • 21. ARTERIAL LINE DEFINITION: It is the method of direct continuous monitoring of systemic arterial pressure by inserting a catheter into peripheral artery either in arm or in leg. The catheter is connected with a transducer with electrical signals. 21Prof.Dr. RS Mehta, BPKIHS
  • 22. PURPOSE: The arterial line provides a way to constantly measure a patient's blood pressure and may be essential to the stabilization of the patient. Continuous measurement of arterial blood pressure in case of open heart surgery. 22Prof.Dr. RS Mehta, BPKIHS
  • 23.  Arterial lines may be useful in patients with very high or low blood pressures. The arterial line also provides access for frequent blood sampling. 23Prof.Dr. RS Mehta, BPKIHS
  • 24. COMPLICATIONS:  The major complications associated with the arterial line are bleeding, infection, and rarely, a lack of blood flow to the tissue supplied by the artery. 24Prof.Dr. RS Mehta, BPKIHS
  • 25. NURSES ROLE / NURSING CARE  Never give any medication through an arterial line.  Always check the pressure of the pressurized bag and maintain a pressure of 300mm of hg.  Cover the cannula cap with adhesive tape.  Flush properly the arterial line every hour and every time after a blood sample is drawn.  Always compress the site after removal of arterial line for 10 min. 25Prof.Dr. RS Mehta, BPKIHS
  • 26. BEDSIDE MONITOR  A bedside monitor is a display of major body functions on a device that looks like a television screen or computer monitor. 26Prof.Dr. RS Mehta, BPKIHS
  • 27.  It is a comprehensive patient monitoring systems that can be configured to continuously measure and display a number of parameters via electrodes and sensors that are connected to the patient. 27Prof.Dr. RS Mehta, BPKIHS
  • 28.  These may include the electrical activity of the heart via an EKG, respiration rate (breathing), blood pressure, body temperature, cardiac output, and amount of oxygen and carbon dioxide in the blood. 28Prof.Dr. RS Mehta, BPKIHS
  • 29.  Each patient bed in an ICU has a physiologic monitor that measure these body activities. All monitors are networked to a central nurses' station. 29Prof.Dr. RS Mehta, BPKIHS
  • 30. PURPOSES:  The monitor is typically used when the doctor wants to measure functions like the heart rate, respiratory rate, blood pressure and temperature. In addition, special functions such as capnography, oximetry, electroencephalography and pulmonary artery catheter readings are also used in certain situations. 30Prof.Dr. RS Mehta, BPKIHS
  • 31.  The bedside monitor has alarms that signal the nurse if a body function needs attention. 31Prof.Dr. RS Mehta, BPKIHS
  • 32. NURSES ROLE:  Check properly each connection so as to get a desired reading.  Any abnormality in a reading is signalled by an alarm so inform doctor immediately. 32Prof.Dr. RS Mehta, BPKIHS
  • 33. CentralVenous Line or Catheter  A central venous catheter is a special IV line that is inserted into a large vein in the body. Several veins are used for central venous catheters including those located in the shoulder (subclavian vein), neck (jugular vein), and groin (femoral vein) 33Prof.Dr. RS Mehta, BPKIHS
  • 35. Common sites for central venous catheter insertion 1 Prof.Dr. RS Mehta, BPKIHS
  • 36. PROCEDURE  The most common used method is seldinger technique. Prof.Dr. RS Mehta, BPKIHS
  • 37.  In some patients, a central venous catheter may be inserted into the elbow vein (anticubital vein) and advanced into the subclavian vein. 37Prof.Dr. RS Mehta, BPKIHS
  • 39. PURPOSE  These special IVs are used when the patient either does not have adequate veins in the arms or needs special medications and/or nutrition that cannot be given through the smaller arm veins.  Serve as a guide of fluid balance in critically ill patients.  Determine the function of the right side of the heart 39Prof.Dr. RS Mehta, BPKIHS
  • 40. complication  Bleeding and infection are complications associated with IV catheters. As previously mentioned, collapse of a lung is a rare complication of central venous catheters. If this occurs, a chest tube (thoracostomy tube) may be required to re-expand the lung. 40Prof.Dr. RS Mehta, BPKIHS
  • 41.  Arterial puncture, cardiac puncture  Pneumothorax, Hemomothorax  Air emboli,Thrombosis  Cardiac temponade  Cardiac arrhythmias Carotid Artery Puncture Perforation of SVC or R.Atrium/Ventricle Pleural Effusion 41Prof.Dr. RS Mehta, BPKIHS
  • 42. NURSES ROLE  Monitor for the signs of complications.  Assess for patency of the CVP line.  Sterile dressing should be done to prevent infection( CVP care per the hospital protocol)  The length of the indwelling catheter should be recorded and regularly monitored. 42Prof.Dr. RS Mehta, BPKIHS
  • 43. ICP monitor  ICU patients who have sustained head trauma, brain hemorrhage, brain surgery, or conditions in which the brain may swell might require intracranial pressure monitoring. 43Prof.Dr. RS Mehta, BPKIHS
  • 44. PURPOSE  The purpose of ICP monitoring is to continuously measure the pressure surrounding the brain. If the pressure surrounding the brain gets too high, it can cause decreased blood flow to the brain and potentially lead to brain damage. 44Prof.Dr. RS Mehta, BPKIHS
  • 45.  The ICP monitor is usually inserted by a neurosurgeon while the patient is in the ICU or operating room. After using numbing medicine (local anesthetics), the neurosurgeon makes a skin incision and inserts the ICP monitor into the brain through a very small hole created in the skull. 45Prof.Dr. RS Mehta, BPKIHS
  • 46.  The ICP monitor is usually inserted in the left or right top-front part of the brain. Some ICP monitors can drain spinal fluid if necessary. 46Prof.Dr. RS Mehta, BPKIHS
  • 47. complication  Potential complications associated with ICP monitoring include infection and brain hemorrhage, which are very infrequent. 47Prof.Dr. RS Mehta, BPKIHS
  • 48. Nurses role  Optimizing cerebral tissue perfusion.  Preventing infection.  Maintaining patient airway.  Maintaining negative fluid balance.  Prevent infection( dressing) 48Prof.Dr. RS Mehta, BPKIHS
  • 49. PULSE OXIMETER A pulse oximeter is the device that measures and displays the oxygen arterial saturation. The study is called pulse oxymetry. The pulse oximeter is a small device that has to be in contact with the skin to detect the oxygen saturation. 49Prof.Dr. RS Mehta, BPKIHS
  • 50.  The device is usually place on the patient's finger, earlobe, toe or nose. The pulse oximeter gives off light that determines the oxygen saturation of the blood. 50Prof.Dr. RS Mehta, BPKIHS
  • 51. Breathing Machine (Mechanical Ventilator  A breathing machine helps the patient breathe. It is designed to help patients who cannot breathe adequately on their own. The breathing machine does not fix any problems of the lungs. 51Prof.Dr. RS Mehta, BPKIHS
  • 52.  It is a device that simply pushes air and oxygen into the lungs and withdraws carbon dioxide from the lungs. The lungs must function in order for the breathing machine to be effective. 52Prof.Dr. RS Mehta, BPKIHS
  • 53. PURPOSE  A breathing machine is used whenever a patient cannot breathe without assistance. Doctors, nurses and respiratory therapists all work to make sure a breathing machine is not used any longer than necessary. 53Prof.Dr. RS Mehta, BPKIHS
  • 54.  The goal when a breathing machine is first used is to get the patient to be able to breathe on their own, so that the breathing machine can be removed. 54Prof.Dr. RS Mehta, BPKIHS
  • 55. complications  Patients who require breathing machine support are at increased risk to develop pneumonia. Occasionally, patients may develop a collapsed lung. Both of these complications require treatment 55Prof.Dr. RS Mehta, BPKIHS
  • 56. NURSES ROLE  Promoting effective airway clearance.  preventing trauma and infection.  Check  Ventilator functioning properly  Blockage of air passage  Too much sputum, secretions  When sedation drugs are used  ABG, hypoxia 56Prof.Dr. RS Mehta, BPKIHS
  • 57. b. Suction periodically as per need c. Change the mode setup as adviced. d. Give sedatives as adviced. 57Prof.Dr. RS Mehta, BPKIHS
  • 58. INFUSION PUMP  An intravenous (IV) infusion pump is a machine that carefully controls the rate at which IV fluids and/or IV medications are given. 58Prof.Dr. RS Mehta, BPKIHS
  • 59. PURPOSE  Under some circumstances, the rate at which IV fluids and/or IV medications are given needs to be closely controlled. 59Prof.Dr. RS Mehta, BPKIHS
  • 60.  These pumps are very reliable. Mechanical problems are possible, but very rare. If the IV infusion pump does not work correctly, an alarm will sound. 60Prof.Dr. RS Mehta, BPKIHS
  • 61. NURSES ROLE  Using aseptic technique and universal precautions, iv infusion should be set.  Set the flow rate as prescribed calculating the amount of fluid.  Observe for the signs of infiltration or other complications such as thrombophlebitis. Fluid or electrolyte overload and embolism before administration. 61Prof.Dr. RS Mehta, BPKIHS
  • 62. Resuscitation Cart (Crash Cart)  The resuscitation cart contains all of the equipment and medications needed for advanced life support and CPR (cardiopulmonary resuscitation). 62Prof.Dr. RS Mehta, BPKIHS
  • 63. purpose  This emergency equipment is used only if the patient's heart or lungs stop working. The cart is brought to the patient's bedside when the patient's heart or lungs are failing or have failed. 63Prof.Dr. RS Mehta, BPKIHS
  • 64. NURSES ROLE  Keep the resuscitation cart ready all the time.  Check the devices and ensure that the devices are kept in charging.  Check for the emergency (life saving) medication for their expiry date. 64Prof.Dr. RS Mehta, BPKIHS
  • 65. DEFIBRILLATOR  A defibrillator is a device that is designed to pass electrical current through a patient’s heart. The passing of electrical current through the heart is called defibrillation. A defibrillation is done through pads placed on the patient’s chest. 65Prof.Dr. RS Mehta, BPKIHS
  • 66. purpose  A defibrillation is used to restore a patient’s heart rhythm to normal. Abnormal heart rhythms may be treated with medications while other rhythms need to be treated with defibrillation. 66Prof.Dr. RS Mehta, BPKIHS
  • 67.  Life threatening heart rhythms need defibrillation immediately while other heart rhythms may be defibrillated in a scheduled fashion.  Defibrillation may be done using the manual defibrillator or the automatic external defibrillator (AED). 67Prof.Dr. RS Mehta, BPKIHS
  • 68. Complication  The defibrillator pads may cause a skin irritation and leave a temporary redden area where they contacted the chest. Unfortunately defibrillation does not always return the patient’s heart rhythm back to normal. 68Prof.Dr. RS Mehta, BPKIHS
  • 69. NURSES ROLE  Keep the patient in comfortable position and obtain 12 lead ECG.  Give the patient 100 % oxygen by inhalation.  Apply electrode paste on the DC paddle, rub it and apply the paste at the patient’s chest in the second intercostal space at the right side of breast line and at the apex of the heart. 69Prof.Dr. RS Mehta, BPKIHS
  • 70.  TURN OFF the oxygen to the patient as a spark from paddle could blow the oxygen on the fire.  Be sure to say “ ALL CLEAR”. No one should touch the patient or the bed during cardioversion.  Check the rhythm on ECG monitor.  Keep the patient in comfortable position and give 100% oxygen by inhalation. 70Prof.Dr. RS Mehta, BPKIHS
  • 71.  Report and record the procedure and clean the paddle area with spirit swab.  Keep the difibrilator on continue electrical charging. 71Prof.Dr. RS Mehta, BPKIHS
  • 72. MAINTENANCE OF ICU EQUIPMENTS  Since ICU equipment is used continuously on critically ill patients, it is essential that equipment be properly maintained, particularly devices that are used for life support and resuscitation. 72Prof.Dr. RS Mehta, BPKIHS
  • 73. Contd…  Staff in the ICU should perform daily checks on equipment and inform biomedical engineering staff when equipment needs maintenance, repair, or replacement.  For mechanically complex devices, service and preventive maintenance contracts are available from the manufacturer or third- party servicing companies, and should be kept current at all times. 73Prof.Dr. RS Mehta, BPKIHS
  • 74. Health care team roles  Equipment in the ICU is used by a team specialized in their use. The team usually comprises a critical care attending physician (also called an intensivist), critical care nurses, an infectious disease team, critical care respiratory therapists, pharmacologists, physical therapists, and dietitians. 74Prof.Dr. RS Mehta, BPKIHS
  • 75.  Radiologic technologists perform mobile x ray examinations (bedside radiography). Either nurses or clinical laboratory personnel perform point-of-care blood analysis. Equipment in the ICU is maintained and repaired by hospital biomedical engineering staff and/or the equipment manufacturer. 75Prof.Dr. RS Mehta, BPKIHS
  • 76. Thank you 76Prof.Dr. RS Mehta, BPKIHS
  • 77. 3.CARDIAC PACING  Cardiac Pacing is the repetitive delivery of very low electrical energies to the heart to initiate and maintain cardiac rhythm. 77Prof.Dr. RS Mehta, BPKIHS
  • 78. METHODS  Percussive pacing  Transcutaneous  Epicardial  Transvenous  Permanent pacing 78Prof.Dr. RS Mehta, BPKIHS
  • 79. Types of PA catheters 1. The thermo dilution catheter:  is the one described above; using this catheter, thermo dilution cardiac output & other divided haemodynamic parameters may be measured 2. Pacing:  Some PAC’s have the capacity to provide intra cardiac pacing 79Prof.Dr. RS Mehta, BPKIHS
  • 80. 3. Mixed venous oxygen saturation:  Special fiber-optic PAC can be used to monitor mixed venous oxygen saturation SVO2 continuously by the principle of absorption and reflectance of light through blood  The normal SVO2 is 75% and a 5– 10 % increase or decrease is considered significant  A significant decrease in SVO2 may be due to: (a) a decrease in the cardiac output (b) increase in metabolic rate (c) decrease in arterial oxygen saturation. 80Prof.Dr. RS Mehta, BPKIHS
  • 81. 4. Ejection fraction catheter : New-catheters with faster thermistor response times can be used to determine the right ventricular ejection fraction in addition to the cardiac output 5. Continuous cardiac output measurement : Continuous cardiac output measuring PACs contain an integrated thermal filament at level of the RV This filament is activated in a programmed sequence to provide small amounts of heat, which is then detected in the PA by a thermistor The data by the device yields a rapidly updated, near continuous value for cardiac output 81Prof.Dr. RS Mehta, BPKIHS
  • 84. Components of a Pulmonary Artery Catheter 84Prof.Dr. RS Mehta, BPKIHS
  • 86. POSITION OF PATIENT 3-way tap manometer Fluid Bag Patient in supine position Central Venous Access 86Prof.Dr. RS Mehta, BPKIHS
  • 88. DEFINITIONS  Cardiac output = CO (mL/min): volume of blood pumped by heart each minute • Stroke volume = SV (mL/beat): volume of blood pumped by heart with each ventricular contraction • Heart rate = HR (beats/min) • Starlings law: the degree to which the ventricular walls are stretched by returning blood determines the stroke volume (more in = more out) 88Prof.Dr. RS Mehta, BPKIHS
  • 89. Regulation of Cardiodynamics  Intrinsic: Within the heart (SV) – force of contraction related to degree of stretch of myocardium  Lots of stretch = increased force production  Extrinsic: outside the heart (NS: Autonomic or Hormonal) – Heart rate influenced by both sympathetic and parasympathetic (autonomic) nervous system – Stroke volume influenced by blood pressure 89Prof.Dr. RS Mehta, BPKIHS
  • 91. Definition  Hemodynamic regulation is known as optimization of heart rate, preload, afterlo ad, and contractility.  Heart rate, or heart pulse, is the speed of the heartbeat measured by the number of poundings of the heart per unit of time — typically beats per minute (bpm). 91Prof.Dr. RS Mehta, BPKIHS
  • 92. Contd…  Preload is the end diastolic pressure that stretches the right or left ventricle of the heart to its greatest geometric dimensions under variable physiologic demand  Afterload is the tension or stress developed in the wall of the left ventricle during ejection. 92Prof.Dr. RS Mehta, BPKIHS
  • 93. MONITORING  Oxygenation  Ventilation  Circulation  Temperature  Monitoring patients on ventilator  Monitoring patients with raised ICP 93Prof.Dr. RS Mehta, BPKIHS
  • 94. MONITORING CIRCULATION  Clinical parameters ◦ Pulse- Rate, Rhythm, Volume, Character ◦ NIBP ◦ Pulse Oximeter- Plethysomography ◦ Cyanosis ◦ Temperature ◦ Capillary Refill ◦ Urine Output ◦ Peripheral Temperature ◦ JVP ◦ Pedal Edema ◦ Basal Rales 94Prof.Dr. RS Mehta, BPKIHS
  • 96.  ICU nurses play a vital role in the patient’s care, including the following: ◦ Taking regular blood tests ◦ Changing the patient’s treatment in line with test results ◦ Giving the patient the drugs and fluids that the doctors have prescribed ◦ Recording a patient’s blood pressure, heart rate and oxygen levels ◦ Clearing fluid and mucus from the patient’s chest using a suction tube ◦ Turning the patient in his or her bed every few hours to prevent sores on the skin 96Prof.Dr. RS Mehta, BPKIHS
  • 97. Contd… ◦ Cleaning the patient’s teeth and moistening the mouth with a wet sponge ◦ Washing the patient in bed ◦ Changing the sheets ◦ Changing a patient’s surgical stockings, which help circulation when he or she is inactive (lying still) for a long time ◦ Putting drops in the patient’s eyes to make it easier to blink 97Prof.Dr. RS Mehta, BPKIHS
  • 98. Nurses role to patient with CVP  Position the patient in Semi Fowler position.  Removes clothing that could constrict the neck or upper chest  Provide adequate lightening to visualize effectively the external jugular veins.  Prevent the infection from the ports by change dressing. 98Prof.Dr. RS Mehta, BPKIHS
  • 99. Contd…  Label the date of insertion and change.  Observe for complication such as pneumothorax, hemothorax, hematoma, cardiac tamponade, air embolism and colonization of micro-organism. 99Prof.Dr. RS Mehta, BPKIHS
  • 100. THANK YOU! 100Prof.Dr. RS Mehta, BPKIHS