QUALITY CONTROL IN
LABORATORY
WHAT IS QC ?
• QC is a technique that is used to detect &
correct errors before they result in a defective
product or service.
• It refers to the measures that must be
included during each assay run to verify that
the test is working properly.
• QC must be practical, achievable & affordable.
TERMINOLOGY
• PRECISION: This indicates how close test
measurements to each other when the same
test is run on the same sample repeatedly.
• Precision does not imply accuracy.
• Precision does not imply accuracy.
• ACCURACY: How close to the true value a
measurement is.
• The closer to the actual value, the more
accurate.
• STANDARD: This is a substance of constant
composition of sufficient purity to be used for
comparison purpose.
• CONTROL: This is a sample i.e. chemically &
physically similar to the unknown specimen.
physically similar to the unknown specimen.
• STANDARD DEVIATION: This is a statistical
expression of scatter or dispersion of values
around a central average value.
Calculation of Standard Deviation
Calculation of Standard Deviation
Coefficient of Variation
Coefficient of Variation
• The Coefficient of Variation (CV) is the
standard Deviation (SD) expressed as a
percentage of the mean
-Also known as Relative Standard
deviation (RSD)
• CV % = SD x 100
Mean
CALCULATION OF MEAN
Data set
(30.0, 32.0, 31.5, 33.5, 32.0, 33.0, 29.0,29.5, 31.0,
32.5, 34.5, 33.5, 31.5, 30.5, 30.0, 34.0,32.0, 32.0,
32.5, 34.5, 33.5, 31.5, 30.5, 30.0, 34.0,32.0, 32.0,
35.0, 32.5.) mg/dL
The sum of the values (X1 + X2 + X3 … X20)
divided by the number (n) of observations
The mean of these 20 observations is (639.5 Ă· 20)
= 32.0 mg/dL
P
PRECISE & INACCURATE
RECISE & INACCURATE
IMPRECISE & INACCURATE
IMPRECISE & INACCURATE
PRECISE & ACCURATE
PRECISE & ACCURATE
L. J. CHART
• L-J chart is a graph that quality control data is
plotted on to give a visual indication whether
a laboratory test is working well.
• It is named after S.LEVEY & E.R.JENNINGS in
1950.
Understanding Quality Control in Laboratory Practice
Understanding Quality Control in Laboratory Practice
lj with Manufacturer & Lab Calculated Mean & SD
213
234
255
276
297
318
339
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Glucose (Jan 2011) with Manufacturer Mean & Ranges (SD – 21 ; CV- 7.6%)
Level 3
266.58
269.31
272.04
274.77
277.5
280.23
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Glucose (Jan 2011) with Lab defined Mean & Ranges(SD – 2.73; CV- 0.99%)
Level 3
Normal Distribution
Normal Distribution
• All values are symmetrically distributed around the mean
• Characteristic “bell-shaped” curve
• Assumed for all quality control statistics
Understanding Quality Control in Laboratory Practice
INTERNAL QUALITY CONTROL
• Use of standard glassware , reagent
, equipment
• Well trained staff
• Selection of accurate & precise
method.
• At least one primary std. is included
with each batch of unknown specimen
analysis.
• Occasionally a different primary std . of
higher concentration is included to
check the reliability of routine primary
higher concentration is included to
check the reliability of routine primary
std.
• The batch result are accepted if the
values of control sera are within first SD.
Types of Errors
• Intrinsic : due to an imprecise &
inaccurate method.
• Systemic : all high values , all low
values.
• Random : technical errors.
• Intrinsic errors can be eliminated by
A) selecting precise & accurate method.
B) by using fresh batches of reagents.
• systemic errors can be eliminated by using diff.
concentration of primary std. & analyzing
concentration of primary std. & analyzing
control serum.
• Technical with a proper analytical application
can be correct random errors.
Systematic vs. Random Errors
Systematic vs. Random Errors
Systematic Error
Avoidable error due to
controllable variables in a
measurement.
Random Errors
Unavoidable errors that
are always present in any
measurement. Impossible
measurement. measurement. Impossible
to eliminate
PREVENTIVE PHASE
• Collection of specimen
• Separation of serum
• Specimen analysis
• Calculation of test values etc..
RETROSPECTIVE PHASE
• Optimum condition variance
• Routine condition variance
OPTIMUM CONDITION VARIANCE
• Refers to the results obtained under
optimum conditions i.e. by using
i)freshly prepared reagents
ii)by using standardized glassware
ROUTINE CONDITIONS VARIANCE
• Refers to the results obtained by using routine
requirements i.e. by using routinely stored
reagents and glassware in regular use.
WESTGARD RULES
• 1-2S
• 1-3S
• 2-2S
• 2-2S
• R-4S
• 4-1S
• 10X
• Multi control QC rules (WESTGARD RULES)
given by Dr. James Westgard of the University of
Wisconsin in an article in 1981 on laboratory
quality control that set the basis for evaluating
analytical run quality for medical laboratories.
• The Westgard system -based on the principles of
statistical process control used in manufacturing
statistical process control used in manufacturing
nationwide since the 1950s
• Six basic rules in the Westgard scheme: 1-3s, 2-
2s, R-4s, 1-2s, 4-1s, and 10x. These rules are used
individually or in combination (multi-rule) to
evaluate the quality of analytical runs.
• Detect random or systematic errors
• Warning 12SD or 1-2s:
It is violated if the single IQC
value exceeds the mean by ± 2SD.
• Rejection 22SD or 2-2s:
• This rule detects systematic error and is applied within and across
runs.
• It is violated within the run when two consecutive control values
exceed the "same" (mean + 2s or mean - 2s) limit.
• The rule is violated across runs when the previous value for a
particular control level exceeds the "same"
(mean + 2s or mean - 2s) limit.
(mean + 2s or mean - 2s) limit.
Within run violation Across run violation
• Rejection 13SD or 1-3s:
• It is violated when the single IQC
value exceeds the mean by ±3SD.
• This rule is applied within control
material only.
• The 1-3s rule identifies unacceptable
random error or possibly the
beginning of a large systematic error.
• Rejection 41SD or 4-1s:
It is violated if four consecutive IQC
values exceed the same mean plus 1s
or the same mean minus 1s control limit.
• Rejection 10x:
• This rule detects systematic bias and is applied both within and
across control materials.
• It is violated across control materials if the last 10 consecutive
values, regardless of control level, are on the same side of the
mean.
• The rule is violated within the control materials if the last 10
values for the same control level are on the same side of the
mean.
Westgard Procedure Flowchart
Westgard Procedure Flowchart
EXTERNAL QUALITY CONTROL
• All the participating laboratories daily
analyze the same lot of control material.
• The results are tabulated monthly & sent
the sponsoring groups for the data
the sponsoring groups for the data
analysis.
• Summary reports are prepared by the
program sponsor & are distributed to all
participating laboratories .
• The mean of values of all reference
laboratories is taken as the “ true “ or
correct value & is used for comparison
with the individual laboratory reported
values.
values.
• If the difference between the reported
value & the true value is statistically
significant then the reporting lab is
alerted.
THANK YOU

More Related Content

PPTX
Troubleshooting IQC / EQAS
PPTX
Quality control
PPTX
Share_Total_Quality_Management.pptx
PPTX
Control of analytical quality using stable control materials postgrad
PPT
Quality control in histopathology laboratory
PPTX
Troubleshooting iqc eqas 19.07.2018
PPT
WESTGARD RULES
PDF
Quality management system BY M .REJA
Troubleshooting IQC / EQAS
Quality control
Share_Total_Quality_Management.pptx
Control of analytical quality using stable control materials postgrad
Quality control in histopathology laboratory
Troubleshooting iqc eqas 19.07.2018
WESTGARD RULES
Quality management system BY M .REJA

Similar to Understanding Quality Control in Laboratory Practice (20)

PDF
quality control in clinical laboratory
PPTX
Westgard's rules and LJ (Levey Jennings) Charts.
PPTX
Clinical lab qc sethu
PPTX
medical -laboratory- quality- control --
PPTX
Quality assurance
PPTX
Quality Control in Clinical Laboratory Demo.pptx
PPTX
(IQC) - Internal Quality Control.pptx
PPTX
Quality Control in a Medical Testing Laboratory
PDF
Clinical_Chemistry_QC.pdffffffffffffffff
PPTX
Internal Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
PPTX
Qc and qa
PDF
Quality Control in Laboratory
PPT
Quality control lecture CPath master 2014 Ain Shams
PPT
General Principles of Quality control.ppt
PPTX
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
PPT
Laboratory Quality Control .ppt
PPTX
Quality assurance in the department of clinical biochemistry
PPSX
Basics of laboratory internal quality control, Ola Elgaddar, 2012
PPTX
QUALITY CONTROL IN BIOCHEMISTRY BMLS 3rd
PPT
Haemoglobin quality control by maintaining levey jennings chart
quality control in clinical laboratory
Westgard's rules and LJ (Levey Jennings) Charts.
Clinical lab qc sethu
medical -laboratory- quality- control --
Quality assurance
Quality Control in Clinical Laboratory Demo.pptx
(IQC) - Internal Quality Control.pptx
Quality Control in a Medical Testing Laboratory
Clinical_Chemistry_QC.pdffffffffffffffff
Internal Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
Qc and qa
Quality Control in Laboratory
Quality control lecture CPath master 2014 Ain Shams
General Principles of Quality control.ppt
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
Laboratory Quality Control .ppt
Quality assurance in the department of clinical biochemistry
Basics of laboratory internal quality control, Ola Elgaddar, 2012
QUALITY CONTROL IN BIOCHEMISTRY BMLS 3rd
Haemoglobin quality control by maintaining levey jennings chart
Ad

Recently uploaded (20)

PPTX
master seminar digital applications in india
PPTX
Final Presentation General Medicine 03-08-2024.pptx
PDF
GENETICS IN BIOLOGY IN SECONDARY LEVEL FORM 3
PDF
102 student loan defaulters named and shamed – Is someone you know on the list?
PPTX
human mycosis Human fungal infections are called human mycosis..pptx
PDF
2.FourierTransform-ShortQuestionswithAnswers.pdf
PDF
STATICS OF THE RIGID BODIES Hibbelers.pdf
PPTX
Microbial diseases, their pathogenesis and prophylaxis
PDF
Computing-Curriculum for Schools in Ghana
PDF
Saundersa Comprehensive Review for the NCLEX-RN Examination.pdf
PDF
A systematic review of self-coping strategies used by university students to ...
PDF
Microbial disease of the cardiovascular and lymphatic systems
PPTX
PPT- ENG7_QUARTER1_LESSON1_WEEK1. IMAGERY -DESCRIPTIONS pptx.pptx
 
PDF
RMMM.pdf make it easy to upload and study
PDF
VCE English Exam - Section C Student Revision Booklet
PDF
The Lost Whites of Pakistan by Jahanzaib Mughal.pdf
PDF
Anesthesia in Laparoscopic Surgery in India
PDF
grade 11-chemistry_fetena_net_5883.pdf teacher guide for all student
PPTX
Final Presentation General Medicine 03-08-2024.pptx
PDF
Supply Chain Operations Speaking Notes -ICLT Program
master seminar digital applications in india
Final Presentation General Medicine 03-08-2024.pptx
GENETICS IN BIOLOGY IN SECONDARY LEVEL FORM 3
102 student loan defaulters named and shamed – Is someone you know on the list?
human mycosis Human fungal infections are called human mycosis..pptx
2.FourierTransform-ShortQuestionswithAnswers.pdf
STATICS OF THE RIGID BODIES Hibbelers.pdf
Microbial diseases, their pathogenesis and prophylaxis
Computing-Curriculum for Schools in Ghana
Saundersa Comprehensive Review for the NCLEX-RN Examination.pdf
A systematic review of self-coping strategies used by university students to ...
Microbial disease of the cardiovascular and lymphatic systems
PPT- ENG7_QUARTER1_LESSON1_WEEK1. IMAGERY -DESCRIPTIONS pptx.pptx
 
RMMM.pdf make it easy to upload and study
VCE English Exam - Section C Student Revision Booklet
The Lost Whites of Pakistan by Jahanzaib Mughal.pdf
Anesthesia in Laparoscopic Surgery in India
grade 11-chemistry_fetena_net_5883.pdf teacher guide for all student
Final Presentation General Medicine 03-08-2024.pptx
Supply Chain Operations Speaking Notes -ICLT Program
Ad

Understanding Quality Control in Laboratory Practice

  • 2. WHAT IS QC ? • QC is a technique that is used to detect & correct errors before they result in a defective product or service. • It refers to the measures that must be included during each assay run to verify that the test is working properly. • QC must be practical, achievable & affordable.
  • 3. TERMINOLOGY • PRECISION: This indicates how close test measurements to each other when the same test is run on the same sample repeatedly. • Precision does not imply accuracy. • Precision does not imply accuracy. • ACCURACY: How close to the true value a measurement is. • The closer to the actual value, the more accurate.
  • 4. • STANDARD: This is a substance of constant composition of sufficient purity to be used for comparison purpose. • CONTROL: This is a sample i.e. chemically & physically similar to the unknown specimen. physically similar to the unknown specimen. • STANDARD DEVIATION: This is a statistical expression of scatter or dispersion of values around a central average value.
  • 5. Calculation of Standard Deviation Calculation of Standard Deviation
  • 6. Coefficient of Variation Coefficient of Variation • The Coefficient of Variation (CV) is the standard Deviation (SD) expressed as a percentage of the mean -Also known as Relative Standard deviation (RSD) • CV % = SD x 100 Mean
  • 7. CALCULATION OF MEAN Data set (30.0, 32.0, 31.5, 33.5, 32.0, 33.0, 29.0,29.5, 31.0, 32.5, 34.5, 33.5, 31.5, 30.5, 30.0, 34.0,32.0, 32.0, 32.5, 34.5, 33.5, 31.5, 30.5, 30.0, 34.0,32.0, 32.0, 35.0, 32.5.) mg/dL The sum of the values (X1 + X2 + X3 … X20) divided by the number (n) of observations The mean of these 20 observations is (639.5 Ă· 20) = 32.0 mg/dL
  • 11. L. J. CHART • L-J chart is a graph that quality control data is plotted on to give a visual indication whether a laboratory test is working well. • It is named after S.LEVEY & E.R.JENNINGS in 1950.
  • 14. lj with Manufacturer & Lab Calculated Mean & SD 213 234 255 276 297 318 339 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 Glucose (Jan 2011) with Manufacturer Mean & Ranges (SD – 21 ; CV- 7.6%) Level 3 266.58 269.31 272.04 274.77 277.5 280.23 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 Glucose (Jan 2011) with Lab defined Mean & Ranges(SD – 2.73; CV- 0.99%) Level 3
  • 15. Normal Distribution Normal Distribution • All values are symmetrically distributed around the mean • Characteristic “bell-shaped” curve • Assumed for all quality control statistics
  • 17. INTERNAL QUALITY CONTROL • Use of standard glassware , reagent , equipment • Well trained staff • Selection of accurate & precise method.
  • 18. • At least one primary std. is included with each batch of unknown specimen analysis. • Occasionally a different primary std . of higher concentration is included to check the reliability of routine primary higher concentration is included to check the reliability of routine primary std. • The batch result are accepted if the values of control sera are within first SD.
  • 19. Types of Errors • Intrinsic : due to an imprecise & inaccurate method. • Systemic : all high values , all low values. • Random : technical errors.
  • 20. • Intrinsic errors can be eliminated by A) selecting precise & accurate method. B) by using fresh batches of reagents. • systemic errors can be eliminated by using diff. concentration of primary std. & analyzing concentration of primary std. & analyzing control serum. • Technical with a proper analytical application can be correct random errors.
  • 21. Systematic vs. Random Errors Systematic vs. Random Errors Systematic Error Avoidable error due to controllable variables in a measurement. Random Errors Unavoidable errors that are always present in any measurement. Impossible measurement. measurement. Impossible to eliminate
  • 22. PREVENTIVE PHASE • Collection of specimen • Separation of serum • Specimen analysis • Calculation of test values etc..
  • 23. RETROSPECTIVE PHASE • Optimum condition variance • Routine condition variance
  • 24. OPTIMUM CONDITION VARIANCE • Refers to the results obtained under optimum conditions i.e. by using i)freshly prepared reagents ii)by using standardized glassware
  • 25. ROUTINE CONDITIONS VARIANCE • Refers to the results obtained by using routine requirements i.e. by using routinely stored reagents and glassware in regular use.
  • 26. WESTGARD RULES • 1-2S • 1-3S • 2-2S • 2-2S • R-4S • 4-1S • 10X
  • 27. • Multi control QC rules (WESTGARD RULES) given by Dr. James Westgard of the University of Wisconsin in an article in 1981 on laboratory quality control that set the basis for evaluating analytical run quality for medical laboratories. • The Westgard system -based on the principles of statistical process control used in manufacturing statistical process control used in manufacturing nationwide since the 1950s • Six basic rules in the Westgard scheme: 1-3s, 2- 2s, R-4s, 1-2s, 4-1s, and 10x. These rules are used individually or in combination (multi-rule) to evaluate the quality of analytical runs. • Detect random or systematic errors
  • 28. • Warning 12SD or 1-2s: It is violated if the single IQC value exceeds the mean by ± 2SD.
  • 29. • Rejection 22SD or 2-2s: • This rule detects systematic error and is applied within and across runs. • It is violated within the run when two consecutive control values exceed the "same" (mean + 2s or mean - 2s) limit. • The rule is violated across runs when the previous value for a particular control level exceeds the "same" (mean + 2s or mean - 2s) limit. (mean + 2s or mean - 2s) limit. Within run violation Across run violation
  • 30. • Rejection 13SD or 1-3s: • It is violated when the single IQC value exceeds the mean by ±3SD. • This rule is applied within control material only. • The 1-3s rule identifies unacceptable random error or possibly the beginning of a large systematic error. • Rejection 41SD or 4-1s: It is violated if four consecutive IQC values exceed the same mean plus 1s or the same mean minus 1s control limit.
  • 31. • Rejection 10x: • This rule detects systematic bias and is applied both within and across control materials. • It is violated across control materials if the last 10 consecutive values, regardless of control level, are on the same side of the mean. • The rule is violated within the control materials if the last 10 values for the same control level are on the same side of the mean.
  • 33. EXTERNAL QUALITY CONTROL • All the participating laboratories daily analyze the same lot of control material. • The results are tabulated monthly & sent the sponsoring groups for the data the sponsoring groups for the data analysis. • Summary reports are prepared by the program sponsor & are distributed to all participating laboratories .
  • 34. • The mean of values of all reference laboratories is taken as the “ true “ or correct value & is used for comparison with the individual laboratory reported values. values. • If the difference between the reported value & the true value is statistically significant then the reporting lab is alerted.