SlideShare a Scribd company logo
Chapter 1Measurement1
After you have read this chapter, you should be able to…Explain the difference between analog and digital measurement systems.Define accuracy, hysteresis, sensitivity.List the SI units for length, time, mass, and electric current.Perform conversion from other units to Si units or vice versa. Identify the standard transmission signal and its P&ID symbols.Draw a typical first-order time response curve.2
IntroductionA measurement refers to the conversion of the variable into some corresponding analog of the variable, such as pneumatic pressure, an electrical voltage or current, or a digitally encoded signal.3
Measurement System A measurement system consists of :A sensor that performs the initial measurement and energy conversion of a variable into analogues digital, electrical, or pneumatic information.A signal conditioning for further transformation to complete the measurement function.The sensor is also called a “transducer” -a device that converts any signal from one form to another.4
Figure 1.2 & fig 1.3 – example of measure5A human can regulate the level using a sight tube, S, to compare the level, h, to the objective, H, and adjust a valve to change the level.
6An automatic level-control system replaces the human with a controller and uses a sensor to measure the level.
Analog Measurement SystemFigure 1.14 (pg. 18)7An analog control system such as this allows continuous variation of some parameter, such as heat input, as a function of error.
Digital Measurement SystemFigure 1.15 (pg. 19)8In supervisory control, the computer monitors measurements and updates set points, but the loops are still analog in nature.
Units, Standards, and DefinitionsThe field of measurement has many sets of units, standards, and definitions to describe its characteristics.In this section, we will go through common units, standards and definitions used in the measurement technology.9
UnitsTo ensure precise technical communication among individuals employed in technological disciplines, it is essential to use a well-defined set of units of measurement.SI units  (International System on Units )Other units (CGS, English system)In United States, the English system of units is still common in use. We should be able to translate between the SI unit to other unit or vice versa. Refer Appendix 1.10
Examples:Express a pressure of p =2.1x103 dyne/cm2 in pascals. 1 Pa=1 N/m2	Solution : 100cm= 1 m and 105 dyne = 1 newton; thus 	p = ? N/m2 =? pascals (Pa) 2.	Express 6.00 ft in meters. 	Solution : Using 39.37 in./m and 12 in./ft; thus 			?11
3.  Find the number of feet in 5.7m. 	Solution : 1 m=39.37in and 1 ft=12 in; 	thus ?4.   Express 6.00 ft in meters. 	Solution : Using 39.37 in./m and 12 in./ft; thus  ?5.	Find the mass in kilograms of a 2-lb object.	Solution : The conversion factor between mass in kilograms and pounds to be 0.454 kg/lb. Therefore, we have	m =??12
SI Units13
Analog Data RepresentationPart of the specification in measurement system is the range of variables involved.Example : slide 7A thermistor sensor is used to measure temperature in an oven. The range of temperature inside the oven is from 0oC to 250oC. - 0oC to 250oC is the range of variables in the process (in this case oven).14
Analog Data RepresentationTwo analog standards are in common use as a means of representing the range of variables.1.Current signals 	: 4 – 20 mA2. Pneumatic signals	: 3 – 15 psiThese signals are used to transmit variable information over some distance, such as to and from the control room and plant. 15
Analog Data RepresentationElectric current and pneumatic pressures are the most common means of information transmission in the industrial environment.16
Analog Data RepresentationFigure 1.6(a) pg 917The physical diagram of a control loop and its corresponding block diagram look similar. Note the use of current- and pressure-transmission signals.
Current Signal : 4-20 mAThe common current transmission signal is 4 to 20 mA.In the previous example, range of temperature inside the oven is from 0oC to 250oC. 0oC might be represented by 4 mAand 250oC by 20 mA, with all temperatures in between represented by a proportional current.18
Why current is used for  transmission instead of voltage?19One of the advantages of current as a transmission signal is that it is nearly independent of line resistance.
Pneumatic Signal : 3 – 15 psiThe common pneumatic transmission signal is 3 to 15 psi.When a sensor measures some variable in a range (for example, temperature in oven 0oC to 250oC),  it is converted into proportional pressure of gas in pipe.The gas is usually dry air.The English system standard is still widely used in the United States, despite the move to the SI units. The equivalent SI range is 20 to 100 kPa.20
DefinitionsErrorAccuracyBlock DiagramTransfer functionSensitivityHysteresis and ReproducibilityResolutionLinearity21
ErrorError is the difference between the actual value of a variable and the measured indication of its value.Example : The current temperature inside the oven is 900C. However, the measurement from temperature sensor inside the oven is 890C.Therefore, the error is 10C.22
AccuracyTo specify the maximum overall error to be expected from a device, such as measurement of a variable.Can appear in several forms such as :Measured variable% of the instrument full-scale (FS) reading% of instrument span% of actual reading23
AccuracyMeasured variableExample : The accuracy of temperature sensor used to measure the oven temperature is ± 1oC.Thus, if the current temperature inside the oven is 90oC, a reading between 89oC to 91oC is acceptable.24
Accuracy2. % of the instrument full-scale (FS) readingExample :An analog multimeter with 5-V full-scale range gives an accuracy of ± 00.5% FS.This means the uncertainty in any measurement is ±0.025V.If we want to measure a 5-V voltage using this analog multimeter, the measurement between (5 -0.025)V and (5+0.025)V is acceptable.25
Accuracy3. % of instrument spanPercentage of the range of instrument measurement capability.Example :A pressure sensor device is measuring ±3% of span for a 3 to 15 psi range of pressure.Thus, the accuracy of the device is 			(±0.03)(15-3)=±0.36 psi.26
Accuracy4. % of actual readingExample :For a 2% of reading voltmeter, we will have an inaccuracy of ±0.04V for a reading of 2V.Please look into Example 1.8, 1.9,1.10 for further understanding about accuracy.27
Block DiagramA block has an input of some variable, x(t) and an output of another variable y(t).Figure 1.21 (pg. 27)28A transfer function shows how a system-block output variable varies in response to an input variable, as a function of both static inputvalue and time.
Transfer FunctionT(x,y,t) describes the relationship between the input and output for the block.29A transfer function shows how a system-block output variable varies in response to an input variable, as a function of both static inputvalue and time.
SensitivityA measure of change in output of an instrument for a change in input.High sensitivity is desirable in an instrument.Sensitivity must be evaluated together with other parameters, such as linearity of output to input, range, and accuracy.The value of sensitivity is indicated by transfer function.Example : when temperature transducer outputs 5 mV per degree Celcius, the sensitivity is 5 mV/oC.30
Hysteresis and Reproducibility31Hysteresis is a predictable error resulting from differences in the transfer function as the input variable increases or decreases.
ResolutionResolution of a device is a minimum measurable value of the input variable.This characteristic of the instrument can be changed only by redesign.Example 1.12 : A force sensor measures a range of 0 to 150N with a resolution of 0.1% FS. Find the smallest change in force that can be measured.Solution :Because the resolution is 0.1% FS, we have a resolution of (0.001)(150N) = 0.15N, which is the smallest measurable change in force.32
LinearityIn both sensor and signal conditioning, a linear relationship between input and output is highly desirable. Example :Consider a sensor that outputs a voltage as a function of pressure from 0 to 100 psi with a linearity of 5% FS. This means, at some point on the curve of voltage versus pressure, the deviation between actual pressure and linearly indicated pressure for a given voltage deviates by 5% of 100psi, or 5 psi.33
Linearity     Comparison of an actual curve and its best-fit straight line, where the maximum deviation is 5% FS.34
DrawingsAn electrical schematic is a drawing that employs a standard set of symbols and definitions so that anyone who knows the standards can understand the operation of the circuit.-The standard was developed and approved by American National Standards Institute (ANSI) and Instrumentation, Systems, and Automation (ISA) society. -  Known as : ANSI/ISA S5.1-1984(R1992) Instrumentation Symbols and Identification.35
DrawingsPiping & Instrumentation Diagram (P&ID)36Computers and programmable logic controllers are included in the P&ID.
Sensor Time ResponseThe sensing element inside the sensor has a time dependence that specifies how the output changes in time when the input is changing in time.Time response analysis is always applied to the output of the sensor.Figure 1.2737
Sensor Time ResponseFigure 1.2738The dynamic transfer function specifies how a sensor output varies when the input changes instantaneously in time (i.e., a step change).
2.2.5	Sensor Time ResponseFirst Order Response and Time Constant39Characteristic first-order exponential time response of a sensor to a step change of input.
Tutorial Problems – Chapter 1Section 1.5 – 	Q1.8Section 1.6 – 	Q 1.15, 1.16, 1,17, 1.18,1.19,1.20,1.21,1.23, 1.243. Additional questions -Describe what elements would be necessary to do (a) analog measurement and (b) digital measurement system.By referring  to Figure 1.36 (Supplementary Problems QS1.8), identify all the sensors available in the drawing and the type of signals that connect the sensors (4-20 mA or 3-15 psi). Refer to Appendix 5 for P&ID symbols.40

More Related Content

PPTX
Mahindra & Mahindra training report.
PPTX
Presentation1.pptx
PPTX
Piezoelectric accelerometer
PPT
Vibration and frequency measuring instruments
PPT
Lecture1 measurement & intrumentation
PPTX
Region 5 - Bicol Region Philippines
PPTX
Word of Mouth Marketing
Mahindra & Mahindra training report.
Presentation1.pptx
Piezoelectric accelerometer
Vibration and frequency measuring instruments
Lecture1 measurement & intrumentation
Region 5 - Bicol Region Philippines
Word of Mouth Marketing

What's hot (20)

PPTX
Electronic Measurement and Instrumentation
PPT
Sensor and transducers
PPTX
Oscilloscope
PDF
Multimeter basics
PPT
Transducers
PPTX
Functional elements of an instrument
PDF
Basics of Sensors & Transducers
PPTX
Strain gauge
PPT
Sensors and transducers 1.ppt
PDF
Electronic Measurement - Power Factor Meter
PPTX
Transducers & transmitters
PPT
Sensor and transducers lect 1
PPTX
Transducers,Active Transducers and Passive Transducers
PPT
Static and Dynamic characteristics of Measuring Instrument
PPT
Electronic passive components
PPTX
Measurement of resistance
PPTX
Control systems engineering
PPTX
Unit 1(part-2)sensors and transducer
PPT
Characteristic of transducetr.ppt
PPTX
Classification of transducers
Electronic Measurement and Instrumentation
Sensor and transducers
Oscilloscope
Multimeter basics
Transducers
Functional elements of an instrument
Basics of Sensors & Transducers
Strain gauge
Sensors and transducers 1.ppt
Electronic Measurement - Power Factor Meter
Transducers & transmitters
Sensor and transducers lect 1
Transducers,Active Transducers and Passive Transducers
Static and Dynamic characteristics of Measuring Instrument
Electronic passive components
Measurement of resistance
Control systems engineering
Unit 1(part-2)sensors and transducer
Characteristic of transducetr.ppt
Classification of transducers
Ad

Viewers also liked (14)

PPTX
Electrode Interfacing
PPT
Static characteristics of Instruments
PPT
Introduction to sensors
PDF
Bio Potential and Bio Electrodes
PPT
Sensors
PPTX
Introduction to sensors
PPTX
Characteristics of instruments
PPTX
Transducers
PPT
Sensor Characteristics and Selection
PPTX
Sensors and transducers
PPSX
Accuracy & Precision
PPT
Sensors and their applications
PPTX
Accuracy and Precision
Electrode Interfacing
Static characteristics of Instruments
Introduction to sensors
Bio Potential and Bio Electrodes
Sensors
Introduction to sensors
Characteristics of instruments
Transducers
Sensor Characteristics and Selection
Sensors and transducers
Accuracy & Precision
Sensors and their applications
Accuracy and Precision
Ad

Similar to Chapter 1 Measurement System (20)

PDF
Introduction to Instrumentation
PPTX
Ch 01 (Measurement Systems).pptx
PPTX
Ch 01 (Measurement Systems).pptx
PDF
Instrumentation_1_Review-IntroductionToInstrumentation.pdf
PPT
SENSORS_AND_INST.UNIT_1_REG_17__Presentation.ppt
PPTX
Lecture 02 instrument transducers manipal
PPTX
Biomedical instrumentation in TY BSc Physics
PPTX
Measurement
PPT
Introduction to Measurement Transducers.ppt
PDF
Mechatronics study material-Question Bank.pdf
PDF
Instrumentation_ Measurment_course_Chapter_2_.pdf
PDF
Basic concepts of instrumentation
PPT
Fundamentals on Measurement and uncertainty
PPTX
measurments
PPT
01. Sensors and Actuators.ppt
PPTX
Metrolog and Measurments Lecture number 1
PDF
Basic Instrument
PPTX
Lecture Notes: EEEC6430312 Measurements And Instrumentation - Instrument Typ...
PPTX
Ch 4 _mmm mesurement systems and basic of mesurement method.pptx
PPT
Metrology and Gauging Dynamic Characteristics.ppt
Introduction to Instrumentation
Ch 01 (Measurement Systems).pptx
Ch 01 (Measurement Systems).pptx
Instrumentation_1_Review-IntroductionToInstrumentation.pdf
SENSORS_AND_INST.UNIT_1_REG_17__Presentation.ppt
Lecture 02 instrument transducers manipal
Biomedical instrumentation in TY BSc Physics
Measurement
Introduction to Measurement Transducers.ppt
Mechatronics study material-Question Bank.pdf
Instrumentation_ Measurment_course_Chapter_2_.pdf
Basic concepts of instrumentation
Fundamentals on Measurement and uncertainty
measurments
01. Sensors and Actuators.ppt
Metrolog and Measurments Lecture number 1
Basic Instrument
Lecture Notes: EEEC6430312 Measurements And Instrumentation - Instrument Typ...
Ch 4 _mmm mesurement systems and basic of mesurement method.pptx
Metrology and Gauging Dynamic Characteristics.ppt

Chapter 1 Measurement System

  • 2. After you have read this chapter, you should be able to…Explain the difference between analog and digital measurement systems.Define accuracy, hysteresis, sensitivity.List the SI units for length, time, mass, and electric current.Perform conversion from other units to Si units or vice versa. Identify the standard transmission signal and its P&ID symbols.Draw a typical first-order time response curve.2
  • 3. IntroductionA measurement refers to the conversion of the variable into some corresponding analog of the variable, such as pneumatic pressure, an electrical voltage or current, or a digitally encoded signal.3
  • 4. Measurement System A measurement system consists of :A sensor that performs the initial measurement and energy conversion of a variable into analogues digital, electrical, or pneumatic information.A signal conditioning for further transformation to complete the measurement function.The sensor is also called a “transducer” -a device that converts any signal from one form to another.4
  • 5. Figure 1.2 & fig 1.3 – example of measure5A human can regulate the level using a sight tube, S, to compare the level, h, to the objective, H, and adjust a valve to change the level.
  • 6. 6An automatic level-control system replaces the human with a controller and uses a sensor to measure the level.
  • 7. Analog Measurement SystemFigure 1.14 (pg. 18)7An analog control system such as this allows continuous variation of some parameter, such as heat input, as a function of error.
  • 8. Digital Measurement SystemFigure 1.15 (pg. 19)8In supervisory control, the computer monitors measurements and updates set points, but the loops are still analog in nature.
  • 9. Units, Standards, and DefinitionsThe field of measurement has many sets of units, standards, and definitions to describe its characteristics.In this section, we will go through common units, standards and definitions used in the measurement technology.9
  • 10. UnitsTo ensure precise technical communication among individuals employed in technological disciplines, it is essential to use a well-defined set of units of measurement.SI units (International System on Units )Other units (CGS, English system)In United States, the English system of units is still common in use. We should be able to translate between the SI unit to other unit or vice versa. Refer Appendix 1.10
  • 11. Examples:Express a pressure of p =2.1x103 dyne/cm2 in pascals. 1 Pa=1 N/m2 Solution : 100cm= 1 m and 105 dyne = 1 newton; thus p = ? N/m2 =? pascals (Pa) 2. Express 6.00 ft in meters. Solution : Using 39.37 in./m and 12 in./ft; thus ?11
  • 12. 3. Find the number of feet in 5.7m. Solution : 1 m=39.37in and 1 ft=12 in; thus ?4. Express 6.00 ft in meters. Solution : Using 39.37 in./m and 12 in./ft; thus ?5. Find the mass in kilograms of a 2-lb object. Solution : The conversion factor between mass in kilograms and pounds to be 0.454 kg/lb. Therefore, we have m =??12
  • 14. Analog Data RepresentationPart of the specification in measurement system is the range of variables involved.Example : slide 7A thermistor sensor is used to measure temperature in an oven. The range of temperature inside the oven is from 0oC to 250oC. - 0oC to 250oC is the range of variables in the process (in this case oven).14
  • 15. Analog Data RepresentationTwo analog standards are in common use as a means of representing the range of variables.1.Current signals : 4 – 20 mA2. Pneumatic signals : 3 – 15 psiThese signals are used to transmit variable information over some distance, such as to and from the control room and plant. 15
  • 16. Analog Data RepresentationElectric current and pneumatic pressures are the most common means of information transmission in the industrial environment.16
  • 17. Analog Data RepresentationFigure 1.6(a) pg 917The physical diagram of a control loop and its corresponding block diagram look similar. Note the use of current- and pressure-transmission signals.
  • 18. Current Signal : 4-20 mAThe common current transmission signal is 4 to 20 mA.In the previous example, range of temperature inside the oven is from 0oC to 250oC. 0oC might be represented by 4 mAand 250oC by 20 mA, with all temperatures in between represented by a proportional current.18
  • 19. Why current is used for transmission instead of voltage?19One of the advantages of current as a transmission signal is that it is nearly independent of line resistance.
  • 20. Pneumatic Signal : 3 – 15 psiThe common pneumatic transmission signal is 3 to 15 psi.When a sensor measures some variable in a range (for example, temperature in oven 0oC to 250oC), it is converted into proportional pressure of gas in pipe.The gas is usually dry air.The English system standard is still widely used in the United States, despite the move to the SI units. The equivalent SI range is 20 to 100 kPa.20
  • 22. ErrorError is the difference between the actual value of a variable and the measured indication of its value.Example : The current temperature inside the oven is 900C. However, the measurement from temperature sensor inside the oven is 890C.Therefore, the error is 10C.22
  • 23. AccuracyTo specify the maximum overall error to be expected from a device, such as measurement of a variable.Can appear in several forms such as :Measured variable% of the instrument full-scale (FS) reading% of instrument span% of actual reading23
  • 24. AccuracyMeasured variableExample : The accuracy of temperature sensor used to measure the oven temperature is ± 1oC.Thus, if the current temperature inside the oven is 90oC, a reading between 89oC to 91oC is acceptable.24
  • 25. Accuracy2. % of the instrument full-scale (FS) readingExample :An analog multimeter with 5-V full-scale range gives an accuracy of ± 00.5% FS.This means the uncertainty in any measurement is ±0.025V.If we want to measure a 5-V voltage using this analog multimeter, the measurement between (5 -0.025)V and (5+0.025)V is acceptable.25
  • 26. Accuracy3. % of instrument spanPercentage of the range of instrument measurement capability.Example :A pressure sensor device is measuring ±3% of span for a 3 to 15 psi range of pressure.Thus, the accuracy of the device is (±0.03)(15-3)=±0.36 psi.26
  • 27. Accuracy4. % of actual readingExample :For a 2% of reading voltmeter, we will have an inaccuracy of ±0.04V for a reading of 2V.Please look into Example 1.8, 1.9,1.10 for further understanding about accuracy.27
  • 28. Block DiagramA block has an input of some variable, x(t) and an output of another variable y(t).Figure 1.21 (pg. 27)28A transfer function shows how a system-block output variable varies in response to an input variable, as a function of both static inputvalue and time.
  • 29. Transfer FunctionT(x,y,t) describes the relationship between the input and output for the block.29A transfer function shows how a system-block output variable varies in response to an input variable, as a function of both static inputvalue and time.
  • 30. SensitivityA measure of change in output of an instrument for a change in input.High sensitivity is desirable in an instrument.Sensitivity must be evaluated together with other parameters, such as linearity of output to input, range, and accuracy.The value of sensitivity is indicated by transfer function.Example : when temperature transducer outputs 5 mV per degree Celcius, the sensitivity is 5 mV/oC.30
  • 31. Hysteresis and Reproducibility31Hysteresis is a predictable error resulting from differences in the transfer function as the input variable increases or decreases.
  • 32. ResolutionResolution of a device is a minimum measurable value of the input variable.This characteristic of the instrument can be changed only by redesign.Example 1.12 : A force sensor measures a range of 0 to 150N with a resolution of 0.1% FS. Find the smallest change in force that can be measured.Solution :Because the resolution is 0.1% FS, we have a resolution of (0.001)(150N) = 0.15N, which is the smallest measurable change in force.32
  • 33. LinearityIn both sensor and signal conditioning, a linear relationship between input and output is highly desirable. Example :Consider a sensor that outputs a voltage as a function of pressure from 0 to 100 psi with a linearity of 5% FS. This means, at some point on the curve of voltage versus pressure, the deviation between actual pressure and linearly indicated pressure for a given voltage deviates by 5% of 100psi, or 5 psi.33
  • 34. Linearity Comparison of an actual curve and its best-fit straight line, where the maximum deviation is 5% FS.34
  • 35. DrawingsAn electrical schematic is a drawing that employs a standard set of symbols and definitions so that anyone who knows the standards can understand the operation of the circuit.-The standard was developed and approved by American National Standards Institute (ANSI) and Instrumentation, Systems, and Automation (ISA) society. - Known as : ANSI/ISA S5.1-1984(R1992) Instrumentation Symbols and Identification.35
  • 36. DrawingsPiping & Instrumentation Diagram (P&ID)36Computers and programmable logic controllers are included in the P&ID.
  • 37. Sensor Time ResponseThe sensing element inside the sensor has a time dependence that specifies how the output changes in time when the input is changing in time.Time response analysis is always applied to the output of the sensor.Figure 1.2737
  • 38. Sensor Time ResponseFigure 1.2738The dynamic transfer function specifies how a sensor output varies when the input changes instantaneously in time (i.e., a step change).
  • 39. 2.2.5 Sensor Time ResponseFirst Order Response and Time Constant39Characteristic first-order exponential time response of a sensor to a step change of input.
  • 40. Tutorial Problems – Chapter 1Section 1.5 – Q1.8Section 1.6 – Q 1.15, 1.16, 1,17, 1.18,1.19,1.20,1.21,1.23, 1.243. Additional questions -Describe what elements would be necessary to do (a) analog measurement and (b) digital measurement system.By referring to Figure 1.36 (Supplementary Problems QS1.8), identify all the sensors available in the drawing and the type of signals that connect the sensors (4-20 mA or 3-15 psi). Refer to Appendix 5 for P&ID symbols.40