Detection and treatment of ventricular tachyarrhythmias by an ICD involves a series of
sequential steps, each of which provides an opportunity to minimize unnecessary shocks
7
Optimal ICD programming
 Programming of the detection rate
 Detection duration
 Antitachycardia pacing (ATP)
 Discriminate supraventricular tachycardia (SVT) from VT
 Programming to minimize the sensing of noise
8
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Rate and Duration for Initial Detection
 ICDs detect VT/VF if the RR intervals are shorter than the detection interval for a
programmable number of intervals or duration
 An increased detection rate or prolonged duration is the most robust programming tool for
prevention of unnecessary ICD therapy
9
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
10
Rate and Duration for Initial Detection
Primary prevention patients experience faster VTs with rates less likely to overlap SVT than secondary prevention
patients Permits programming of faster VT rate cutoffs, ,minimizes unnecessary shocks in primary prevention patients
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
11
12
Rate and Duration for Initial Detection
PREPARE (Primary Prevention Parameters Evaluation) study
The Detection rates 182 bpm, duration set for 30 of 40 intervals (≈9 seconds), patients less likely to receive a shock
(9% versus 17%, P<0. .01), No difference in arrhythmic syncope between groups
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
SVT-VT Discrimination
 Withhold inappropriate therapy
 Hardware dependent
13
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Single chamber discrimination
 Sudden onset
 Rate stability
 Ventricular complexes morphology
14
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Single-Chamber Algorithms
Sudden onset
 Sudden onset criterion bases action on the past
8 intervals:
 Takes the interval average of every other
detected ventricular event (4 out of 8)
 Compares the current interval average to each
of the four previous interval averages
 If the sudden onset difference is less than the
programmed delta, the onset is classified as
gradual, favoring sinus tachycardia. Otherwise,
the onset is classified as sudden, favoring VT.
15
Single-Chamber Algorithms
Stability
 The stability delta is the widest difference
between the second longest and second
shortest cycles, among 12 (programmable
value) consecutive cycles of tachycardia
detection (discards the fastest and the slowest
intervals and subtracts the shortest interval
from the longest interval).
If the stability difference is less than the
programmed delta, the rhythm is VT.
Otherwise favors AF
Does not, discriminate sinus tachycardia,
atrial tachycardia or atrial flutter from VT.
In this example: stability ignores longest
(395 ms) and shortest (268 ms) intervals
and subtracts second longest (378 ms)
from second shortest (271 ms): 378 ms -
271 ms = 107 m (> 80 ms the
programmed value); therefore interval
stability indicates SVT (AF)
16
17
Compare electrograms during tachycardia with a template acquired during normal rhythm
Single-Chamber Algorithms
Morphology
Single-Chamber Algorithms
Morphology
 Tachycardia morphology similar
from template by programmable
threshold is SVT
Example: SVT = Match (% match above 60% threshold)
The algorithm breaks the waveform apart into triangular chunks and compares the A
pieces, the B pieces, etc. At each point, it determines how much discrepancy there is between
template and test. In this case, there is none, which results in a 100% match. In such a
scenario, the algorithm would determine the rhythm was an SVT (sinus in origin) rather
than a VT.
18
Single-Chamber Algorithms
Morphology
 Tachycardia morphology differs
from template by programmable
threshold is VT
Example: VT = Non-Match (35% match is less than 60% threshold)
Each part is compared (A to A, B to B) etc. This drawing shows how these particular pieces do not
match exactly; the far right shows areas that one waveform has and the other does not (in other words,
the blue areas is the one that does not match). The MD algorithm assesses this numerically; in this case,
the waveforms are only a 35% match (in favor of VT).
19
Single-Chamber Algorithms
 Morphology algorithm greater sensitivity and specificity
 Sensitivity 92% to 99% and specificities of 90% to 97%
 Interval algorithms (onset and stability) unreliable at higher rates (>180–200 bpm
 Morphology is the only SVT-VT discriminator reliably applied at rates >200 bpm
20
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Dual-Chamber SVT-VT Discriminators
21
 Use information collected simultaneously from the atria and ventricles
 Dual chamber devices discriminators, depending on the AV relationship used to classify the
arrhythmia (V>A, V<A, V=A).
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Dual-Chamber SVT-VT Discriminators
The ventricular rate is faster than the atrial one in the majority (more than 80%) of the ventricular tachycardias or
fibrillations, this proportion even increases with very fast rhythms. Therefore, the 2 other branches (V=A and V<A) apply
for less than 20% of ventricular tachycardias
VT arm (V>A), When the ventricular rate is faster than the atrial rate, the tachycardia is classified as
VT. No other discrimination criterion is analyzed
22
Dual-Chamber SVT-VT Discriminators
 When the ventricular rate is slower than
the atrial rate, one must distinguish AF,
atrial tachycardia or flutter, and VT,
 Rate stability
 Stability of AV association
 QRS morphology
The rate stability and morphology criteria operate as described for single chamber devices. AV association operates on the same
principle as rate stability, i.e. the AV association delta measures the difference between the second longest and second shortest AV
intervals among the 12 tachycardia detection cycles. When Interval Stability classifies a rhythm disorder as VT, AV association goes into
effect.
23
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Dual-Chamber SVT-VT Discriminators
24
 Ventricular and the atrial rates are same, using the following criteria:
 AV interval delta
 Sudden onset
 QRS morphology
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Dual-Chamber SVT-VT Discriminators
25
 AV Interval discrimination is evaluated first.
 If AV interval points to VT, then no other discrimination is applied, and VT therapy begins.
 If AV interval points to SVT, then the rhythm is evaluated by 2 other discriminators in the
V=A rate branch: the sudden onset and morphology criteria
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Dual-Chamber SVT-VT Discriminators
26
Dual-Chamber SVT-VT Discriminators
 Use atrioventricular (AV) association to distinguish SVT from VT
 Proper atrial and ventricular sensing is essential
 Atrial undersensing caused by AF, lead dislodgement, functional, resulting from
the atrial blanking that occurs during and immediately after ventricular sensed or
paced events
 Atrial undersensing lead to misclassification of SVT as VT
27
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Dual-Chamber SVT-VT Discriminators
 Oversensing by large FFRWs , misclassification of VT as SVT, inappropriately withheld
therapy
 Prevent both far-field R-wave (FFRW) oversensing and the undersensing of small fibrillation
waves on the atrial channel
28
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Single- Versus Dual-Chamber ICD
 No benefit of dual-chamber over single-chamber
 Dual-chamber discrimination is superior when overlap in VT and SVT rates
 Dual-chamber ICDs for secondary prevention and for patients with slow VT
 In primary prevention patients, little overlap between SVT and VT heart rates, the
benefit of using high rate cutoffs and prolonged detection appears to overwhelm
any modest additional contribution of dual-chamber SVT-VT discrimination
29
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Antitachycardia Pacing
 ATP is rapid pacing at a cycle length
shorter than VT
 Terminates VT by penetrating the circuit
 ATP terminate slow VT (<188–200
bpm),
 Success 85% to 90% and a 1% to 5%
risk of acceleration
 VT in primary prevention are
monomorphic, potential application of
ATP
30
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
ATP
31
 Rate of ATP programmed at 69% to 88% of VT cycle length
 Each drive train typically has 8 stimuli
 First ATP most effective, 95% of responsive fast VTs terminated successfully
 Only a small minority of patients are responsive to >3 or 4 ATPs
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Factors affecting efficacy of ATP
 Burst ATP-Interstimulus interval in the
train remains constant
 Ramp ATP-Each subsequent stimulus in
a train is decremented
 Burst and ramp pacing have similar
efficacy and safety for slow VT
 Fast VT, burst pacing
32
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Redetection and Reconfirmation
 Contemporary devices reconfirm the
persistence of arrhythmia during and
after charging.
 Charge is dissipated if arrhythmia
terminates
 Noncommitted shock if possible
 The ADVANCE-CRT trial reported
greater efficacy of biventricular ATP in
patients with ischemic cardiomyopathy.
33
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Antitachycardia Pacing
Analysis of the response to ATP
 Used to optimize programming
 Provides measure of temporal proximity of the pacing site to the VT
 Differential diagnosis of SVT and VT
 Transient VA and AV block during ATP is diagnostic of VT and SVT, respectively
 An AAV response is diagnostic of atrial tachycardia, and a VVA response is diagnostic of VT
34
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Programming Therapy Zones
 ICDs have 2 to 3 zones defined by the longest RR interval in each zone
 The lower boundary between sinus and VT zone is programmed at approximately
180 to 188 bpm in primary prevention patients and at 30 to 60 ms greater than the
cycle length of the slowest observed VT (150–160 bpm in most trials) in secondary
prevention patients
 Programming in the VT zone is focused on preventing therapy for SVT and
nonsustained VT
35
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Programming Therapy Zones
 If 2 VT zones are programmed, the rate
at which fewer trials of ATP therapy are
desired forms the boundary between the
2 zones.
 In secondary prevention, slower VT zone
is programmed with 3 to 4 sequences of
ramp/burst ATP compared with 1-burst
ATP followed by shocks in the faster VT
zone.
 SVT discriminators are programmed on
in VT zones. Implantable cardioverter defibrillator (ICD) rate detection zones. Some
ICDs permit programming of an additional monitor-only zone. The upper
panel shows programming for secondary prevention patients. The lower
panel shows programming for primary prevention patients.
36
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Programming Therapy Zones
 The boundary between the VT and VF zones is determined by the rate at which
ATP before charge is no longer desired
 SVT discriminators are either not programmable in the VF zone (Boston Scientific
and St Jude Medical) or are applied with some limitations (Medtronic)
37
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
ATP should not be routinely programmed
38
 Inherited channelopathies (Brugada syndrome, Long and Short QT
syndrome)
 Catecholaminergic polymorphic VT
 Early repolarization syndromes
 Arrhythmia is polymorphic VT or VF
 Arrhythmias lack reentry and not interrupted by pacing
Shock Strength, Polarity, and Defibrillation Threshold
39
 Implant testing assesses VF sensing and defibrillation effectiveness
 SAFE-ICD study (Safety of Two Strategies of ICD Management at Implantation) revealed
comparable rates of ICD complications and long-term sudden death in patients who did and
did not undergo routine DFT testing
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Shock Strength, Polarity, and Defibrillation Threshold
 Strongly considered in nonstandard lead
positioning, right-sided devices,
amiodarone therapy, pediatric
implantations, and use of an s-ICD
40
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Shock Strength, Polarity, and Defibrillation Threshold
 If the DFT is known, the first shock in the VT zone is programmed ≥10 J above it,
and subsequent shocks are set to maximal energy
 The advantage of lower-energy shocks (conservation of battery life, diminished
postshock stunning or block) is minimal with current biphasic technology.
 High-energy shocks are more likely to be successful for VT/VF and SVT without
patient discomfort or increase in charge time
 Low-energy shocks also convey a risk of conversion of VT to VF
 Maximal shock strength is generally preferred
41
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Shock Strength, Polarity, and Defibrillation Threshold
 Right ventricular coil as the anode resulted in slightly lower DFT in some patients
 Reverse-shock polarity, can be tried in patients with high DFTs.
42
Optimization of Sensing to Prevent Shocks
43
 ICD should reliably sense every QRS complex during normal rhythm and small fibrillatory
waves during VF
 Do not sense T waves
 Ignore extracardiac and nonphysiological signals
 ICDs use dynamic sensitivity or gain to achieve these sensing requirements
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Optimization of Sensing to Prevent Shocks
 After a sensed or paced ventricular event,
sensitivity is reduced as a function of the
amplitude of the R wave after the
postventricular blanking period expires.
 Sensitivity is increased until a subsequent
event is sensed or the maximum
programmed sensitivity is reached
 Strategy reduces risk of T-wave
oversensing while maintaining sensitivity
for the small deflections of VF
44
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Preventing T-Wave Oversensing
45
 Low-amplitude R waves (<3 mV), large T waves, or long QT intervals promote T-wave
oversensing, result in double counting and inappropriate therapy.
 Hypertrophic cardiomyopathy, long-QT syndrome, and Brugada syndrome predispose T-
wave oversensing
 Decreasing ventricular sensitivity reduces the risk of T-wave oversensing but requires a
sufficiently large R- to T-wave ratio and may risk VF undersensing
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Surveillance of Lead Fracture
 Lead Integrity Alert algorithm for detection
of lead fracture based on sudden changes in
lead impedance and noise detection
 The lead-integrity alert is triggered if 2 of 3
events occur:
 (1) Sudden increase in lead impedance
 (2) ≥2 nonsustained tachycardia events (≥5
beats) with intervals <220 ms
 (3) sensing integrity counters with ≥30 short
RR-interval counts <140 ms within 3
consecutive days indicative of
nonphysiological signals
46
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Noise-Detection Algorithms
 Algorithms for discrimination of VT from myopotentials and nonphysiological noise such as
electromagnetic interference
 Noise have higher frequency with very short intervals
 If noise is detected, the device raises the floor of the dynamic sensitivity above the level of
the noise to prevent oversensing
 This is most useful to prevent myopotential detection
47
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
ICD Programming During Electrical Storm
 Recurrent ICD shocks in the course of
electrical storm (defined as ≥3 VT/VF
episodes in 24 hours)
 Lead to heart failure, sympathetic
overdrive, and psychological trauma,
which potentiates the risk for recurrent
VT.
48
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
ICD Programming During Electrical Storm
49
 Programming for minimizing the number of shocks
 Duration to detection is increased
 Number of intervals to declare an end of episode decreased
 Increasing the lower pacing rate may help suppress recurrent VT/VF
 Hospitalized patients, ICD therapies are turned off during electrical storm
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Remote Monitoring
 Automated, wireless remote
monitoring with clinician and
patient alerts shortens the time to
detection of changes in clinical
status and device function and may
reduce mortality risk
50
Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
Conclusions
 Proper ICD programming minimizes unnecessary ICD therapy to reduce patient
morbidity and mortality.
 Current evidence supports the use of various strategies to reduce shocks without
compromising therapy effectiveness.
 The extension of detection duration to prevent treatment of self-terminating
tachycardia has proven to be the most useful of these strategies, followed by the
application of SVT-VT discrimination and the use of ATP to terminate VT
51
Disclosures
 Dr Friedman has received research support from Medtronic (grant administered by
Mayo Clinic for investigator-initiated study), Biotronik, and Cameron Health, and
is a speaker or consultant for Bard, Biotronik, Leadex, and Sorin.
 Dr Madhavan reports no conflicts
52
53
Thank You

More Related Content

PPTX
Basics of Electrophysiologic study, part 1 (2020)
PPTX
Tachycardia discriminating algorithms and trouble shooting of ICDs
PPTX
Ep diagnosis and ablation of avnrt
PPT
Pacemaker timing &amp; advanced dual chamber concepts
PPTX
Echo assesment of rv function
PPTX
Coronary lesion assessment
PPTX
LV Dyssynchrony assessment
PPTX
FFR(fractional flow reserve)
Basics of Electrophysiologic study, part 1 (2020)
Tachycardia discriminating algorithms and trouble shooting of ICDs
Ep diagnosis and ablation of avnrt
Pacemaker timing &amp; advanced dual chamber concepts
Echo assesment of rv function
Coronary lesion assessment
LV Dyssynchrony assessment
FFR(fractional flow reserve)

What's hot (20)

PPTX
CRT Non Responders - A practical guide
PPT
Shunt quantification
PPTX
Lesion complexity
PPTX
Asd device closure
PPTX
Electrophysiology AVNRT
PPTX
POST CRT OPTIMISATION
PPT
Evaluation of prosthetic valve function and clinical utility.
PPTX
Trouble shoooting ICD AND CRT
PPTX
His bundle pacing
PPTX
Cardiac dyssynchrony ppt by dr awadhesh
PPTX
Contrast Echocardiography
PPTX
Echocardiography of CHD in Adults
PPTX
His bundle pacing by Dr. Vaibhav Yawalkar MD DM Cardiology, Consultant Interv...
PPTX
LEADLESS PACEMAKER AND SUBCUTANEOUS ICD
PPTX
CORONARY BALLOONS PRACTICAL ASPECTS.pptx
PPTX
ECG interpretation postCRT
PDF
Contrast echocardiography
PPTX
Right heart catheters
PPTX
Percutaneous Pulmonary Valve Interventions
PPTX
Coronary guide wires
CRT Non Responders - A practical guide
Shunt quantification
Lesion complexity
Asd device closure
Electrophysiology AVNRT
POST CRT OPTIMISATION
Evaluation of prosthetic valve function and clinical utility.
Trouble shoooting ICD AND CRT
His bundle pacing
Cardiac dyssynchrony ppt by dr awadhesh
Contrast Echocardiography
Echocardiography of CHD in Adults
His bundle pacing by Dr. Vaibhav Yawalkar MD DM Cardiology, Consultant Interv...
LEADLESS PACEMAKER AND SUBCUTANEOUS ICD
CORONARY BALLOONS PRACTICAL ASPECTS.pptx
ECG interpretation postCRT
Contrast echocardiography
Right heart catheters
Percutaneous Pulmonary Valve Interventions
Coronary guide wires
Ad

Similar to AICD programming (20)

PPTX
ICD troubleshooting
PDF
Real-Time Detection of Fatal Ventricular Dysrhythmias for Automated External ...
PPTX
Implantable cardioverter defibrillator ICD.pptx
PPTX
Rol actual del cardiodesfibrilador implantable subcutáneo en la prevención de...
PPTX
How to read ICD episode_Dr Kush Bhagat.pptx
PPTX
Basics of ICD - Dr Hafeesh Fazulu - Pushpagiri
PDF
植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區
PPTX
khairi.pptx
PPTX
DR KUSH PRESENTATION.pptx
PPTX
Implantable cardiodefrillator in cardiology
PPT
PPTX
Aicd in asian settings ppt
PPTX
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
PPTX
Effortles Trial
PPTX
Anethesia and cardiac implantable electronic devices
PPT
defibrillation and cardioversion
PDF
1-dimensional convolutional neural networks for predicting sudden cardiac
PPTX
Pacemakers and implantable cardiac defibrillators
PDF
Research proposal focus cardiovascular and other disease biology - Pubrica ...
ICD troubleshooting
Real-Time Detection of Fatal Ventricular Dysrhythmias for Automated External ...
Implantable cardioverter defibrillator ICD.pptx
Rol actual del cardiodesfibrilador implantable subcutáneo en la prevención de...
How to read ICD episode_Dr Kush Bhagat.pptx
Basics of ICD - Dr Hafeesh Fazulu - Pushpagiri
植入性心臟電子儀器(CIED )的基本原理及設定_20130907北區
khairi.pptx
DR KUSH PRESENTATION.pptx
Implantable cardiodefrillator in cardiology
Aicd in asian settings ppt
IImplantable Cardioverter Defibrillators (ICDs) - Dr Prithvi puwar
Effortles Trial
Anethesia and cardiac implantable electronic devices
defibrillation and cardioversion
1-dimensional convolutional neural networks for predicting sudden cardiac
Pacemakers and implantable cardiac defibrillators
Research proposal focus cardiovascular and other disease biology - Pubrica ...
Ad

More from PRAVEEN GUPTA (20)

PPTX
Praveen's Indian cardiology board review and self-assessment
PPTX
Evaluation and management of Pacemaker malfunction
PPTX
Air pollution and cardiovascular disease
PPTX
Chronic thromboembolic pulmonary hypertension (CTEPH)
PPTX
ICD in Non-ischemic cardiomyopathy
PPTX
2015 Jones criteria for the diagnosis of rheumatic fever
PPTX
Non-Invasive physiological assessment of coronary circulatory function
PPTX
Cardiology mcq
PPTX
Heart block
PPTX
Pulseless electrical activity, Ventricular flutter, Ventricular fibrillation
PPTX
Kawasaki diseases
PPTX
Hybrid atrial fibrillation ablation
PPTX
Cardiology Journal club
PPTX
Cardiovascular complication of cancer chemotherapy
PPTX
Cardiology Journal scan
PPTX
SELECT-LV study (2017)
PPTX
Wearable defibrillator
PPTX
Diabetic cardiomyopathy (Pathophysiology)
PPTX
Heart sounds
PPTX
ECG/X-ray Quiz
Praveen's Indian cardiology board review and self-assessment
Evaluation and management of Pacemaker malfunction
Air pollution and cardiovascular disease
Chronic thromboembolic pulmonary hypertension (CTEPH)
ICD in Non-ischemic cardiomyopathy
2015 Jones criteria for the diagnosis of rheumatic fever
Non-Invasive physiological assessment of coronary circulatory function
Cardiology mcq
Heart block
Pulseless electrical activity, Ventricular flutter, Ventricular fibrillation
Kawasaki diseases
Hybrid atrial fibrillation ablation
Cardiology Journal club
Cardiovascular complication of cancer chemotherapy
Cardiology Journal scan
SELECT-LV study (2017)
Wearable defibrillator
Diabetic cardiomyopathy (Pathophysiology)
Heart sounds
ECG/X-ray Quiz

Recently uploaded (20)

PPTX
school management -TNTEU- B.Ed., Semester II Unit 1.pptx
PPTX
PPT- ENG7_QUARTER1_LESSON1_WEEK1. IMAGERY -DESCRIPTIONS pptx.pptx
PDF
O7-L3 Supply Chain Operations - ICLT Program
PDF
Microbial disease of the cardiovascular and lymphatic systems
PDF
grade 11-chemistry_fetena_net_5883.pdf teacher guide for all student
PDF
Supply Chain Operations Speaking Notes -ICLT Program
PDF
Abdominal Access Techniques with Prof. Dr. R K Mishra
PDF
Saundersa Comprehensive Review for the NCLEX-RN Examination.pdf
PDF
3rd Neelam Sanjeevareddy Memorial Lecture.pdf
PDF
STATICS OF THE RIGID BODIES Hibbelers.pdf
PDF
OBE - B.A.(HON'S) IN INTERIOR ARCHITECTURE -Ar.MOHIUDDIN.pdf
PDF
Black Hat USA 2025 - Micro ICS Summit - ICS/OT Threat Landscape
PDF
GENETICS IN BIOLOGY IN SECONDARY LEVEL FORM 3
PPTX
Introduction-to-Literarature-and-Literary-Studies-week-Prelim-coverage.pptx
PPTX
202450812 BayCHI UCSC-SV 20250812 v17.pptx
PDF
Chapter 2 Heredity, Prenatal Development, and Birth.pdf
PPTX
human mycosis Human fungal infections are called human mycosis..pptx
PDF
The Lost Whites of Pakistan by Jahanzaib Mughal.pdf
PDF
Computing-Curriculum for Schools in Ghana
PDF
FourierSeries-QuestionsWithAnswers(Part-A).pdf
school management -TNTEU- B.Ed., Semester II Unit 1.pptx
PPT- ENG7_QUARTER1_LESSON1_WEEK1. IMAGERY -DESCRIPTIONS pptx.pptx
O7-L3 Supply Chain Operations - ICLT Program
Microbial disease of the cardiovascular and lymphatic systems
grade 11-chemistry_fetena_net_5883.pdf teacher guide for all student
Supply Chain Operations Speaking Notes -ICLT Program
Abdominal Access Techniques with Prof. Dr. R K Mishra
Saundersa Comprehensive Review for the NCLEX-RN Examination.pdf
3rd Neelam Sanjeevareddy Memorial Lecture.pdf
STATICS OF THE RIGID BODIES Hibbelers.pdf
OBE - B.A.(HON'S) IN INTERIOR ARCHITECTURE -Ar.MOHIUDDIN.pdf
Black Hat USA 2025 - Micro ICS Summit - ICS/OT Threat Landscape
GENETICS IN BIOLOGY IN SECONDARY LEVEL FORM 3
Introduction-to-Literarature-and-Literary-Studies-week-Prelim-coverage.pptx
202450812 BayCHI UCSC-SV 20250812 v17.pptx
Chapter 2 Heredity, Prenatal Development, and Birth.pdf
human mycosis Human fungal infections are called human mycosis..pptx
The Lost Whites of Pakistan by Jahanzaib Mughal.pdf
Computing-Curriculum for Schools in Ghana
FourierSeries-QuestionsWithAnswers(Part-A).pdf

AICD programming

  • 1. Programming AICD Presenter Dr Praveen Gupta Moderator Dr Raja Selveraj Department of Cardiology 05.09.2017 JIPMER Pondicherry 1
  • 2. 2
  • 3. 3
  • 4. 4
  • 5. 5
  • 6. 6
  • 7. Detection and treatment of ventricular tachyarrhythmias by an ICD involves a series of sequential steps, each of which provides an opportunity to minimize unnecessary shocks 7
  • 8. Optimal ICD programming  Programming of the detection rate  Detection duration  Antitachycardia pacing (ATP)  Discriminate supraventricular tachycardia (SVT) from VT  Programming to minimize the sensing of noise 8 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 9. Rate and Duration for Initial Detection  ICDs detect VT/VF if the RR intervals are shorter than the detection interval for a programmable number of intervals or duration  An increased detection rate or prolonged duration is the most robust programming tool for prevention of unnecessary ICD therapy 9 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 10. 10 Rate and Duration for Initial Detection Primary prevention patients experience faster VTs with rates less likely to overlap SVT than secondary prevention patients Permits programming of faster VT rate cutoffs, ,minimizes unnecessary shocks in primary prevention patients Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 11. 11
  • 12. 12 Rate and Duration for Initial Detection PREPARE (Primary Prevention Parameters Evaluation) study The Detection rates 182 bpm, duration set for 30 of 40 intervals (≈9 seconds), patients less likely to receive a shock (9% versus 17%, P<0. .01), No difference in arrhythmic syncope between groups Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 13. SVT-VT Discrimination  Withhold inappropriate therapy  Hardware dependent 13 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 14. Single chamber discrimination  Sudden onset  Rate stability  Ventricular complexes morphology 14 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 15. Single-Chamber Algorithms Sudden onset  Sudden onset criterion bases action on the past 8 intervals:  Takes the interval average of every other detected ventricular event (4 out of 8)  Compares the current interval average to each of the four previous interval averages  If the sudden onset difference is less than the programmed delta, the onset is classified as gradual, favoring sinus tachycardia. Otherwise, the onset is classified as sudden, favoring VT. 15
  • 16. Single-Chamber Algorithms Stability  The stability delta is the widest difference between the second longest and second shortest cycles, among 12 (programmable value) consecutive cycles of tachycardia detection (discards the fastest and the slowest intervals and subtracts the shortest interval from the longest interval). If the stability difference is less than the programmed delta, the rhythm is VT. Otherwise favors AF Does not, discriminate sinus tachycardia, atrial tachycardia or atrial flutter from VT. In this example: stability ignores longest (395 ms) and shortest (268 ms) intervals and subtracts second longest (378 ms) from second shortest (271 ms): 378 ms - 271 ms = 107 m (> 80 ms the programmed value); therefore interval stability indicates SVT (AF) 16
  • 17. 17 Compare electrograms during tachycardia with a template acquired during normal rhythm Single-Chamber Algorithms Morphology
  • 18. Single-Chamber Algorithms Morphology  Tachycardia morphology similar from template by programmable threshold is SVT Example: SVT = Match (% match above 60% threshold) The algorithm breaks the waveform apart into triangular chunks and compares the A pieces, the B pieces, etc. At each point, it determines how much discrepancy there is between template and test. In this case, there is none, which results in a 100% match. In such a scenario, the algorithm would determine the rhythm was an SVT (sinus in origin) rather than a VT. 18
  • 19. Single-Chamber Algorithms Morphology  Tachycardia morphology differs from template by programmable threshold is VT Example: VT = Non-Match (35% match is less than 60% threshold) Each part is compared (A to A, B to B) etc. This drawing shows how these particular pieces do not match exactly; the far right shows areas that one waveform has and the other does not (in other words, the blue areas is the one that does not match). The MD algorithm assesses this numerically; in this case, the waveforms are only a 35% match (in favor of VT). 19
  • 20. Single-Chamber Algorithms  Morphology algorithm greater sensitivity and specificity  Sensitivity 92% to 99% and specificities of 90% to 97%  Interval algorithms (onset and stability) unreliable at higher rates (>180–200 bpm  Morphology is the only SVT-VT discriminator reliably applied at rates >200 bpm 20 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 21. Dual-Chamber SVT-VT Discriminators 21  Use information collected simultaneously from the atria and ventricles  Dual chamber devices discriminators, depending on the AV relationship used to classify the arrhythmia (V>A, V<A, V=A). Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 22. Dual-Chamber SVT-VT Discriminators The ventricular rate is faster than the atrial one in the majority (more than 80%) of the ventricular tachycardias or fibrillations, this proportion even increases with very fast rhythms. Therefore, the 2 other branches (V=A and V<A) apply for less than 20% of ventricular tachycardias VT arm (V>A), When the ventricular rate is faster than the atrial rate, the tachycardia is classified as VT. No other discrimination criterion is analyzed 22
  • 23. Dual-Chamber SVT-VT Discriminators  When the ventricular rate is slower than the atrial rate, one must distinguish AF, atrial tachycardia or flutter, and VT,  Rate stability  Stability of AV association  QRS morphology The rate stability and morphology criteria operate as described for single chamber devices. AV association operates on the same principle as rate stability, i.e. the AV association delta measures the difference between the second longest and second shortest AV intervals among the 12 tachycardia detection cycles. When Interval Stability classifies a rhythm disorder as VT, AV association goes into effect. 23 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 24. Dual-Chamber SVT-VT Discriminators 24  Ventricular and the atrial rates are same, using the following criteria:  AV interval delta  Sudden onset  QRS morphology Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 25. Dual-Chamber SVT-VT Discriminators 25  AV Interval discrimination is evaluated first.  If AV interval points to VT, then no other discrimination is applied, and VT therapy begins.  If AV interval points to SVT, then the rhythm is evaluated by 2 other discriminators in the V=A rate branch: the sudden onset and morphology criteria Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 27. Dual-Chamber SVT-VT Discriminators  Use atrioventricular (AV) association to distinguish SVT from VT  Proper atrial and ventricular sensing is essential  Atrial undersensing caused by AF, lead dislodgement, functional, resulting from the atrial blanking that occurs during and immediately after ventricular sensed or paced events  Atrial undersensing lead to misclassification of SVT as VT 27 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 28. Dual-Chamber SVT-VT Discriminators  Oversensing by large FFRWs , misclassification of VT as SVT, inappropriately withheld therapy  Prevent both far-field R-wave (FFRW) oversensing and the undersensing of small fibrillation waves on the atrial channel 28 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 29. Single- Versus Dual-Chamber ICD  No benefit of dual-chamber over single-chamber  Dual-chamber discrimination is superior when overlap in VT and SVT rates  Dual-chamber ICDs for secondary prevention and for patients with slow VT  In primary prevention patients, little overlap between SVT and VT heart rates, the benefit of using high rate cutoffs and prolonged detection appears to overwhelm any modest additional contribution of dual-chamber SVT-VT discrimination 29 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 30. Antitachycardia Pacing  ATP is rapid pacing at a cycle length shorter than VT  Terminates VT by penetrating the circuit  ATP terminate slow VT (<188–200 bpm),  Success 85% to 90% and a 1% to 5% risk of acceleration  VT in primary prevention are monomorphic, potential application of ATP 30 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 31. ATP 31  Rate of ATP programmed at 69% to 88% of VT cycle length  Each drive train typically has 8 stimuli  First ATP most effective, 95% of responsive fast VTs terminated successfully  Only a small minority of patients are responsive to >3 or 4 ATPs Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 32. Factors affecting efficacy of ATP  Burst ATP-Interstimulus interval in the train remains constant  Ramp ATP-Each subsequent stimulus in a train is decremented  Burst and ramp pacing have similar efficacy and safety for slow VT  Fast VT, burst pacing 32 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 33. Redetection and Reconfirmation  Contemporary devices reconfirm the persistence of arrhythmia during and after charging.  Charge is dissipated if arrhythmia terminates  Noncommitted shock if possible  The ADVANCE-CRT trial reported greater efficacy of biventricular ATP in patients with ischemic cardiomyopathy. 33 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 34. Antitachycardia Pacing Analysis of the response to ATP  Used to optimize programming  Provides measure of temporal proximity of the pacing site to the VT  Differential diagnosis of SVT and VT  Transient VA and AV block during ATP is diagnostic of VT and SVT, respectively  An AAV response is diagnostic of atrial tachycardia, and a VVA response is diagnostic of VT 34 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 35. Programming Therapy Zones  ICDs have 2 to 3 zones defined by the longest RR interval in each zone  The lower boundary between sinus and VT zone is programmed at approximately 180 to 188 bpm in primary prevention patients and at 30 to 60 ms greater than the cycle length of the slowest observed VT (150–160 bpm in most trials) in secondary prevention patients  Programming in the VT zone is focused on preventing therapy for SVT and nonsustained VT 35 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 36. Programming Therapy Zones  If 2 VT zones are programmed, the rate at which fewer trials of ATP therapy are desired forms the boundary between the 2 zones.  In secondary prevention, slower VT zone is programmed with 3 to 4 sequences of ramp/burst ATP compared with 1-burst ATP followed by shocks in the faster VT zone.  SVT discriminators are programmed on in VT zones. Implantable cardioverter defibrillator (ICD) rate detection zones. Some ICDs permit programming of an additional monitor-only zone. The upper panel shows programming for secondary prevention patients. The lower panel shows programming for primary prevention patients. 36 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 37. Programming Therapy Zones  The boundary between the VT and VF zones is determined by the rate at which ATP before charge is no longer desired  SVT discriminators are either not programmable in the VF zone (Boston Scientific and St Jude Medical) or are applied with some limitations (Medtronic) 37 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 38. ATP should not be routinely programmed 38  Inherited channelopathies (Brugada syndrome, Long and Short QT syndrome)  Catecholaminergic polymorphic VT  Early repolarization syndromes  Arrhythmia is polymorphic VT or VF  Arrhythmias lack reentry and not interrupted by pacing
  • 39. Shock Strength, Polarity, and Defibrillation Threshold 39  Implant testing assesses VF sensing and defibrillation effectiveness  SAFE-ICD study (Safety of Two Strategies of ICD Management at Implantation) revealed comparable rates of ICD complications and long-term sudden death in patients who did and did not undergo routine DFT testing Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 40. Shock Strength, Polarity, and Defibrillation Threshold  Strongly considered in nonstandard lead positioning, right-sided devices, amiodarone therapy, pediatric implantations, and use of an s-ICD 40 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 41. Shock Strength, Polarity, and Defibrillation Threshold  If the DFT is known, the first shock in the VT zone is programmed ≥10 J above it, and subsequent shocks are set to maximal energy  The advantage of lower-energy shocks (conservation of battery life, diminished postshock stunning or block) is minimal with current biphasic technology.  High-energy shocks are more likely to be successful for VT/VF and SVT without patient discomfort or increase in charge time  Low-energy shocks also convey a risk of conversion of VT to VF  Maximal shock strength is generally preferred 41 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 42. Shock Strength, Polarity, and Defibrillation Threshold  Right ventricular coil as the anode resulted in slightly lower DFT in some patients  Reverse-shock polarity, can be tried in patients with high DFTs. 42
  • 43. Optimization of Sensing to Prevent Shocks 43  ICD should reliably sense every QRS complex during normal rhythm and small fibrillatory waves during VF  Do not sense T waves  Ignore extracardiac and nonphysiological signals  ICDs use dynamic sensitivity or gain to achieve these sensing requirements Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 44. Optimization of Sensing to Prevent Shocks  After a sensed or paced ventricular event, sensitivity is reduced as a function of the amplitude of the R wave after the postventricular blanking period expires.  Sensitivity is increased until a subsequent event is sensed or the maximum programmed sensitivity is reached  Strategy reduces risk of T-wave oversensing while maintaining sensitivity for the small deflections of VF 44 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 45. Preventing T-Wave Oversensing 45  Low-amplitude R waves (<3 mV), large T waves, or long QT intervals promote T-wave oversensing, result in double counting and inappropriate therapy.  Hypertrophic cardiomyopathy, long-QT syndrome, and Brugada syndrome predispose T- wave oversensing  Decreasing ventricular sensitivity reduces the risk of T-wave oversensing but requires a sufficiently large R- to T-wave ratio and may risk VF undersensing Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 46. Surveillance of Lead Fracture  Lead Integrity Alert algorithm for detection of lead fracture based on sudden changes in lead impedance and noise detection  The lead-integrity alert is triggered if 2 of 3 events occur:  (1) Sudden increase in lead impedance  (2) ≥2 nonsustained tachycardia events (≥5 beats) with intervals <220 ms  (3) sensing integrity counters with ≥30 short RR-interval counts <140 ms within 3 consecutive days indicative of nonphysiological signals 46 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 47. Noise-Detection Algorithms  Algorithms for discrimination of VT from myopotentials and nonphysiological noise such as electromagnetic interference  Noise have higher frequency with very short intervals  If noise is detected, the device raises the floor of the dynamic sensitivity above the level of the noise to prevent oversensing  This is most useful to prevent myopotential detection 47 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 48. ICD Programming During Electrical Storm  Recurrent ICD shocks in the course of electrical storm (defined as ≥3 VT/VF episodes in 24 hours)  Lead to heart failure, sympathetic overdrive, and psychological trauma, which potentiates the risk for recurrent VT. 48 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 49. ICD Programming During Electrical Storm 49  Programming for minimizing the number of shocks  Duration to detection is increased  Number of intervals to declare an end of episode decreased  Increasing the lower pacing rate may help suppress recurrent VT/VF  Hospitalized patients, ICD therapies are turned off during electrical storm Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 50. Remote Monitoring  Automated, wireless remote monitoring with clinician and patient alerts shortens the time to detection of changes in clinical status and device function and may reduce mortality risk 50 Madhavan M, Friedman PA. Optimal programming of implantable cardiac-defibrillators. Circulation. 2013 Aug 6;128(6):659-72.
  • 51. Conclusions  Proper ICD programming minimizes unnecessary ICD therapy to reduce patient morbidity and mortality.  Current evidence supports the use of various strategies to reduce shocks without compromising therapy effectiveness.  The extension of detection duration to prevent treatment of self-terminating tachycardia has proven to be the most useful of these strategies, followed by the application of SVT-VT discrimination and the use of ATP to terminate VT 51
  • 52. Disclosures  Dr Friedman has received research support from Medtronic (grant administered by Mayo Clinic for investigator-initiated study), Biotronik, and Cameron Health, and is a speaker or consultant for Bard, Biotronik, Leadex, and Sorin.  Dr Madhavan reports no conflicts 52