Overview of Quality Control and
its implementation in the
laboratory
Dipesh Tamrakar
MSc. Clinical Biochemistry (IOM-TUTH)
BSc. MLT, CMLT
• Why do we need to about TQM/QA/QC?
• What are its practical approaches in the lab?
• Can we do it regularly in the lab of our setup?
Overview
• Brief on TQM, QA, QC
• Quality Indicators
• Lab errors
• Quality Control
• Implementation of QC in the lab
WHAT IS LABORATORY TEST ?
• “A laboratory test is intervention of any kind (diagnostic or
therapeutic) ; more likely to benefit than to harm the patient
and can be done at a reasonable cost and with reasonable
risk”.
• Laboratory tests provide information (with strong impact)
which help doctors to provide better and more effective care
for their patients.
• Laboratory results are component of 70% medical claims =
70% of medical decisions = 70% of medical records
ROLES OF LABORATORY MEDICINE
1. Perform laboratory investigations
2. Improve testing appropriateness
3. Improve patient outcomes
4. Improve decision making strategies
5. Improve cost effectiveness
6. Educate clinical users
7. Clinical optimization of operational efficiencies
8. Survey performance through appropriately designed quality control
9. Assist and improve on laboratory clinic interface
10. Improve information accessibility to the patient
11. Speed-up bringing in new diagnostic EVIDENCE BASED technologies
12. Redesign diagnostic paths towards an easier integrated service
STATUS and CHALLENGES
• Status: Laboratory experts are often perceived as a service for doctors.
• In order to have a successful healthcare ecosystem, they need to have the
same status as other experts in their own medical disciplines.
• Laboratory experts involve themselves more in structuring and ordering
tests
• Poor communication between medical departments and laboratories.
• Initiatives to make laboratory experts act as medical consultants are being
neglected that weakens the communication.
• Modern healthcare systems are oriented towards developing better
communication between doctors and laboratory experts.
• This form of communication has an enormous influence on the treatment
outcomes and medical expenses in general.
OFTEN KNOWN AS:
• Total Quality Control
• Total Quality Leadership
• Continuous Quality
Improvement
• Quality Management Science
• Quality assurance relates to how
a process is performed or how a
product is made.
• Quality control can be defined as
"part of quality management
focused on fulfilling quality
requirements."
• Quality control is more the
inspection aspect of quality
management.
• Quality Assurance - the overall program that ensures that the final
results reported by the laboratory are correct.
• It is much more concerned with :Right result from the Right test at
the Right time on the Right specimen from the Right patient with the
Right result interpretation based on correct reference data at the
Right price
Quality Control
• One of the most scientific and significant tool to correct errors in
analytical phase in the clinical laboratories
• Designed to monitor the routine performance of testing process
• Detect possible errors, reduce and correct deficiencies before the test
results are reported
• From service user: Quality of a good service for the price of the
product
• Internal quality control (IQC) and external quality assessment (EQA)
programs are used to assess and improve analytical performance.
IQC Vs EQA
• IQC is self-assessing tool by
using control samples for the
continuous and immediate
monitoring of the analytical part.
• IQC : to observe precise and
accurate results on daily
analytical phase
• EQA is intended for periodic and
retrospective observing of
laboratory results by an
autonomous external agency to
specify to the laboratory about
the accuracy or bias in their
systems methods and results.
• EQA : to monitor long term
accuracy and bias of their results
and stability of methods over a
longer period of time.
Aims of Quality assurance
• Accuracy and precision
• Satisfactory
• Quality report
• Economical
• Aims at preventing defects rather
than detecting the defects
• Minimize errors
• Timely report
Quality Assurance
An overall management plan to
guarantee the integrity of data (The
“system”)
PRECISION
The degree of fluctuation in the
measurements
Quality Control
A series of analytical measurements
used to assess the quality of the
analytical data (The “tools”)
ACCURACY
The closeness of measurements to the
true value
Quality Control is used to monitor both the precision and the accuracy
of the assay in order to provide reliable results.
• Precise and inaccurate
• Precise and accurate
• Imprecise and inaccurate
LABORATORY ERRORS ???
• From the patient’s perspective, every error is perceived
the same, whether it is made in pre-analytic, analytic or
post-analytic phase.
• Most of the errors occur outside the laboratory due to
various pre-analytic factors.
• A large number of errors occur in post-analytic phase as
well.
• In recent time, errors in analytic phase got significantly
reduced.
Overview of Quality Control and its implementation in the laboratory.pptx
How to implement a laboratory quality control
program???
FLOW OF LAB PROCESS
•Data and Lab
Management
•Safety
•Customer
Service
Patient/Client Prep
Sample Collection
Sample Receipt
and Accessioning
Sample Transport
Quality Control
Record Keeping
Reporting
Personnel Competency
Test Evaluations
Testing
The Quality Assurance Cycle
Overview of Quality Control and its implementation in the laboratory.pptx
Pre-analytical stage
• Test requisition slip
• Patient preparation
• Sample tube labeling
• Sample collection
• Sample transportation
• Reagent storage
• Reagent preparation
• Quality control sera
storage
• Callibrator storage
• QC alliquet storage
• Quality monitoring of
refrigerator, waterbath
• Maintenance and
cleaning of instruments
Machine maintenance: Daily, Weekly, Monthly and Annually
Preanalytical errors
• The importance of correctly carrying out all pre-analytical activities
performed by laboratory and non-laboratory staff in safeguarding the
quality and reliability of the final laboratory information.
• An error in any step of pre-analytical phase can adversely affect both
the quality of analytical results and the interpretation of information
provided.
Quality indicators of preanalytical phase
Misidentification error
• Misidentified requests
• Misidentified samples
• Samples with fewer than 2 identifiers
• Unlabeled samples
Inappropriate test request
• Requests without clinical questions (OPD)
• Inappropriate requests (OPD)
• Inappropriate requests (IPD)
Test transcription errors
• Erroneous data entry of
• Test name
• Missed test
• Added test
Unintelligible request
• Inappropriate sample type
• Samples collected in wrong container
• Incorrect fill level
• Inappropriate sample-anticoagulant volume
Unsuitable samples for transportation and storage problems
• Samples not properly stored before analysis
• Samples damaged during transportation
• Samples transported at inappropriate temperature
• Samples with excessive transportation time
• Contaminated samples
• Hemolyzed sample
• Clotted samples
• Inappropriate time in sample collection
Quality indicators for analytical phase
• Test with inappropriate IQC performances (CV higher than expected)
• Test uncovered by an EQA
• Unacceptable performances in EQA schemes
• Data transcription errors (manual or LIS failure)
Selection of Control & Calibrators
• Has a known concentration of the substance (analyte) being measured
• Used to adjust instrument, kit, test system in order to standardize the assay
Sources of QC samples:
• Appropriate diagnostic sample ( repeated samples for consecutive days)
• Obtained from:
• Another laboratory
• EQA provider
• Commercial product
• IQC depends on the analyzers available in the lab.
• EQA depends on availability of the resources
Characteristics of a Good Control
• The composition of the control material should be as similar to the
patient sample as possible, reacting in the same manner
• The analyte concentration should be at medically significant levels
• Stable under storage for longer period of time prior to preparation &
even after prep
• Low vial-to-vial variability
• Requiring a minimum preparation or ready to use
• Reasonable in price
Storage of QC Samples
• Controls are aliquoted into smaller ‘user friendly’ volumes for storage
• Establish a storage protocol:
• store at -20oC
• in use vials stored at 4oC
• use 0.5 ml vial maximum of one week
Monitoring QC Data
• Use Levey-Jennings chart
• Plot control values each run
• Acceptance of run
• Monitor precision and accuracy of repeated measurements
• Review charts at defined intervals, take necessary action, and
document
Findings Over Time
• Ideally should have control values clustered about the mean (+/-2 SD)
with little variation in the upward or downward direction
• Imprecision = large amount of scatter about the mean. Usually
caused by errors in technique
• Inaccuracy = may see as a trend or a shift, usually caused by change in
the testing process
• Random error = no pattern. Usually poor technique, malfunctioning
equipment
Overview of Quality Control and its implementation in the laboratory.pptx
Westgard Rules
• “Multirule Quality Control”
• Uses a combination of decision criteria or control rules
• Allows determination of whether an analytical run is “in-control” or
“out-of-control”
• Atleast run 2 control samples
12S rule R4S rule
13S rule 41S rule
22S rule 10X rule
Westgard – 12S Rule
• “warning rule”
• One of two control results falls
outside ±2SD
• Alerts tech to possible problems
• Not cause for rejecting a run
• Must then evaluate the 13S rule
12S rule
violation
Westgard – 13S Rule
• If either of the two control
results falls outside of ±3SD, rule
is violated
• Run must be rejected
• If 13S not violated, check 22S
13S rule
violation
Westgard – 22S Rule
• 2 consecutive control values for the
same level fall outside of ±2SD in the
same direction, or
• Both controls in the same run exceed
±2SD
• Patient results cannot be reported
• Requires corrective action
22S rule
violation
Westgard – R4S Rule
• One control exceeds the mean by –
2SD, and the other control exceeds
the mean by +2SD
• The range between the two results
will therefore exceed 4 SD
• Random error has occurred, test run
must be rejected
R4S rule
violation
Westgard – 41S Rule
• Requires control data from previous
runs
• Four consecutive QC results for one
level of control are outside ±1SD, or
• Both levels of control have
consecutive results that are outside
±1SD
41S rule
violation
Westgard – 10X Rule
• Requires control data from previous
runs
• Ten consecutive QC results for one
level of control are on one side of the
mean, or
• Both levels of control have five
consecutive results that are on the
same side of the mean
10x rule
violation
Westgard Multirule QC
When a rule is violated!!!!!
• Warning rule = use other rules to inspect the control points
• Rejection rule = “out of control”
• Stop testing
• Identify and correct problem
• Repeat testing on patient samples and controls
• Do not report patient results until problem is solved and controls
indicate proper performance
Types of Errors
Random=Fluctuations
Temperature
Personnel
Systematic
Reagent problem
Device problem
Solving “out-of-control” problems
Overview of Quality Control and its implementation in the laboratory.pptx
Quality indicators for post analytical phase
• Turn around time
• Laboratory reports
• Notifications of critical values
• Interpretative comments
Post-Analytical Stage
• Delta Check
• Report entry
• Report verification at 2 different levels
• Report print out
• Turn around time
• Emergency parameters: 1 hour
• Routine parameters: 2 hours
• Special immuno parameters: 2:30 hours
Sample storage for last 2 days at 4 degree C
How to get it done in the lab?
• Plan: define the course of action
• Organized process
• systematically monitored by using accurate measures
• Staff: who does what
• Lead the action
• Control the system process
Six sigma matrics
• Six Sigma metrics are being adopted as the universal
measure of quality to be applied to their processes and
the processes of their suppliers.
• Six Sigma provides a more quantitative framework for
evaluating process performance and more objective
evidence for process improvement.
• Two methods can be used to assess process
performance in terms of a sigma metric (+6S to -6S).
• outcomes by inspection & variation and predict process
performance.
• Sigma = (TEa − bias)/SD [TEa=allowable total error]
• Methods with Six Sigma performance are considered
“world class”; methods with sigma performance less than
THANK
YOU
Points to take away ???

More Related Content

PPT
Quality control in the medical laboratory
PPTX
Quality control in clinical laboratory
PPTX
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
PPTX
Quality Control in a Medical Testing Laboratory
PPTX
External Quality Assurance of Serological Testing
PPTX
Validation of lab instruments and quantitative test methods
PPT
Quality control
PPTX
Total Quality Management in Medical Laboratories
Quality control in the medical laboratory
Quality control in clinical laboratory
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
Quality Control in a Medical Testing Laboratory
External Quality Assurance of Serological Testing
Validation of lab instruments and quantitative test methods
Quality control
Total Quality Management in Medical Laboratories

What's hot (20)

PDF
Internal quality control in clinical laboratories hematology(2)
PDF
Quality assurance in a medical laboratory
PDF
Quality assurance in medical laboratory
PPTX
Quality Control In Clinical Laboratory
PPTX
Quality assurance in the department of clinical biochemistry
PPTX
Preanalytical quality control practices in clinical laboratory
PPTX
Post analytical variables in Laboratory
PPTX
quality control in clinical pathology
PPT
Laboratory Internal Quality Control presentation master revision, 2014
PPTX
Clinical lab qc sethu
PPT
Quality control lecture CPath master 2014 Ain Shams
PPTX
QUALITY control in hematology by Dr yogeeta Tanty.pptx
PDF
Total Quality Management (TQM) by Dr Anurag Yadav
PPT
Pre analytic and postanalytic test management
PPTX
ISO 15189 MEDICAL LABORTORIES REQUIRMENTS FOR QUALITY AND COMPETENCE
PDF
Quality control in clinical laboratories
PPTX
Quality control
PPTX
Quality assurance in haematology
PPTX
Understanding statistics in laboratory quality control
PPT
Automated Urine Analysis
Internal quality control in clinical laboratories hematology(2)
Quality assurance in a medical laboratory
Quality assurance in medical laboratory
Quality Control In Clinical Laboratory
Quality assurance in the department of clinical biochemistry
Preanalytical quality control practices in clinical laboratory
Post analytical variables in Laboratory
quality control in clinical pathology
Laboratory Internal Quality Control presentation master revision, 2014
Clinical lab qc sethu
Quality control lecture CPath master 2014 Ain Shams
QUALITY control in hematology by Dr yogeeta Tanty.pptx
Total Quality Management (TQM) by Dr Anurag Yadav
Pre analytic and postanalytic test management
ISO 15189 MEDICAL LABORTORIES REQUIRMENTS FOR QUALITY AND COMPETENCE
Quality control in clinical laboratories
Quality control
Quality assurance in haematology
Understanding statistics in laboratory quality control
Automated Urine Analysis
Ad

Similar to Overview of Quality Control and its implementation in the laboratory.pptx (20)

PPT
Laboratory Quality Control .ppt
PPTX
Internal Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
PDF
DMLT (2nd Year) : Quality Control, Standardization and Harmonization in Clini...
PPT
Qualitycontrolinclinicallaboratory
PPTX
Quality control
PPTX
Quality control in clinical laboratories
PDF
Quality Control in Laboratory
PPT
chem II 2019.ppt
PPTX
01 Introduction to Clinical Biochemistry II 03-07.pptx
PDF
qualitycontrolinbiochemistry-190731013722.pdf
PPTX
Quality Control
PPT
Quality management
PPTX
1. Introduction To Quality Assurance.pptx
PPTX
Troubleshooting iqc eqas 19.07.2018
PPTX
Troubleshooting IQC / EQAS
PDF
quality control in clinical laboratory
PPTX
Perez.pptx
PPTX
Assuring of Test Results based on Statistical tools.pptx
PPTX
Quality Control of zook xxookkh http.pptx
PPSX
Basics of laboratory internal quality control, Ola Elgaddar, 2012
Laboratory Quality Control .ppt
Internal Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
DMLT (2nd Year) : Quality Control, Standardization and Harmonization in Clini...
Qualitycontrolinclinicallaboratory
Quality control
Quality control in clinical laboratories
Quality Control in Laboratory
chem II 2019.ppt
01 Introduction to Clinical Biochemistry II 03-07.pptx
qualitycontrolinbiochemistry-190731013722.pdf
Quality Control
Quality management
1. Introduction To Quality Assurance.pptx
Troubleshooting iqc eqas 19.07.2018
Troubleshooting IQC / EQAS
quality control in clinical laboratory
Perez.pptx
Assuring of Test Results based on Statistical tools.pptx
Quality Control of zook xxookkh http.pptx
Basics of laboratory internal quality control, Ola Elgaddar, 2012
Ad

More from Dipesh Tamrakar (20)

PPTX
Therapeutic drug monitoring (TDM)
PPTX
Case presentation about TSH variants
PPTX
Newborn Screening
PPTX
DNA Sequencing
PPTX
Case triple vessel disease
PPTX
Case membranous nephropathy
PPTX
Case addisons disease
PPTX
Thyroid dysfunction - hypothyroidism
PPTX
Basics on statistical data analysis
PPTX
Thyroid gland
PPTX
Myocardial infraction & Cardiac Biomarkers (Laboratory Diagnosis)
PPT
Myocardial infarction and its laboratory diagnosis
PPT
Microscopy
PPTX
Diagnosis of Diabetes Mellitus
PPTX
Nucleotide metabolism
PPTX
Eicosanoids
PPTX
Triacylglycerol and compound lipid metabolism
PPTX
Phenylalanine & tyrosine amino acid metabolism
PPTX
Sulfur containing amino acid metabolism
PPTX
Inhibitors & uncouplers of oxidative phosphorylation & ETC
Therapeutic drug monitoring (TDM)
Case presentation about TSH variants
Newborn Screening
DNA Sequencing
Case triple vessel disease
Case membranous nephropathy
Case addisons disease
Thyroid dysfunction - hypothyroidism
Basics on statistical data analysis
Thyroid gland
Myocardial infraction & Cardiac Biomarkers (Laboratory Diagnosis)
Myocardial infarction and its laboratory diagnosis
Microscopy
Diagnosis of Diabetes Mellitus
Nucleotide metabolism
Eicosanoids
Triacylglycerol and compound lipid metabolism
Phenylalanine & tyrosine amino acid metabolism
Sulfur containing amino acid metabolism
Inhibitors & uncouplers of oxidative phosphorylation & ETC

Recently uploaded (20)

PPTX
First aid in common emergency conditions.pptx
PPTX
Medical aspects of impairment including all the domains mentioned in ICF
PDF
Selvita_Development-Strategy-2022-2025.pdf
PPTX
CBT FOR OCD TREATMENT WITHOUT MEDICATION
PPTX
PEDIATRIC OSCE, MBBS, by Dr. Sangit Chhantyal(IOM)..pptx
PPTX
AI_in_Pharmaceutical_Technology_Presentation.pptx
PDF
DAY-6. Summer class. Ppt. Cultural Nursing
PDF
Dr Masood Ahmed Expertise And Sucess Story
PDF
Pharmacology slides archer and nclex quest
PPTX
Nursing Care Aspects for High Risk newborn.pptx
PPTX
ABG advance Arterial Blood Gases Analysis
PPTX
Immunity....(shweta).................pptx
PPTX
Genaralised anxiety disorder presentation
PDF
MINERAL & VITAMIN CHARTS fggfdtujhfd.pdf
PPTX
HEMODYNAMICS - I DERANGEMENTS OF BODY FLUIDS.pptx
PPTX
Rheumatic heart diseases with Type 2 Diabetes Mellitus
PPT
Recent advances in Diagnosis of Autoimmune Disorders
PPTX
Trichuris trichiura infection
PPT
KULIAH UG WANITA Prof Endang 121110 (1).ppt
PPT
Adrenergic drugs (sympathomimetics ).ppt
First aid in common emergency conditions.pptx
Medical aspects of impairment including all the domains mentioned in ICF
Selvita_Development-Strategy-2022-2025.pdf
CBT FOR OCD TREATMENT WITHOUT MEDICATION
PEDIATRIC OSCE, MBBS, by Dr. Sangit Chhantyal(IOM)..pptx
AI_in_Pharmaceutical_Technology_Presentation.pptx
DAY-6. Summer class. Ppt. Cultural Nursing
Dr Masood Ahmed Expertise And Sucess Story
Pharmacology slides archer and nclex quest
Nursing Care Aspects for High Risk newborn.pptx
ABG advance Arterial Blood Gases Analysis
Immunity....(shweta).................pptx
Genaralised anxiety disorder presentation
MINERAL & VITAMIN CHARTS fggfdtujhfd.pdf
HEMODYNAMICS - I DERANGEMENTS OF BODY FLUIDS.pptx
Rheumatic heart diseases with Type 2 Diabetes Mellitus
Recent advances in Diagnosis of Autoimmune Disorders
Trichuris trichiura infection
KULIAH UG WANITA Prof Endang 121110 (1).ppt
Adrenergic drugs (sympathomimetics ).ppt

Overview of Quality Control and its implementation in the laboratory.pptx

  • 1. Overview of Quality Control and its implementation in the laboratory Dipesh Tamrakar MSc. Clinical Biochemistry (IOM-TUTH) BSc. MLT, CMLT
  • 2. • Why do we need to about TQM/QA/QC? • What are its practical approaches in the lab? • Can we do it regularly in the lab of our setup?
  • 3. Overview • Brief on TQM, QA, QC • Quality Indicators • Lab errors • Quality Control • Implementation of QC in the lab
  • 4. WHAT IS LABORATORY TEST ? • “A laboratory test is intervention of any kind (diagnostic or therapeutic) ; more likely to benefit than to harm the patient and can be done at a reasonable cost and with reasonable risk”. • Laboratory tests provide information (with strong impact) which help doctors to provide better and more effective care for their patients. • Laboratory results are component of 70% medical claims = 70% of medical decisions = 70% of medical records
  • 5. ROLES OF LABORATORY MEDICINE 1. Perform laboratory investigations 2. Improve testing appropriateness 3. Improve patient outcomes 4. Improve decision making strategies 5. Improve cost effectiveness 6. Educate clinical users 7. Clinical optimization of operational efficiencies 8. Survey performance through appropriately designed quality control 9. Assist and improve on laboratory clinic interface 10. Improve information accessibility to the patient 11. Speed-up bringing in new diagnostic EVIDENCE BASED technologies 12. Redesign diagnostic paths towards an easier integrated service
  • 6. STATUS and CHALLENGES • Status: Laboratory experts are often perceived as a service for doctors. • In order to have a successful healthcare ecosystem, they need to have the same status as other experts in their own medical disciplines. • Laboratory experts involve themselves more in structuring and ordering tests • Poor communication between medical departments and laboratories. • Initiatives to make laboratory experts act as medical consultants are being neglected that weakens the communication. • Modern healthcare systems are oriented towards developing better communication between doctors and laboratory experts. • This form of communication has an enormous influence on the treatment outcomes and medical expenses in general.
  • 7. OFTEN KNOWN AS: • Total Quality Control • Total Quality Leadership • Continuous Quality Improvement • Quality Management Science
  • 8. • Quality assurance relates to how a process is performed or how a product is made. • Quality control can be defined as "part of quality management focused on fulfilling quality requirements." • Quality control is more the inspection aspect of quality management.
  • 9. • Quality Assurance - the overall program that ensures that the final results reported by the laboratory are correct. • It is much more concerned with :Right result from the Right test at the Right time on the Right specimen from the Right patient with the Right result interpretation based on correct reference data at the Right price
  • 10. Quality Control • One of the most scientific and significant tool to correct errors in analytical phase in the clinical laboratories • Designed to monitor the routine performance of testing process • Detect possible errors, reduce and correct deficiencies before the test results are reported • From service user: Quality of a good service for the price of the product • Internal quality control (IQC) and external quality assessment (EQA) programs are used to assess and improve analytical performance.
  • 11. IQC Vs EQA • IQC is self-assessing tool by using control samples for the continuous and immediate monitoring of the analytical part. • IQC : to observe precise and accurate results on daily analytical phase • EQA is intended for periodic and retrospective observing of laboratory results by an autonomous external agency to specify to the laboratory about the accuracy or bias in their systems methods and results. • EQA : to monitor long term accuracy and bias of their results and stability of methods over a longer period of time.
  • 12. Aims of Quality assurance • Accuracy and precision • Satisfactory • Quality report • Economical • Aims at preventing defects rather than detecting the defects • Minimize errors • Timely report
  • 13. Quality Assurance An overall management plan to guarantee the integrity of data (The “system”) PRECISION The degree of fluctuation in the measurements Quality Control A series of analytical measurements used to assess the quality of the analytical data (The “tools”) ACCURACY The closeness of measurements to the true value Quality Control is used to monitor both the precision and the accuracy of the assay in order to provide reliable results.
  • 14. • Precise and inaccurate • Precise and accurate • Imprecise and inaccurate
  • 15. LABORATORY ERRORS ??? • From the patient’s perspective, every error is perceived the same, whether it is made in pre-analytic, analytic or post-analytic phase. • Most of the errors occur outside the laboratory due to various pre-analytic factors. • A large number of errors occur in post-analytic phase as well. • In recent time, errors in analytic phase got significantly reduced.
  • 17. How to implement a laboratory quality control program???
  • 18. FLOW OF LAB PROCESS
  • 19. •Data and Lab Management •Safety •Customer Service Patient/Client Prep Sample Collection Sample Receipt and Accessioning Sample Transport Quality Control Record Keeping Reporting Personnel Competency Test Evaluations Testing The Quality Assurance Cycle
  • 21. Pre-analytical stage • Test requisition slip • Patient preparation • Sample tube labeling • Sample collection • Sample transportation • Reagent storage • Reagent preparation • Quality control sera storage • Callibrator storage • QC alliquet storage • Quality monitoring of refrigerator, waterbath • Maintenance and cleaning of instruments Machine maintenance: Daily, Weekly, Monthly and Annually
  • 22. Preanalytical errors • The importance of correctly carrying out all pre-analytical activities performed by laboratory and non-laboratory staff in safeguarding the quality and reliability of the final laboratory information. • An error in any step of pre-analytical phase can adversely affect both the quality of analytical results and the interpretation of information provided.
  • 23. Quality indicators of preanalytical phase Misidentification error • Misidentified requests • Misidentified samples • Samples with fewer than 2 identifiers • Unlabeled samples Inappropriate test request • Requests without clinical questions (OPD) • Inappropriate requests (OPD) • Inappropriate requests (IPD)
  • 24. Test transcription errors • Erroneous data entry of • Test name • Missed test • Added test Unintelligible request • Inappropriate sample type • Samples collected in wrong container • Incorrect fill level • Inappropriate sample-anticoagulant volume
  • 25. Unsuitable samples for transportation and storage problems • Samples not properly stored before analysis • Samples damaged during transportation • Samples transported at inappropriate temperature • Samples with excessive transportation time • Contaminated samples • Hemolyzed sample • Clotted samples • Inappropriate time in sample collection
  • 26. Quality indicators for analytical phase • Test with inappropriate IQC performances (CV higher than expected) • Test uncovered by an EQA • Unacceptable performances in EQA schemes • Data transcription errors (manual or LIS failure)
  • 27. Selection of Control & Calibrators • Has a known concentration of the substance (analyte) being measured • Used to adjust instrument, kit, test system in order to standardize the assay Sources of QC samples: • Appropriate diagnostic sample ( repeated samples for consecutive days) • Obtained from: • Another laboratory • EQA provider • Commercial product • IQC depends on the analyzers available in the lab. • EQA depends on availability of the resources
  • 28. Characteristics of a Good Control • The composition of the control material should be as similar to the patient sample as possible, reacting in the same manner • The analyte concentration should be at medically significant levels • Stable under storage for longer period of time prior to preparation & even after prep • Low vial-to-vial variability • Requiring a minimum preparation or ready to use • Reasonable in price
  • 29. Storage of QC Samples • Controls are aliquoted into smaller ‘user friendly’ volumes for storage • Establish a storage protocol: • store at -20oC • in use vials stored at 4oC • use 0.5 ml vial maximum of one week
  • 30. Monitoring QC Data • Use Levey-Jennings chart • Plot control values each run • Acceptance of run • Monitor precision and accuracy of repeated measurements • Review charts at defined intervals, take necessary action, and document
  • 31. Findings Over Time • Ideally should have control values clustered about the mean (+/-2 SD) with little variation in the upward or downward direction • Imprecision = large amount of scatter about the mean. Usually caused by errors in technique • Inaccuracy = may see as a trend or a shift, usually caused by change in the testing process • Random error = no pattern. Usually poor technique, malfunctioning equipment
  • 33. Westgard Rules • “Multirule Quality Control” • Uses a combination of decision criteria or control rules • Allows determination of whether an analytical run is “in-control” or “out-of-control” • Atleast run 2 control samples 12S rule R4S rule 13S rule 41S rule 22S rule 10X rule
  • 34. Westgard – 12S Rule • “warning rule” • One of two control results falls outside ±2SD • Alerts tech to possible problems • Not cause for rejecting a run • Must then evaluate the 13S rule 12S rule violation
  • 35. Westgard – 13S Rule • If either of the two control results falls outside of ±3SD, rule is violated • Run must be rejected • If 13S not violated, check 22S 13S rule violation
  • 36. Westgard – 22S Rule • 2 consecutive control values for the same level fall outside of ±2SD in the same direction, or • Both controls in the same run exceed ±2SD • Patient results cannot be reported • Requires corrective action 22S rule violation
  • 37. Westgard – R4S Rule • One control exceeds the mean by – 2SD, and the other control exceeds the mean by +2SD • The range between the two results will therefore exceed 4 SD • Random error has occurred, test run must be rejected R4S rule violation
  • 38. Westgard – 41S Rule • Requires control data from previous runs • Four consecutive QC results for one level of control are outside ±1SD, or • Both levels of control have consecutive results that are outside ±1SD 41S rule violation
  • 39. Westgard – 10X Rule • Requires control data from previous runs • Ten consecutive QC results for one level of control are on one side of the mean, or • Both levels of control have five consecutive results that are on the same side of the mean 10x rule violation
  • 41. When a rule is violated!!!!! • Warning rule = use other rules to inspect the control points • Rejection rule = “out of control” • Stop testing • Identify and correct problem • Repeat testing on patient samples and controls • Do not report patient results until problem is solved and controls indicate proper performance
  • 45. Quality indicators for post analytical phase • Turn around time • Laboratory reports • Notifications of critical values • Interpretative comments
  • 46. Post-Analytical Stage • Delta Check • Report entry • Report verification at 2 different levels • Report print out • Turn around time • Emergency parameters: 1 hour • Routine parameters: 2 hours • Special immuno parameters: 2:30 hours
  • 47. Sample storage for last 2 days at 4 degree C
  • 48. How to get it done in the lab? • Plan: define the course of action • Organized process • systematically monitored by using accurate measures • Staff: who does what • Lead the action • Control the system process
  • 49. Six sigma matrics • Six Sigma metrics are being adopted as the universal measure of quality to be applied to their processes and the processes of their suppliers. • Six Sigma provides a more quantitative framework for evaluating process performance and more objective evidence for process improvement. • Two methods can be used to assess process performance in terms of a sigma metric (+6S to -6S). • outcomes by inspection & variation and predict process performance. • Sigma = (TEa − bias)/SD [TEa=allowable total error] • Methods with Six Sigma performance are considered “world class”; methods with sigma performance less than
  • 51. Points to take away ???