Multidimensional Approaches for Noise
Cancellation of ECG signal
Akash Kumar Bhoi, Karma Sonam Sherpa, Devakishore Phurailatpam, Jitendra Singh Tamang, Pankaj Kumar Giri
Abstract: In many situations, the Electrocardiogram (ECG) is
recorded during ambulatory or strenuous conditions such that the
signal is corrupted by different types of noise, sometimes
originating from another physiological process of the body. Hence,
noise removal is an important aspect of signal processing. Here five
different filters i.e. median, Low Pass Butter worth, FIR, Weighted
Moving Average and Stationary Wavelet Transform (SWT) with
their filtering effect on noisy ECG are presented. Comparative
analyses among these filtering techniques are described and
statically results are evaluated.
Index Terms- Electrocardiography, Active noise reduction, Filters,
Noise cancellation.
I. INTRODUCTION
A physician can detect a heart problem from this
information and can suggest timely measures. But during the
acquisition of ECG signal, it may get corrupted by different
types of noises [21] which make it difficult for the physician to
give his diagnosis. Power Line Interference (PLI) is one such
kind of noise which superimposes on the vital information. The
frequency range of ECG signal is 0.05Hz to 150Hz, and the
frequency of the PLI noise is 50/60 Hz which lies within the
frequency spectrum of the ECG signal, so for the meaningful and
accurate detection, steps have to be taken to filter out or discard
all these noise sources. Hamilton PS in his article compared
adaptive and non adaptive filters for reduction of power line
interference in the ECG [5]. Iders YZ, Saki MC, Gcer HA have
developed a method for line interference reduction to be used in
signal averaged electrocardiography [6].
Cramer E, McManus CD, Neubert D has introduced a global
filtering approach. In this method two types of the digital filters
are used where, one is using lest square method and other is
using special summation method [7]. Different scientists have
tried for removing the power line interference and base line
wonder specifically from the ECG signal [8-17]. Zschorlich VR
and Zschorlich VR, have also designed digital filters to cope
with EMG signals [18-19]. Webster has explained the
instrumentation requirements for the ECG [20].
Akash Kumar Bhoi is with the Applied Electronics & Instrumentation
Engineering Department, Sikkim Manipal Institute of Technology (SMIT), India
(email: akash730@gmail.com).
Karma Sonam Sherpa is with the Electrical & Electronics Engineering
Department, Sikkim Manipal Institute of Technology (SMIT), India (email:
karmasherpa23@gmail.com).
The wavelet coefficients represent a measure of similarity in
the frequency content between a signal and a chosen wavelet
function [1]. These coefficients are computed as a convolution of
the signal and the scaled wavelet function, which can be
interpreted as a dilated band-pass filter because of its band-pass
like spectrum [2]. Sander et al. designed a 50/60Hz notch filter to
eliminate baseline drift from high resolution ECG Signal [3].
Markovsky et al. used band pass, kalman adaptive filter for
removal of resuscitation artifacts from human ECG signal [4].
For wavelet transform, daubechies wavelets were used
because the scaling functions of this wavelet filter are similar to
the shape of the ECG. From the decomposition of the ECG
signal it was seen that the low frequency component cause the
baseline shift, theses component were deducted to get a signal
without baseline drift. Also the high frequency components of
the signal were removed for getting denoised signal [22].
Below chapters sequentially elaborate the filtering
performances of five different filters.
II. METHODOLOGY
Fig.1. Block diagram of proposed methodology
“Fig. 1” illustrates the workflow of proposed
methodology. Filtering operations are performed for the selected
initial waveform (i.e. 4 sec data) “fig.2” of the full length noisy
ECG signal for better visualization of filtering results “fig.3-7”.
The statistical analysis results are shown in “fig.8-12”.
Devakishore Phurailatpam is with the Electrical & Electronics Engineering
Department, National Institute of Technology, Manipur, India (email:
bungcha@gmail.com).
Jitendra Singh Tamang is with Electronics & Communication Department,
SMIT, India (email: js.tamang@gmail.com).
Pankaj Kumar Giri is with and Applied Electronics & Instrumentation
Engineering Department, SMIT, India (email: pankajdav09@gmail.com).
ECG signal Noise Cancellation
Filtering Efficiency
Analysis
International Conference on Communication and Signal Processing, April 2-4, 2015, India
ISBN 978-1-4799-8080-2 Adhiparasakthi Engineering College, Melmaruvathur
060
Fig.2. Input Noisy ECG signal
A. Median Filter
The median filter is a nonlinear digital filtering technique,
often used to remove noise. Such noise reduction is a typical pre-
processing step to improve the results of later processing. The
main idea of the median filter is to run through the signal entry
by entry, replacing each entry with the median of neighboring
entries. The pattern of neighbors is called the "window", which
slides, entry by entry, over the entire signal. The function
considers the signal to be 0 beyond the endpoints. The output has
the same length as x.
For odd n, y (k) is the median of ( − ( − 1)/2: + ( −
1)/2).
For even n, y(k) is the median of ( − /2), ( − ( /2) +
1), . . . , ( + ( /2) − 1).
In this case, medfilt1 sorts the numbers, then takes the
average of the n/2 and (n/2) +1 elements with window=15.
Fig.3. Filtered ECG signal (using Median filter)
B. Low Pass Butter Worth Filter
The generalized equation representing an “nth” Order
Butterworth filter, the frequency response (1) is given as:
( ) = (1)
where, n represents the filter order, Omega ω is equal to 2πƒ ,
Epsilon ε is the maximum pass band gain, (Amax).
If Amax is defined at a frequency equal to the cut-off -3dB
corner point (ƒc), ε will then be equal to one and therefore ε2
will also be one. However, if you now wish to define Amax at a
different voltage gain value, for example 1dB, or 1.1220 (1dB =
20logAmax) then the new value of epsilon, ε is found by (2):
= (2)
where, H0 = the Maximum Pass band Gain, Amax. H1 = the
Minimum Pass band Gain.
Transpose the equation to give:
= 1.1220 = (1 + ) (3)
gives ε = 0.5088
The Frequency Response (3) of a filter can be defined
mathematically by its Transfer Function with the standard
Voltage Transfer Function H(jω) (4) written as:
( ) =
( )
( )
(4)
where:
Vout = the output signal voltage.
Vin = the input signal voltage.
ω = the radian frequency (2πƒ)
Fig.4. Filtered ECG signal (using LP Butterworth filter with order 10)
C. Finite Impulse Response (FIR) Filter
fir1 implements the classical method of windowed linear-
phase FIR digital filter design. It designs filters in standard low
pass, band pass, high pass, and band pass configurations.
Here the output of filtered signal contains the n + 1
coefficients of an order n low pass FIR filter. This is a
Hamming-windowed, linear-phase filter with cutoff frequency
Wn. The output filter coefficients, b, (5) are ordered in
descending powers of z.
061
( ) = (1) + (2) + ⋯. + ( + 1) (5)
Wn, the cutoff frequency, is a number between 0 and 1,
where 1 corresponds to half the sampling frequency (the Nyquist
frequency).
Fig.5. Filtered ECG signal (using FIR1 filter with order 15)
D. Weighted Moving Average
A weighted average is any average that has multiplying
factors to give different weights to data at different positions in
the sample window. Mathematically (6), the moving average is
the convolution of the datum points with a fixed weighting
function.
=
( ) ⋯ ( ) ( )
( ) ⋯
(6)
The denominator is a triangle number equal to
( )
. In the
more general case the denominator (8) will always be the sum of
the individual weights. When calculating the WMA across
successive values, the difference between the numerators of
WMAM+1 (9) and WMAM is − − ⋅⋅⋅ − − +
1. If we denote the sum + ⋅⋅⋅ + − + 1 by Total M,
then
+1 = + +1
− − +1
(7)
= + +
1 − − + 1 (8)
=
+ ( − 1) + ⋯ + 2 + 1
(9)
Fig.6. Filtered ECG signal (using WMA filter with window size 7)
E. Stationary Wavelet Transform (SWT)
The Stationary Wavelet Transform (SWT) is a wavelet
transform algorithm designed to overcome the lack of
translation-invariance of the discrete wavelet transform (DWT).
Translation-invariance is achieved by removing the down
samplers and up samplers in the DWT and up sampling the filter
coefficients by a factor of 2( )
in the jth
level of the algorithm.
Fig.7. Filtered ECG signal (using SWT filter with level 5 & wavelet ‘db1’)
III. STATISTICAL ANALYSIS
A. Amplitude
Fig.8. Amplitude changes after filtering ECG signal
0.10963
0.103966
0.109587
0.107243
0.108252
0.110528
0.1
0.102
0.104
0.106
0.108
0.11
0.112
Noisy FIR btrworth Median Moving
Window
Swt
AMP
062
Nonlinearities that give rise to amplitude distortion can be
analyzed from the amplitude changes after filtering from the bar graph
where, the input noisy ECG signal amplitude compared with filtered
ECG signals of five different filters.
B. Frequency
Fig.9. Frequency changes after filtering ECG signal
The use of these filters is to remove undesired frequency
imbedded in the ECG signal. The effect of filtering clearly
shown in SWT while, compared to other filters.
C. RMS
RMS is a statistical measure of the magnitude of a varying
quantity.
In the case of a set of n values { , , … … } the RMS (10) is
= ( + + ⋯. + ) (10)
Fig.10. RMS changes after filtering ECG signal
D. Power Spectral Density
The power spectral density (PSD) of a wide sense stationary
random process X (t) is computed (11) from the Fourier
transform of the autocorrelation function R(τ);
( ) = ( ).
∞
∞
(11)
where, the autocorrelation (12) function
( ) = [ ( + ) ( )] (12)
The nonparametric methods are methods in which the estimate
of PSD is made directly from a signal itself. One type of such
methods is called periodogram.
The periodogram estimate for PSD for discrete time
sequence , , ,…., is defined as square magnitude of the
Fourier transform (13) of data:
%( ) =
1
. . (13)
Fig.11. PSD changes after filtering ECG signal
E. Kurtosis
Fig.12. kurtosis changes after filtering ECG signal
In probability theory and statistics, kurtosis is any measure
of the "peakedness" of the probability distribution of a real-
valued random variable. There are various interpretations of
kurtosis, and of how particular measures should be interpreted;
these are primarily peakedness (width of peak), tail weight, and
0.007378
0.007509
0.007377
0.007281
0.007372
0.00687
0.0064
0.0066
0.0068
0.007
0.0072
0.0074
0.0076
Noisy FIR btrworth Median Moving
Window
Swt
FREQUECY
-91.103
-141.047
-166.582
-82.7685
-140.601
-93.0256
-180
-160
-140
-120
-100
-80
-60
-40
-20
0
Noisy FIR btrworth Median Moving
Window
Swt
RMS
-67.3009
-117.244
-142.78
-58.9664
-116.798
-68.9432
-160
-140
-120
-100
-80
-60
-40
-20
0
Noisy FIR btrworth Median Moving
Window
SwtPSD
4.98065
3.38912
4.92353
3.44756
4.04205
5.06952
0
1
2
3
4
5
6
Noisy FIR btrworth Median Moving
Window
Swt
KURTOSIS
063
lack of shoulders (distribution primarily peak and tails, not in
between).
The fourth standardized moment (14) is defined as
=
[( − ) ]
( [( − ) ])
= (14)
where, is the fourth moment about the mean and is the
standard deviation. The fourth standardized moment (15) is
bounded below by the squared skewness plus 1.
≥ ( ) +1 (15)
IV. DISCUSSION
The morphological changes in ECG waveforms are shown
in the filtering results of median and FIR filters where as there is
no significant change in the noisy and filtered signal in case of
LP Butterworth filter. There are no morphological changes (other
than filtering of noises) in the filtered signals of WMA and SWT
filters and the filtering result is quite promising. The statistical
parameters like PSD, RMS and kurtosis showed that the SWT
filter is performing better than the other filters where, the power
of signal is not utterly disturbed. Moving average type regression
line is drawn for PSD and RSM bar graphs and it clears shows
the variation of filtering operations other than SWT filter. WMA
filter is also performed well as compared to median, LP
Butterworth and FIR-1.
V. CONCLUSION
In many situations, the ECG is recorded during ambulatory
or strenuous conditions such that the signal is corrupted by
different types of noise, sometimes originating from another
physiological process of the body, so significant filtering
technique is required. Here, five different filters performances
are analyzed and their statistical parameters are extracted to
compute the filtering efficiencies. SWT is found to be a suitable
filter for noise cancellation from ECG signal. The further
research involves real-time implementation of these filters for
removal of different types of noise from ECG signal.
REFERENCES
[1] T. Nguyen G. Strang. “Wavelets and Filter Banks”. Wellesley-Cambridge
Press, 1996.
[2] C. Valens. “A really friendly guide to wavelets”. eBook, 2004. Available:
http://perso.wanadoo.fr.
[3] Sander A, Voss A, “Griessbach G “An Optimized Filter System For
Eliminating 50 Hz Interference from High Resolution ECG” Biomed Tech
Berl; 1995 40(4): 82-7.
[4] Markovsky Ivan A Anton, Van H and Sabine 2008 “Application of
Filtering methods for Removal of Resuscitation Artifacts from ECG
signals” IEEE conference of Engg. In Medicine and Biology, pp 13-16
[5] Hamilton PS, “A comparison of adaptive and nonadaptive filters for
reduction of power line interference in the ECG”, IEEE Trans Biomed Eng.
1996 Jan; 43(1):105-9.
[6] Ider YZ, Saki MC, Gcer HA, “Removal of power line interference in
signal-averaged electrocardiography systems”, IEEE Trans Biomed Eng.
1995 Jul; 42(7):731-5.
[7] Cramer E, McManus CD, Neubert D, “Estimation and removal of power
line interference in the electrocardiogram: a comparison of digital
approaches”, Comput Biomed Res. 1987 Feb; 20(1):12-28.
[8] McManus CD, Neubert KD, Cramer E, “Characterization and elimination
of AC noise in electrocardiograms: a comparison of digital filtering
methods”, Comput Biomed Res. 1993 Feb; 26(1):48-67.
[9] Sun Y, Chan K, Krishnan SM, “ECG signal conditioning by morphological
filtering”, Comput Biol Med. 2002 Nov;32(6):465-79.
[10] Levkov C, Mihov G, Ivanov R, Daskalov I, Christov I, Dotsinsky I,
“Removal of power-line interference from the ECG: a review of the
subtraction procedure”, Biomed Eng Online. 2005 Aug 23;4:50.
[11] Wu Y, Yang Y, “A new digital filter method for eliminating 50Hz
interference from the ECG”, Chinese journal of medical instrumentation.
1999 May;23(3):145-8.
[12] Ferdjallah M, Barr RE, “Frequency-domain digital filtering techniques for
the removal of powerline noise with application to the electrocardiogram”,
Comput Biomed Res. 1990 Oct; 23(5):473-89.
[13] P.K. Kulkarni, Vinod Kumar, H.K. Verma, “ Removal of power line
interference and baseline wonder using real time digital filter”, Proceedings
of international conference on computer applications in electrical
engineering, recent advances; Roorkee; India, pp 20-25 Sept. 1997.
[14] Ahlstrom M.L. and Tomkins W.J. “Digital filter for ECG real time
processing using microprocessors”, IEEE transactions on BME-32 pp.708-
713; 1885.
[15] Lynn,P.A., “ On line digital filter for biological filters: some fast designs
for small computers”, Medical Biological Engineering and computing
vol.15,pp 91-101.
[16] Challis RE, Kitney RI., “The design of digital filters for biomedical signal
processing. Part 3: The design of Butterworth and Chebychev filters”, J
Biomed Eng. 1983 Apr; 5(2):91-102.
[17] Robertson DG, Dowling JJ, “Design and responses of Butterworth and
critically damped digital filters”, J Electromyogr Kinesiol. 2003 Dec;
13(6):569-73.
[18] Zschorlich VR, “Digital filtering of EMG-signals”, Electromyogr Clin
Neurophysiol. 1989 Mar; 29(2):81-6.
[19] G.S.Furno and W.j. Tompkins, “A learning filter for reducing noise
interference,” IEEE trans. Biomed. Eng, vol. BME-30, pp. 234- 235, 1983.
[20] John G. Webster, Ed., “Medical Instrumentation-Application and Design”
Bosten: Houghton Mifflin, 1978.
[21] Tang Jing-tian, Zou Qing, Tang Yan, Liu Bin, Zhang Xiaokai, “Hilbert-
Huang Transform for ECG De-noising” Bioinformatics and Biomedical
Engineering, ICBBE, Year: 2007, pp. 664 – 667.
[22] Iffat Ara, Md. Najmul Hossain, S. M. Yahea Mahbub “Baseline Drift
Removal and De-Noising of the ECG Signal using Wavelet Transform”
International Journal of Computer Applications (0975 – 8887) Volume 95–
No.16, June 2014.
064

More Related Content

PDF
ECG signal denoising using a novel approach of adaptive filters for real-time...
PDF
Efficient data compression of ecg signal using discrete wavelet transform
PDF
Impedance Cardiography Filtering Using Non-Negative Least-Mean-Square Algorithm
PDF
11.a novel neural network classifier for brain computer
PDF
A novel neural network classifier for brain computer
PDF
Efficient reduction of PLI in ECG signal using new variable step size least m...
PDF
Suppression of power line interference correction of baselinewanders and
PDF
International Journal of Engineering Research and Development (IJERD)
ECG signal denoising using a novel approach of adaptive filters for real-time...
Efficient data compression of ecg signal using discrete wavelet transform
Impedance Cardiography Filtering Using Non-Negative Least-Mean-Square Algorithm
11.a novel neural network classifier for brain computer
A novel neural network classifier for brain computer
Efficient reduction of PLI in ECG signal using new variable step size least m...
Suppression of power line interference correction of baselinewanders and
International Journal of Engineering Research and Development (IJERD)

What's hot (19)

PDF
I047020950101
PDF
DESPECKLING OF SAR IMAGES BY OPTIMIZING AVERAGED POWER SPECTRAL VALUE IN CURV...
PDF
Synthesis and analysis of pzt using impedance method of reactance estimation
PDF
FPGA IMPLEMENTATION OF HUANG HILBERT TRANSFORM FOR CLASSIFICATION OF EPILEPTI...
PDF
Design of dfe based mimo communication system for mobile moving with high vel...
PDF
Estimation of Arm Joint Angles from Surface Electromyography signals using Ar...
PDF
Noisy Speech Enhancement Using Soft Thresholding on Selected Intrinsic Mode F...
PDF
Recovery of low frequency Signals from noisy data using Ensembled Empirical M...
PDF
Denoising of heart sound signal using wavelet transform
PDF
CDMA PARAMETER
PDF
Change Detection from Remotely Sensed Images Based on Stationary Wavelet Tran...
PDF
Remotesensing 06-02069
PDF
A scheme for optical pulse generation using optoelectronic phase locked loops
PDF
A scheme for optical pulse generation using
PDF
The Analysis of Aluminium Cantilever Beam with Piezoelectric Material by Chan...
PDF
Wavelength Conversion of a Laser Beam Using a Continuous- Wave Optical Parame...
PDF
IRJET- Disentangling Brain Activity from EEG Data using Logistic Regression, ...
PDF
ICICCE0324
PDF
Image denoising using new adaptive based median filter
I047020950101
DESPECKLING OF SAR IMAGES BY OPTIMIZING AVERAGED POWER SPECTRAL VALUE IN CURV...
Synthesis and analysis of pzt using impedance method of reactance estimation
FPGA IMPLEMENTATION OF HUANG HILBERT TRANSFORM FOR CLASSIFICATION OF EPILEPTI...
Design of dfe based mimo communication system for mobile moving with high vel...
Estimation of Arm Joint Angles from Surface Electromyography signals using Ar...
Noisy Speech Enhancement Using Soft Thresholding on Selected Intrinsic Mode F...
Recovery of low frequency Signals from noisy data using Ensembled Empirical M...
Denoising of heart sound signal using wavelet transform
CDMA PARAMETER
Change Detection from Remotely Sensed Images Based on Stationary Wavelet Tran...
Remotesensing 06-02069
A scheme for optical pulse generation using optoelectronic phase locked loops
A scheme for optical pulse generation using
The Analysis of Aluminium Cantilever Beam with Piezoelectric Material by Chan...
Wavelength Conversion of a Laser Beam Using a Continuous- Wave Optical Parame...
IRJET- Disentangling Brain Activity from EEG Data using Logistic Regression, ...
ICICCE0324
Image denoising using new adaptive based median filter
Ad

Viewers also liked (8)

PDF
What Is Noise Cancellation? | Phiaton
PPTX
Remote Active Noise Control
PPTX
Active noise control
PPTX
Nlms algorithm for adaptive filter
DOCX
ACTIVE NOISE CANCELLATION IN A LABORATORY DUCT USING FUZZY LOGIC AND NEURAL ...
PDF
M.Tech Thesis on Simulation and Hardware Implementation of NLMS algorithm on ...
PPT
Antinoise system & Noise Cancellation
PPTX
Noice canclellation using adaptive filters with adpative algorithms(LMS,NLMS,...
What Is Noise Cancellation? | Phiaton
Remote Active Noise Control
Active noise control
Nlms algorithm for adaptive filter
ACTIVE NOISE CANCELLATION IN A LABORATORY DUCT USING FUZZY LOGIC AND NEURAL ...
M.Tech Thesis on Simulation and Hardware Implementation of NLMS algorithm on ...
Antinoise system & Noise Cancellation
Noice canclellation using adaptive filters with adpative algorithms(LMS,NLMS,...
Ad

Similar to Multidimensional Approaches for Noise Cancellation of ECG signal (20)

PDF
Artifact elimination in ECG signal using wavelet transform
PDF
K010225156
PDF
E05613238
PDF
Ijetcas14 577
PDF
ECG SIGNAL DENOISING USING EMPIRICAL MODE DECOMPOSITION
PDF
Noise reduction in ECG signals for bio-telemetry
PDF
Bio-medical (EMG) Signal Analysis and Feature Extraction Using Wavelet Transform
PDF
DENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORM
PDF
Analysis of Butterworth and Chebyshev Filters for ECG Denoising Using Wavelets
PDF
M010417478
PDF
Fo3610221025
PDF
J041215358
PDF
De-Noising Corrupted ECG Signals By Empirical Mode Decomposition (EMD) With A...
PDF
I0414752
PPTX
DE-NOISING OF ECG USING WAVELETS AND MULTIWAVELETS
PDF
Noise Cancellation in ECG Signals using Computationally
PDF
Filtering Electrocardiographic Signals using filtered- X LMS algorithm
PDF
Noise reduction in ECG Signals for Bio-telemetryb
PDF
D ESIGN A ND I MPLEMENTATION OF D IGITAL F ILTER B ANK T O R EDUCE N O...
PDF
ADC Lab Analysis
Artifact elimination in ECG signal using wavelet transform
K010225156
E05613238
Ijetcas14 577
ECG SIGNAL DENOISING USING EMPIRICAL MODE DECOMPOSITION
Noise reduction in ECG signals for bio-telemetry
Bio-medical (EMG) Signal Analysis and Feature Extraction Using Wavelet Transform
DENOISING OF ECG SIGNAL USING FILTERS AND WAVELET TRANSFORM
Analysis of Butterworth and Chebyshev Filters for ECG Denoising Using Wavelets
M010417478
Fo3610221025
J041215358
De-Noising Corrupted ECG Signals By Empirical Mode Decomposition (EMD) With A...
I0414752
DE-NOISING OF ECG USING WAVELETS AND MULTIWAVELETS
Noise Cancellation in ECG Signals using Computationally
Filtering Electrocardiographic Signals using filtered- X LMS algorithm
Noise reduction in ECG Signals for Bio-telemetryb
D ESIGN A ND I MPLEMENTATION OF D IGITAL F ILTER B ANK T O R EDUCE N O...
ADC Lab Analysis

More from Sikkim Manipal Institute Of Technology (16)

PPTX
Application of different medical electrodes
PDF
Real Time Acquisition and Analysis of PCG and PPG Signals
PDF
Analysis of RS-Segment to Evaluate the Effect of Ventricular Depolarization d...
PDF
A Comparative Analysis of Neuropathic and Healthy EMG Signal Using PSD
PDF
TBC Algorithm to Detect R-Peak and HRV Analysis for Ischemia Heart Disease
PDF
QRS Complex Detection and Analysis of Cardiovascular Abnormalities: A Review
PDF
Neuromyopathy disease detection using wavelet packet based denoising technique
PDF
Evaluation of frequency domain features for myopathic emg signals in mat lab
PDF
Medical human factor engineering and workspace design
PPTX
PPTX
Action and Resting potential
PPTX
Introduction to biomedical instrumentation
Application of different medical electrodes
Real Time Acquisition and Analysis of PCG and PPG Signals
Analysis of RS-Segment to Evaluate the Effect of Ventricular Depolarization d...
A Comparative Analysis of Neuropathic and Healthy EMG Signal Using PSD
TBC Algorithm to Detect R-Peak and HRV Analysis for Ischemia Heart Disease
QRS Complex Detection and Analysis of Cardiovascular Abnormalities: A Review
Neuromyopathy disease detection using wavelet packet based denoising technique
Evaluation of frequency domain features for myopathic emg signals in mat lab
Medical human factor engineering and workspace design
Action and Resting potential
Introduction to biomedical instrumentation

Recently uploaded (20)

PDF
The CXO Playbook 2025 – Future-Ready Strategies for C-Suite Leaders Cerebrai...
PPT
Mechanical Engineering MATERIALS Selection
PDF
keyrequirementskkkkkkkkkkkkkkkkkkkkkkkkkkkkkkkkkkkkk
PPTX
UNIT-1 - COAL BASED THERMAL POWER PLANTS
DOCX
573137875-Attendance-Management-System-original
PDF
Well-logging-methods_new................
PPTX
Recipes for Real Time Voice AI WebRTC, SLMs and Open Source Software.pptx
PPTX
CYBER-CRIMES AND SECURITY A guide to understanding
PPTX
UNIT 4 Total Quality Management .pptx
PDF
Mitigating Risks through Effective Management for Enhancing Organizational Pe...
PPTX
Foundation to blockchain - A guide to Blockchain Tech
PDF
Mohammad Mahdi Farshadian CV - Prospective PhD Student 2026
PDF
July 2025 - Top 10 Read Articles in International Journal of Software Enginee...
PDF
Embodied AI: Ushering in the Next Era of Intelligent Systems
PPTX
bas. eng. economics group 4 presentation 1.pptx
PDF
Digital Logic Computer Design lecture notes
PPTX
KTU 2019 -S7-MCN 401 MODULE 2-VINAY.pptx
PDF
Operating System & Kernel Study Guide-1 - converted.pdf
PDF
TFEC-4-2020-Design-Guide-for-Timber-Roof-Trusses.pdf
PPTX
Construction Project Organization Group 2.pptx
The CXO Playbook 2025 – Future-Ready Strategies for C-Suite Leaders Cerebrai...
Mechanical Engineering MATERIALS Selection
keyrequirementskkkkkkkkkkkkkkkkkkkkkkkkkkkkkkkkkkkkk
UNIT-1 - COAL BASED THERMAL POWER PLANTS
573137875-Attendance-Management-System-original
Well-logging-methods_new................
Recipes for Real Time Voice AI WebRTC, SLMs and Open Source Software.pptx
CYBER-CRIMES AND SECURITY A guide to understanding
UNIT 4 Total Quality Management .pptx
Mitigating Risks through Effective Management for Enhancing Organizational Pe...
Foundation to blockchain - A guide to Blockchain Tech
Mohammad Mahdi Farshadian CV - Prospective PhD Student 2026
July 2025 - Top 10 Read Articles in International Journal of Software Enginee...
Embodied AI: Ushering in the Next Era of Intelligent Systems
bas. eng. economics group 4 presentation 1.pptx
Digital Logic Computer Design lecture notes
KTU 2019 -S7-MCN 401 MODULE 2-VINAY.pptx
Operating System & Kernel Study Guide-1 - converted.pdf
TFEC-4-2020-Design-Guide-for-Timber-Roof-Trusses.pdf
Construction Project Organization Group 2.pptx

Multidimensional Approaches for Noise Cancellation of ECG signal

  • 1. Multidimensional Approaches for Noise Cancellation of ECG signal Akash Kumar Bhoi, Karma Sonam Sherpa, Devakishore Phurailatpam, Jitendra Singh Tamang, Pankaj Kumar Giri Abstract: In many situations, the Electrocardiogram (ECG) is recorded during ambulatory or strenuous conditions such that the signal is corrupted by different types of noise, sometimes originating from another physiological process of the body. Hence, noise removal is an important aspect of signal processing. Here five different filters i.e. median, Low Pass Butter worth, FIR, Weighted Moving Average and Stationary Wavelet Transform (SWT) with their filtering effect on noisy ECG are presented. Comparative analyses among these filtering techniques are described and statically results are evaluated. Index Terms- Electrocardiography, Active noise reduction, Filters, Noise cancellation. I. INTRODUCTION A physician can detect a heart problem from this information and can suggest timely measures. But during the acquisition of ECG signal, it may get corrupted by different types of noises [21] which make it difficult for the physician to give his diagnosis. Power Line Interference (PLI) is one such kind of noise which superimposes on the vital information. The frequency range of ECG signal is 0.05Hz to 150Hz, and the frequency of the PLI noise is 50/60 Hz which lies within the frequency spectrum of the ECG signal, so for the meaningful and accurate detection, steps have to be taken to filter out or discard all these noise sources. Hamilton PS in his article compared adaptive and non adaptive filters for reduction of power line interference in the ECG [5]. Iders YZ, Saki MC, Gcer HA have developed a method for line interference reduction to be used in signal averaged electrocardiography [6]. Cramer E, McManus CD, Neubert D has introduced a global filtering approach. In this method two types of the digital filters are used where, one is using lest square method and other is using special summation method [7]. Different scientists have tried for removing the power line interference and base line wonder specifically from the ECG signal [8-17]. Zschorlich VR and Zschorlich VR, have also designed digital filters to cope with EMG signals [18-19]. Webster has explained the instrumentation requirements for the ECG [20]. Akash Kumar Bhoi is with the Applied Electronics & Instrumentation Engineering Department, Sikkim Manipal Institute of Technology (SMIT), India (email: akash730@gmail.com). Karma Sonam Sherpa is with the Electrical & Electronics Engineering Department, Sikkim Manipal Institute of Technology (SMIT), India (email: karmasherpa23@gmail.com). The wavelet coefficients represent a measure of similarity in the frequency content between a signal and a chosen wavelet function [1]. These coefficients are computed as a convolution of the signal and the scaled wavelet function, which can be interpreted as a dilated band-pass filter because of its band-pass like spectrum [2]. Sander et al. designed a 50/60Hz notch filter to eliminate baseline drift from high resolution ECG Signal [3]. Markovsky et al. used band pass, kalman adaptive filter for removal of resuscitation artifacts from human ECG signal [4]. For wavelet transform, daubechies wavelets were used because the scaling functions of this wavelet filter are similar to the shape of the ECG. From the decomposition of the ECG signal it was seen that the low frequency component cause the baseline shift, theses component were deducted to get a signal without baseline drift. Also the high frequency components of the signal were removed for getting denoised signal [22]. Below chapters sequentially elaborate the filtering performances of five different filters. II. METHODOLOGY Fig.1. Block diagram of proposed methodology “Fig. 1” illustrates the workflow of proposed methodology. Filtering operations are performed for the selected initial waveform (i.e. 4 sec data) “fig.2” of the full length noisy ECG signal for better visualization of filtering results “fig.3-7”. The statistical analysis results are shown in “fig.8-12”. Devakishore Phurailatpam is with the Electrical & Electronics Engineering Department, National Institute of Technology, Manipur, India (email: bungcha@gmail.com). Jitendra Singh Tamang is with Electronics & Communication Department, SMIT, India (email: js.tamang@gmail.com). Pankaj Kumar Giri is with and Applied Electronics & Instrumentation Engineering Department, SMIT, India (email: pankajdav09@gmail.com). ECG signal Noise Cancellation Filtering Efficiency Analysis International Conference on Communication and Signal Processing, April 2-4, 2015, India ISBN 978-1-4799-8080-2 Adhiparasakthi Engineering College, Melmaruvathur 060
  • 2. Fig.2. Input Noisy ECG signal A. Median Filter The median filter is a nonlinear digital filtering technique, often used to remove noise. Such noise reduction is a typical pre- processing step to improve the results of later processing. The main idea of the median filter is to run through the signal entry by entry, replacing each entry with the median of neighboring entries. The pattern of neighbors is called the "window", which slides, entry by entry, over the entire signal. The function considers the signal to be 0 beyond the endpoints. The output has the same length as x. For odd n, y (k) is the median of ( − ( − 1)/2: + ( − 1)/2). For even n, y(k) is the median of ( − /2), ( − ( /2) + 1), . . . , ( + ( /2) − 1). In this case, medfilt1 sorts the numbers, then takes the average of the n/2 and (n/2) +1 elements with window=15. Fig.3. Filtered ECG signal (using Median filter) B. Low Pass Butter Worth Filter The generalized equation representing an “nth” Order Butterworth filter, the frequency response (1) is given as: ( ) = (1) where, n represents the filter order, Omega ω is equal to 2πƒ , Epsilon ε is the maximum pass band gain, (Amax). If Amax is defined at a frequency equal to the cut-off -3dB corner point (ƒc), ε will then be equal to one and therefore ε2 will also be one. However, if you now wish to define Amax at a different voltage gain value, for example 1dB, or 1.1220 (1dB = 20logAmax) then the new value of epsilon, ε is found by (2): = (2) where, H0 = the Maximum Pass band Gain, Amax. H1 = the Minimum Pass band Gain. Transpose the equation to give: = 1.1220 = (1 + ) (3) gives ε = 0.5088 The Frequency Response (3) of a filter can be defined mathematically by its Transfer Function with the standard Voltage Transfer Function H(jω) (4) written as: ( ) = ( ) ( ) (4) where: Vout = the output signal voltage. Vin = the input signal voltage. ω = the radian frequency (2πƒ) Fig.4. Filtered ECG signal (using LP Butterworth filter with order 10) C. Finite Impulse Response (FIR) Filter fir1 implements the classical method of windowed linear- phase FIR digital filter design. It designs filters in standard low pass, band pass, high pass, and band pass configurations. Here the output of filtered signal contains the n + 1 coefficients of an order n low pass FIR filter. This is a Hamming-windowed, linear-phase filter with cutoff frequency Wn. The output filter coefficients, b, (5) are ordered in descending powers of z. 061
  • 3. ( ) = (1) + (2) + ⋯. + ( + 1) (5) Wn, the cutoff frequency, is a number between 0 and 1, where 1 corresponds to half the sampling frequency (the Nyquist frequency). Fig.5. Filtered ECG signal (using FIR1 filter with order 15) D. Weighted Moving Average A weighted average is any average that has multiplying factors to give different weights to data at different positions in the sample window. Mathematically (6), the moving average is the convolution of the datum points with a fixed weighting function. = ( ) ⋯ ( ) ( ) ( ) ⋯ (6) The denominator is a triangle number equal to ( ) . In the more general case the denominator (8) will always be the sum of the individual weights. When calculating the WMA across successive values, the difference between the numerators of WMAM+1 (9) and WMAM is − − ⋅⋅⋅ − − + 1. If we denote the sum + ⋅⋅⋅ + − + 1 by Total M, then +1 = + +1 − − +1 (7) = + + 1 − − + 1 (8) = + ( − 1) + ⋯ + 2 + 1 (9) Fig.6. Filtered ECG signal (using WMA filter with window size 7) E. Stationary Wavelet Transform (SWT) The Stationary Wavelet Transform (SWT) is a wavelet transform algorithm designed to overcome the lack of translation-invariance of the discrete wavelet transform (DWT). Translation-invariance is achieved by removing the down samplers and up samplers in the DWT and up sampling the filter coefficients by a factor of 2( ) in the jth level of the algorithm. Fig.7. Filtered ECG signal (using SWT filter with level 5 & wavelet ‘db1’) III. STATISTICAL ANALYSIS A. Amplitude Fig.8. Amplitude changes after filtering ECG signal 0.10963 0.103966 0.109587 0.107243 0.108252 0.110528 0.1 0.102 0.104 0.106 0.108 0.11 0.112 Noisy FIR btrworth Median Moving Window Swt AMP 062
  • 4. Nonlinearities that give rise to amplitude distortion can be analyzed from the amplitude changes after filtering from the bar graph where, the input noisy ECG signal amplitude compared with filtered ECG signals of five different filters. B. Frequency Fig.9. Frequency changes after filtering ECG signal The use of these filters is to remove undesired frequency imbedded in the ECG signal. The effect of filtering clearly shown in SWT while, compared to other filters. C. RMS RMS is a statistical measure of the magnitude of a varying quantity. In the case of a set of n values { , , … … } the RMS (10) is = ( + + ⋯. + ) (10) Fig.10. RMS changes after filtering ECG signal D. Power Spectral Density The power spectral density (PSD) of a wide sense stationary random process X (t) is computed (11) from the Fourier transform of the autocorrelation function R(τ); ( ) = ( ). ∞ ∞ (11) where, the autocorrelation (12) function ( ) = [ ( + ) ( )] (12) The nonparametric methods are methods in which the estimate of PSD is made directly from a signal itself. One type of such methods is called periodogram. The periodogram estimate for PSD for discrete time sequence , , ,…., is defined as square magnitude of the Fourier transform (13) of data: %( ) = 1 . . (13) Fig.11. PSD changes after filtering ECG signal E. Kurtosis Fig.12. kurtosis changes after filtering ECG signal In probability theory and statistics, kurtosis is any measure of the "peakedness" of the probability distribution of a real- valued random variable. There are various interpretations of kurtosis, and of how particular measures should be interpreted; these are primarily peakedness (width of peak), tail weight, and 0.007378 0.007509 0.007377 0.007281 0.007372 0.00687 0.0064 0.0066 0.0068 0.007 0.0072 0.0074 0.0076 Noisy FIR btrworth Median Moving Window Swt FREQUECY -91.103 -141.047 -166.582 -82.7685 -140.601 -93.0256 -180 -160 -140 -120 -100 -80 -60 -40 -20 0 Noisy FIR btrworth Median Moving Window Swt RMS -67.3009 -117.244 -142.78 -58.9664 -116.798 -68.9432 -160 -140 -120 -100 -80 -60 -40 -20 0 Noisy FIR btrworth Median Moving Window SwtPSD 4.98065 3.38912 4.92353 3.44756 4.04205 5.06952 0 1 2 3 4 5 6 Noisy FIR btrworth Median Moving Window Swt KURTOSIS 063
  • 5. lack of shoulders (distribution primarily peak and tails, not in between). The fourth standardized moment (14) is defined as = [( − ) ] ( [( − ) ]) = (14) where, is the fourth moment about the mean and is the standard deviation. The fourth standardized moment (15) is bounded below by the squared skewness plus 1. ≥ ( ) +1 (15) IV. DISCUSSION The morphological changes in ECG waveforms are shown in the filtering results of median and FIR filters where as there is no significant change in the noisy and filtered signal in case of LP Butterworth filter. There are no morphological changes (other than filtering of noises) in the filtered signals of WMA and SWT filters and the filtering result is quite promising. The statistical parameters like PSD, RMS and kurtosis showed that the SWT filter is performing better than the other filters where, the power of signal is not utterly disturbed. Moving average type regression line is drawn for PSD and RSM bar graphs and it clears shows the variation of filtering operations other than SWT filter. WMA filter is also performed well as compared to median, LP Butterworth and FIR-1. V. CONCLUSION In many situations, the ECG is recorded during ambulatory or strenuous conditions such that the signal is corrupted by different types of noise, sometimes originating from another physiological process of the body, so significant filtering technique is required. Here, five different filters performances are analyzed and their statistical parameters are extracted to compute the filtering efficiencies. SWT is found to be a suitable filter for noise cancellation from ECG signal. The further research involves real-time implementation of these filters for removal of different types of noise from ECG signal. REFERENCES [1] T. Nguyen G. Strang. “Wavelets and Filter Banks”. Wellesley-Cambridge Press, 1996. [2] C. Valens. “A really friendly guide to wavelets”. eBook, 2004. Available: http://perso.wanadoo.fr. [3] Sander A, Voss A, “Griessbach G “An Optimized Filter System For Eliminating 50 Hz Interference from High Resolution ECG” Biomed Tech Berl; 1995 40(4): 82-7. [4] Markovsky Ivan A Anton, Van H and Sabine 2008 “Application of Filtering methods for Removal of Resuscitation Artifacts from ECG signals” IEEE conference of Engg. In Medicine and Biology, pp 13-16 [5] Hamilton PS, “A comparison of adaptive and nonadaptive filters for reduction of power line interference in the ECG”, IEEE Trans Biomed Eng. 1996 Jan; 43(1):105-9. [6] Ider YZ, Saki MC, Gcer HA, “Removal of power line interference in signal-averaged electrocardiography systems”, IEEE Trans Biomed Eng. 1995 Jul; 42(7):731-5. [7] Cramer E, McManus CD, Neubert D, “Estimation and removal of power line interference in the electrocardiogram: a comparison of digital approaches”, Comput Biomed Res. 1987 Feb; 20(1):12-28. [8] McManus CD, Neubert KD, Cramer E, “Characterization and elimination of AC noise in electrocardiograms: a comparison of digital filtering methods”, Comput Biomed Res. 1993 Feb; 26(1):48-67. [9] Sun Y, Chan K, Krishnan SM, “ECG signal conditioning by morphological filtering”, Comput Biol Med. 2002 Nov;32(6):465-79. [10] Levkov C, Mihov G, Ivanov R, Daskalov I, Christov I, Dotsinsky I, “Removal of power-line interference from the ECG: a review of the subtraction procedure”, Biomed Eng Online. 2005 Aug 23;4:50. [11] Wu Y, Yang Y, “A new digital filter method for eliminating 50Hz interference from the ECG”, Chinese journal of medical instrumentation. 1999 May;23(3):145-8. [12] Ferdjallah M, Barr RE, “Frequency-domain digital filtering techniques for the removal of powerline noise with application to the electrocardiogram”, Comput Biomed Res. 1990 Oct; 23(5):473-89. [13] P.K. Kulkarni, Vinod Kumar, H.K. Verma, “ Removal of power line interference and baseline wonder using real time digital filter”, Proceedings of international conference on computer applications in electrical engineering, recent advances; Roorkee; India, pp 20-25 Sept. 1997. [14] Ahlstrom M.L. and Tomkins W.J. “Digital filter for ECG real time processing using microprocessors”, IEEE transactions on BME-32 pp.708- 713; 1885. [15] Lynn,P.A., “ On line digital filter for biological filters: some fast designs for small computers”, Medical Biological Engineering and computing vol.15,pp 91-101. [16] Challis RE, Kitney RI., “The design of digital filters for biomedical signal processing. Part 3: The design of Butterworth and Chebychev filters”, J Biomed Eng. 1983 Apr; 5(2):91-102. [17] Robertson DG, Dowling JJ, “Design and responses of Butterworth and critically damped digital filters”, J Electromyogr Kinesiol. 2003 Dec; 13(6):569-73. [18] Zschorlich VR, “Digital filtering of EMG-signals”, Electromyogr Clin Neurophysiol. 1989 Mar; 29(2):81-6. [19] G.S.Furno and W.j. Tompkins, “A learning filter for reducing noise interference,” IEEE trans. Biomed. Eng, vol. BME-30, pp. 234- 235, 1983. [20] John G. Webster, Ed., “Medical Instrumentation-Application and Design” Bosten: Houghton Mifflin, 1978. [21] Tang Jing-tian, Zou Qing, Tang Yan, Liu Bin, Zhang Xiaokai, “Hilbert- Huang Transform for ECG De-noising” Bioinformatics and Biomedical Engineering, ICBBE, Year: 2007, pp. 664 – 667. [22] Iffat Ara, Md. Najmul Hossain, S. M. Yahea Mahbub “Baseline Drift Removal and De-Noising of the ECG Signal using Wavelet Transform” International Journal of Computer Applications (0975 – 8887) Volume 95– No.16, June 2014. 064