EMI Analyst™
EMI ANALYST™ Software Suite
I n t u i t i v e  A c c u r a t e  E f f e c t i v e
CS114 Analysis Settings
Using EMI Analyst to perform method CS114
Analyses for MIL-STD-461
EMI Analyst™
Use EMI Analyst to Perform CS114 Analyses
CS114 has conditional injection levels
1. Inject the power drive level recorded during pre-test calibration
or…
2. Inject twice the specified current limit,
…whichever is reached first
Settings for EMI Analyst are shown on the following page.
For more details, read on…
EMI Analyst™
EMI Analyst Settings for CS114 Dual Limit
The CS114 current limit is entered using
the Limit property form
Injected power level is defined using the
Threshold property form
EMI Analyst™
CS114 Limit Curves Map Directly to EMI
Analyst Limit Form
Limit Form Settings MIL-STD-461 CS114 Limit Curves
EMI Analyst™
Threshold Form Settings Capture CS114
Properties
Settings
• Option
• Pretest Calibration – Closed Loop Method
• Calibration Fixture Impedance
• 100 ohms
• Pretest Units
• Watts or dBm
• Margin Units
• Same units as Limit form (dBA)
• Margin
• 6 (dBA)
Threshold Form
EMI Analyst™
CS114 Settings Explained
• Limit Form
• The Limit form contains information about the CS114 current limit.
• Values entered on the Limit form table are vertices of the CS114 current limit.
• Threshold Form
• The Threshold form contains CS114 injection power level. Power levels are
determined by driving current equal to the current limit through a test fixture
whose series impedance is 100 ohms.
• When the Pretest Calibration – Closed Loop Method is selected, power levels
shown on the Threshold form are injected on the cable while monitoring
induced current.
• For CS114, if induced current limit exceeds Limit form values by more than 6
dB, injected power is reduced.
EMI Analyst™
What is CS114?
Method CS114 is a conducted susceptibility test method specified in
MIL-STD-461D, E, F, and G.
• The basic concept is to simulate currents that will be developed on platform
cabling from electromagnetic fields generated by antenna transmissions both
on and off the platform.
• An advantage of this type of requirement is that it provides data that can be
directly related to induced current levels measured during platform-level
evaluations.
(from MIL-STD-461F, paragraph A.5.13)
This type of test is often considered as a bulk current test since current is the parameter
measured. However, it is important to note that the test signal is inductively coupled and that
Faraday's law predicts an induced voltage in a circuit loop with the resultant current flow and
voltage distribution dependent on the various impedances present.
EMI Analyst™
CS114 Requirements Can Be Confusing
What CS114 says:
(from paragraph 5.13.2 of MIL-STD-461F)
The EUT shall not exhibit any malfunction, degradation
of performance, or deviation from specified indications
beyond the tolerances indicated in the individual
equipment or subsystem specification, when subjected
to an injection probe drive level which has been pre-
calibrated to the appropriate current limit shown in
Figure CS114-1 and is modulated as specified below. The
appropriate limit curve in Figure CS114-1 shall be
selected from Table VI. For EUTs intended to be installed
on ships or submarines, an additional common mode
limit of 77 dBμA is applicable from 4 kHz to 1 MHz on
complete power cables (highs and returns - common
mode test). Requirements are also met if the EUT is not
susceptible at forward power levels sensed by the
directional coupler that are below those determined
during calibration provided that the actual current
induced in the cable under test is 6 dB or greater than
the calibration limit.
What CS114 Means:
1. Record the power level needed to generate
the Limit current in a standard test fixture.
2. Inject the test signal on each cable bundle
using the lesser of:
A. The recorded power level
B. Two times the limit current
EMI Analyst™
Pretest Calibration
(from MIL-STD-461F, pargraph 5.13.3 and 5.13.4)
1. Place the injection probe around the center conductor of
the calibration fixture.
2. Terminate one end of the calibration fixture with a 50 ohm
load and terminate the other end with an attenuator
connected to measurement receiver A.
3. Set the signal generator to 10 kHz, unmodulated.
4. Increase the applied signal until measurement receiver A
indicates the current level specified in the applicable limit is
flowing in the center conductor of the calibration fixture.
5. Record the "forward power" to the injection probe indicated
on measurement receiver B.
6. Scan the frequency band from 10 kHz to 200 MHz and
record the forward power needed to maintain the required
current amplitude.
EMI Analyst™
Test Procedure
(from MIL-STD-461F, paragraph 5.13.3 and 5.13.4)
1. Place the injection and monitor probes around a cable bundle interfacing with EUT connector.
2. To minimize errors, maintain the same signal circuit that was used for calibration between the attenuator at
the calibration fixture (oscilloscope, coaxial cables, bulkhead connectors, additional attenuators, etc.) and
connect the circuit to the monitor probe. Additional attenuation may be used, if necessary.
3. Locate the monitor probe 5 cm from the connector. If the overall length of the connector and backshell
exceeds 5 cm, position the monitor probe as close to the connector's backshell as possible.
4. Position the injection probe 5 cm from the monitor probe.
Perform the following procedures on each cable bundle interfacing with each electrical connector on the EUT
including complete power cables (high sides and returns). Also perform the procedures on power cables with
the power returns and chassis grounds (green wires) excluded from the cable bundle. For connectors which
include both interconnecting leads and power, perform the procedures on the entire bundle, on the power
leads (including returns and grounds) grouped separately, and on the power leads grouped with the returns
and grounds removed.
(continued on next slide)
EMI Analyst™
Test Procedure
(from MIL-STD-461F, paragraph 5.13.3 and 5.13.4)
1. Turn on the EUT and allow sufficient time for stabilization.
2. Susceptibility evaluation.
a. Set the signal generator to 10 kHz with 1 kHz pulse modulation, 50% duty cycle.
b. Apply the forward power level determined under 5.13.3.4b(4) to the injection probe while
monitoring the induced current.
c. Scan the required frequency range in accordance with 4.3.10.4.1 and Table III while maintaining
the forward power level at the calibration level determined under 5.13.3.4b(4), or the maximum
current level for the applicable limit, whichever is less stringent.
d. Monitor the EUT for degradation of performance during testing.
e. Whenever susceptibility is noted, determine the threshold level in accordance with 4.3.10.4.3 and
verify that it is above the applicable requirement.
f. For EUTs with redundant cabling for safety critical reasons such as multiple data buses, use
simultaneous multi-cable injection techniques.
EMI Analyst™
CS114 Test Method and Current Limit
EMI Analyst™
Calibration Fixture
(from MIL-STD-461F, paragraph A.5.13)
The calibration fixture with terminations is a 50 ohm transmission line.
Since the injection probe is around the center conductor within the
fixture, a signal is being induced in the loop formed by the center
conductor, the two 50 ohm loads, and the structure of the fixture to
which the 50 ohm loads are terminated. From a loop circuit standpoint,
the two 50 ohm loads are in series, providing a total loop impedance of
100 ohms. Because of the transmission line configuration, inductance
effects are minimized. Measurement of induced current levels is
performed by measuring a corresponding voltage across one of the 50
ohm loads. Since the 50 ohm loads are in series for the induced signal,
the total drive voltage is actually two times that being measured.
The actual current that appears on a tested cable from the pre-
calibrated drive signal depends on the loop impedance associated with
the cable and the source impedance characteristics of the drive probe
and amplifier. If the loop impedance is low, such as would often result
with an overall shielded cable, currents greater than those in the
calibration fixture will result. The maximum required current is limited
to 6 dB above the pre-calibration level.
EMI Analyst™
Visit https://www.emisoftware.com
to learn how EMI Analyst™
can prevent electromagnetic
interference for your next project.

More Related Content

PDF
Understanding Reed-Solomon code
PDF
Transmissor de fm
PDF
vw-golf-7-wiring-diagrams-eng.pdf
PDF
Jerarquías de chomsky 2017 II
PPT
Bca 2nd sem-u-3.2-basic computer programming and micro programmed control
PPTX
Slide minggu ke 4 pertemuan 1
PDF
Motomil Desenho Explodido Jetmil
PDF
Marzu final
Understanding Reed-Solomon code
Transmissor de fm
vw-golf-7-wiring-diagrams-eng.pdf
Jerarquías de chomsky 2017 II
Bca 2nd sem-u-3.2-basic computer programming and micro programmed control
Slide minggu ke 4 pertemuan 1
Motomil Desenho Explodido Jetmil
Marzu final

What's hot (20)

PDF
Manual st10 flash-pro-usb (1)
PDF
Wireless Communication short talk
PDF
Esquemas eletricos-jetta-2.5
PDF
PA linearity
PDF
Manual Volvo
PPTX
MULTIPLEXER DAN DEMULTIPLEXER DIGITAL
PDF
4.Sampling and Hilbert Transform
PPSX
Bab iii transformasi z
DOCX
Qendrueshmeria statike dhe dinamike.Rajmonda Buhajloti
PDF
Esquema elétrico renault master 2.5
PPT
Complements
PDF
PA's - composições sugeridas
PPTX
Modul 02 konsep modulasi, modulasi analog, am
PPTX
Bcd counter
PPTX
Lecture Notes: EEEC6440315 Communication Systems - Inter Symbol Interference...
PDF
BÀI TẬP BỔ TRỢ THEO LESSON TIẾNG ANH - I-LEARN SMART WORLD 7 - CẢ NĂM - CÓ ĐÁ...
PDF
señales digitales
PPT
Transformasi Laplace
PDF
Verilog coding of demux 8 x1
PDF
Signal & system
Manual st10 flash-pro-usb (1)
Wireless Communication short talk
Esquemas eletricos-jetta-2.5
PA linearity
Manual Volvo
MULTIPLEXER DAN DEMULTIPLEXER DIGITAL
4.Sampling and Hilbert Transform
Bab iii transformasi z
Qendrueshmeria statike dhe dinamike.Rajmonda Buhajloti
Esquema elétrico renault master 2.5
Complements
PA's - composições sugeridas
Modul 02 konsep modulasi, modulasi analog, am
Bcd counter
Lecture Notes: EEEC6440315 Communication Systems - Inter Symbol Interference...
BÀI TẬP BỔ TRỢ THEO LESSON TIẾNG ANH - I-LEARN SMART WORLD 7 - CẢ NĂM - CÓ ĐÁ...
señales digitales
Transformasi Laplace
Verilog coding of demux 8 x1
Signal & system
Ad

Viewers also liked (17)

PPSX
An RS 422 IBIS Model for EMI Analyst
DOCX
Spectrum analyzer tutorial
PPTX
Radiated Emissions Example; LVDS - Low Voltage Differential Signaling
PPSX
How to extract shielded cable properties for radiated susceptibility
PPSX
How to Use IBIS Files in EMI Analyst™
PPSX
Conducted Susceptibility; CS Analyst Overview
PPTX
Presentation on emc testing and measurement
PPT
PSA-T Series Spectrum Analyser: PSA1301T/ PSA2701T
PDF
Updated! Debugging EMI Problems Using a Digital Oscilloscope
ODP
Metafísica Aristotèlica
DOCX
Ensayo herramientas ofimáticas
PDF
Práctica Formato de celdas
PDF
ТПЛМ 1091 Дистрибутивні канали й мережі. О.М.Горяїнов (2009)
PPT
PPT
EMI presentation
PPTX
Seis mil años de pecado y el regreso de Cristo - Resumen
An RS 422 IBIS Model for EMI Analyst
Spectrum analyzer tutorial
Radiated Emissions Example; LVDS - Low Voltage Differential Signaling
How to extract shielded cable properties for radiated susceptibility
How to Use IBIS Files in EMI Analyst™
Conducted Susceptibility; CS Analyst Overview
Presentation on emc testing and measurement
PSA-T Series Spectrum Analyser: PSA1301T/ PSA2701T
Updated! Debugging EMI Problems Using a Digital Oscilloscope
Metafísica Aristotèlica
Ensayo herramientas ofimáticas
Práctica Formato de celdas
ТПЛМ 1091 Дистрибутивні канали й мережі. О.М.Горяїнов (2009)
EMI presentation
Seis mil años de pecado y el regreso de Cristo - Resumen
Ad

Similar to Using EMI Analyst™ to Perform Method CS114 Analyses for MIL-STD-461 (20)

PPSX
CE Analyst Overview | Conducted Emissions Analysis Overview
PDF
Conducted susceptibility cs114_cs115_cs116 tests details
PDF
HDT: General overview April 1999
PDF
Ce test report riarudoll.com 国内ラブドール人気 リアルドール通販ショップ
PDF
EMI EMC RE CE
PDF
Iec61000 4-2 ed2-test_standard_noiseken_denkei
PDF
Design & Implementation of a Practical EMI Filter for High Frequency-High Pow...
PDF
Pd_Thesis and temperature variations for
PDF
EMC-Partner-EMC_LIVE_3day_2016-1.2-without-logo-1.pdf
PDF
MIL-STD-461G - CS
PPTX
Webinar: Simple Ideas to Make EMI Issues a Thing of the Past
PDF
Hdmi 1.3c Vs 1.3b
PDF
Emic unit v introduction 18 19
PDF
FrequencyExtendersWithBoosterModule_v12.pdf
PPTX
ELECTROMAGNETIC EMISSION AND SUSCEPTIBILITY STANDARDS AND SPECIFICATIONS,MIL ...
PPT
Si Intro(100413)
PDF
E0261031035
PDF
E0261031035
PDF
International Journal of Engineering and Science Invention (IJESI)
CE Analyst Overview | Conducted Emissions Analysis Overview
Conducted susceptibility cs114_cs115_cs116 tests details
HDT: General overview April 1999
Ce test report riarudoll.com 国内ラブドール人気 リアルドール通販ショップ
EMI EMC RE CE
Iec61000 4-2 ed2-test_standard_noiseken_denkei
Design & Implementation of a Practical EMI Filter for High Frequency-High Pow...
Pd_Thesis and temperature variations for
EMC-Partner-EMC_LIVE_3day_2016-1.2-without-logo-1.pdf
MIL-STD-461G - CS
Webinar: Simple Ideas to Make EMI Issues a Thing of the Past
Hdmi 1.3c Vs 1.3b
Emic unit v introduction 18 19
FrequencyExtendersWithBoosterModule_v12.pdf
ELECTROMAGNETIC EMISSION AND SUSCEPTIBILITY STANDARDS AND SPECIFICATIONS,MIL ...
Si Intro(100413)
E0261031035
E0261031035
International Journal of Engineering and Science Invention (IJESI)

Recently uploaded (20)

PDF
Improvement effect of pyrolyzed agro-food biochar on the properties of.pdf
PPTX
Management Information system : MIS-e-Business Systems.pptx
PDF
August -2025_Top10 Read_Articles_ijait.pdf
PDF
Exploratory_Data_Analysis_Fundamentals.pdf
PPTX
Sorting and Hashing in Data Structures with Algorithms, Techniques, Implement...
PDF
22EC502-MICROCONTROLLER AND INTERFACING-8051 MICROCONTROLLER.pdf
PDF
August 2025 - Top 10 Read Articles in Network Security & Its Applications
PPTX
CURRICULAM DESIGN engineering FOR CSE 2025.pptx
PDF
Human-AI Collaboration: Balancing Agentic AI and Autonomy in Hybrid Systems
PPTX
tack Data Structure with Array and Linked List Implementation, Push and Pop O...
PPTX
Fundamentals of Mechanical Engineering.pptx
PDF
Design Guidelines and solutions for Plastics parts
PPTX
AUTOMOTIVE ENGINE MANAGEMENT (MECHATRONICS).pptx
PDF
Categorization of Factors Affecting Classification Algorithms Selection
PDF
distributed database system" (DDBS) is often used to refer to both the distri...
PPTX
"Array and Linked List in Data Structures with Types, Operations, Implementat...
PPTX
ASME PCC-02 TRAINING -DESKTOP-NLE5HNP.pptx
PDF
Influence of Green Infrastructure on Residents’ Endorsement of the New Ecolog...
PDF
ChapteR012372321DFGDSFGDFGDFSGDFGDFGDFGSDFGDFGFD
PDF
UNIT no 1 INTRODUCTION TO DBMS NOTES.pdf
Improvement effect of pyrolyzed agro-food biochar on the properties of.pdf
Management Information system : MIS-e-Business Systems.pptx
August -2025_Top10 Read_Articles_ijait.pdf
Exploratory_Data_Analysis_Fundamentals.pdf
Sorting and Hashing in Data Structures with Algorithms, Techniques, Implement...
22EC502-MICROCONTROLLER AND INTERFACING-8051 MICROCONTROLLER.pdf
August 2025 - Top 10 Read Articles in Network Security & Its Applications
CURRICULAM DESIGN engineering FOR CSE 2025.pptx
Human-AI Collaboration: Balancing Agentic AI and Autonomy in Hybrid Systems
tack Data Structure with Array and Linked List Implementation, Push and Pop O...
Fundamentals of Mechanical Engineering.pptx
Design Guidelines and solutions for Plastics parts
AUTOMOTIVE ENGINE MANAGEMENT (MECHATRONICS).pptx
Categorization of Factors Affecting Classification Algorithms Selection
distributed database system" (DDBS) is often used to refer to both the distri...
"Array and Linked List in Data Structures with Types, Operations, Implementat...
ASME PCC-02 TRAINING -DESKTOP-NLE5HNP.pptx
Influence of Green Infrastructure on Residents’ Endorsement of the New Ecolog...
ChapteR012372321DFGDSFGDFGDFSGDFGDFGDFGSDFGDFGFD
UNIT no 1 INTRODUCTION TO DBMS NOTES.pdf

Using EMI Analyst™ to Perform Method CS114 Analyses for MIL-STD-461

  • 1. EMI Analyst™ EMI ANALYST™ Software Suite I n t u i t i v e  A c c u r a t e  E f f e c t i v e CS114 Analysis Settings Using EMI Analyst to perform method CS114 Analyses for MIL-STD-461
  • 2. EMI Analyst™ Use EMI Analyst to Perform CS114 Analyses CS114 has conditional injection levels 1. Inject the power drive level recorded during pre-test calibration or… 2. Inject twice the specified current limit, …whichever is reached first Settings for EMI Analyst are shown on the following page. For more details, read on…
  • 3. EMI Analyst™ EMI Analyst Settings for CS114 Dual Limit The CS114 current limit is entered using the Limit property form Injected power level is defined using the Threshold property form
  • 4. EMI Analyst™ CS114 Limit Curves Map Directly to EMI Analyst Limit Form Limit Form Settings MIL-STD-461 CS114 Limit Curves
  • 5. EMI Analyst™ Threshold Form Settings Capture CS114 Properties Settings • Option • Pretest Calibration – Closed Loop Method • Calibration Fixture Impedance • 100 ohms • Pretest Units • Watts or dBm • Margin Units • Same units as Limit form (dBA) • Margin • 6 (dBA) Threshold Form
  • 6. EMI Analyst™ CS114 Settings Explained • Limit Form • The Limit form contains information about the CS114 current limit. • Values entered on the Limit form table are vertices of the CS114 current limit. • Threshold Form • The Threshold form contains CS114 injection power level. Power levels are determined by driving current equal to the current limit through a test fixture whose series impedance is 100 ohms. • When the Pretest Calibration – Closed Loop Method is selected, power levels shown on the Threshold form are injected on the cable while monitoring induced current. • For CS114, if induced current limit exceeds Limit form values by more than 6 dB, injected power is reduced.
  • 7. EMI Analyst™ What is CS114? Method CS114 is a conducted susceptibility test method specified in MIL-STD-461D, E, F, and G. • The basic concept is to simulate currents that will be developed on platform cabling from electromagnetic fields generated by antenna transmissions both on and off the platform. • An advantage of this type of requirement is that it provides data that can be directly related to induced current levels measured during platform-level evaluations. (from MIL-STD-461F, paragraph A.5.13) This type of test is often considered as a bulk current test since current is the parameter measured. However, it is important to note that the test signal is inductively coupled and that Faraday's law predicts an induced voltage in a circuit loop with the resultant current flow and voltage distribution dependent on the various impedances present.
  • 8. EMI Analyst™ CS114 Requirements Can Be Confusing What CS114 says: (from paragraph 5.13.2 of MIL-STD-461F) The EUT shall not exhibit any malfunction, degradation of performance, or deviation from specified indications beyond the tolerances indicated in the individual equipment or subsystem specification, when subjected to an injection probe drive level which has been pre- calibrated to the appropriate current limit shown in Figure CS114-1 and is modulated as specified below. The appropriate limit curve in Figure CS114-1 shall be selected from Table VI. For EUTs intended to be installed on ships or submarines, an additional common mode limit of 77 dBμA is applicable from 4 kHz to 1 MHz on complete power cables (highs and returns - common mode test). Requirements are also met if the EUT is not susceptible at forward power levels sensed by the directional coupler that are below those determined during calibration provided that the actual current induced in the cable under test is 6 dB or greater than the calibration limit. What CS114 Means: 1. Record the power level needed to generate the Limit current in a standard test fixture. 2. Inject the test signal on each cable bundle using the lesser of: A. The recorded power level B. Two times the limit current
  • 9. EMI Analyst™ Pretest Calibration (from MIL-STD-461F, pargraph 5.13.3 and 5.13.4) 1. Place the injection probe around the center conductor of the calibration fixture. 2. Terminate one end of the calibration fixture with a 50 ohm load and terminate the other end with an attenuator connected to measurement receiver A. 3. Set the signal generator to 10 kHz, unmodulated. 4. Increase the applied signal until measurement receiver A indicates the current level specified in the applicable limit is flowing in the center conductor of the calibration fixture. 5. Record the "forward power" to the injection probe indicated on measurement receiver B. 6. Scan the frequency band from 10 kHz to 200 MHz and record the forward power needed to maintain the required current amplitude.
  • 10. EMI Analyst™ Test Procedure (from MIL-STD-461F, paragraph 5.13.3 and 5.13.4) 1. Place the injection and monitor probes around a cable bundle interfacing with EUT connector. 2. To minimize errors, maintain the same signal circuit that was used for calibration between the attenuator at the calibration fixture (oscilloscope, coaxial cables, bulkhead connectors, additional attenuators, etc.) and connect the circuit to the monitor probe. Additional attenuation may be used, if necessary. 3. Locate the monitor probe 5 cm from the connector. If the overall length of the connector and backshell exceeds 5 cm, position the monitor probe as close to the connector's backshell as possible. 4. Position the injection probe 5 cm from the monitor probe. Perform the following procedures on each cable bundle interfacing with each electrical connector on the EUT including complete power cables (high sides and returns). Also perform the procedures on power cables with the power returns and chassis grounds (green wires) excluded from the cable bundle. For connectors which include both interconnecting leads and power, perform the procedures on the entire bundle, on the power leads (including returns and grounds) grouped separately, and on the power leads grouped with the returns and grounds removed. (continued on next slide)
  • 11. EMI Analyst™ Test Procedure (from MIL-STD-461F, paragraph 5.13.3 and 5.13.4) 1. Turn on the EUT and allow sufficient time for stabilization. 2. Susceptibility evaluation. a. Set the signal generator to 10 kHz with 1 kHz pulse modulation, 50% duty cycle. b. Apply the forward power level determined under 5.13.3.4b(4) to the injection probe while monitoring the induced current. c. Scan the required frequency range in accordance with 4.3.10.4.1 and Table III while maintaining the forward power level at the calibration level determined under 5.13.3.4b(4), or the maximum current level for the applicable limit, whichever is less stringent. d. Monitor the EUT for degradation of performance during testing. e. Whenever susceptibility is noted, determine the threshold level in accordance with 4.3.10.4.3 and verify that it is above the applicable requirement. f. For EUTs with redundant cabling for safety critical reasons such as multiple data buses, use simultaneous multi-cable injection techniques.
  • 12. EMI Analyst™ CS114 Test Method and Current Limit
  • 13. EMI Analyst™ Calibration Fixture (from MIL-STD-461F, paragraph A.5.13) The calibration fixture with terminations is a 50 ohm transmission line. Since the injection probe is around the center conductor within the fixture, a signal is being induced in the loop formed by the center conductor, the two 50 ohm loads, and the structure of the fixture to which the 50 ohm loads are terminated. From a loop circuit standpoint, the two 50 ohm loads are in series, providing a total loop impedance of 100 ohms. Because of the transmission line configuration, inductance effects are minimized. Measurement of induced current levels is performed by measuring a corresponding voltage across one of the 50 ohm loads. Since the 50 ohm loads are in series for the induced signal, the total drive voltage is actually two times that being measured. The actual current that appears on a tested cable from the pre- calibrated drive signal depends on the loop impedance associated with the cable and the source impedance characteristics of the drive probe and amplifier. If the loop impedance is low, such as would often result with an overall shielded cable, currents greater than those in the calibration fixture will result. The maximum required current is limited to 6 dB above the pre-calibration level.
  • 14. EMI Analyst™ Visit https://www.emisoftware.com to learn how EMI Analyst™ can prevent electromagnetic interference for your next project.